Dergiler / Natural and Engineering Sciences / 2018 / Cilt: 3 - Sayı: 3

Characterization and Construction of a Robust and Elastic Wall-Less Flow Phantom for High Pressure Flow Rate Using Doppler Ultrasound Applications

Sayfa
359–377
Yayın Tarihi
2018-10-10
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

This wall-less flow phantom will be a suitable instrument for examining in-vitro research studies.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:A Doppler ultrasound is a noninvasive test that can be used to estimate the blood flow through the vessels.
📊
Temel Bulgular & Veri:Presently, few flow phantoms are being used to be qualified for long-term utilize and storage with high physiological flow rate Doppler ultrasound.
💡
Sonuç & Katkı:This wall-less flow phantom will be a suitable instrument for examining in-vitro research studies.
🔬Tür:Araştırma Makalesi
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:BMF · TMM · Wall-less Flow Phantom

Abstract

A Doppler ultrasound is a noninvasive test that can be used to estimate the blood flow through the vessels. Presently, few flow phantoms are being used to be qualified for long-term utilize and storage with high physiological flow rate Doppler ultrasound. The main drawback of the two hydrogel materials items (Konjac (K) and carrageenan (C) (KC)) that it is not fit for long-term storage and easy to deteriorate. Thus, this research study focuses on the characterization and construction of a robust and elastic wall-less flow phantom with suitable acoustical properties of TMM. The mechanisms for the fabrication of a wall-less flow phantom utilizing a physically strong material such as K, C, and gelatin (bovine skin)-based TMM were explained. In addition, the clinical ultrasound (Hitachi Avius (HI)) system was used as the main instrument for data acquisition. Vessel mimicking material (VMM) with dimensions of 15.0 mm depth equal to those of human common carotid arteries (CCA) were obtained with pulsatile flow. The acoustical properties (speed of sound and attenuation were 1533±2 m/s and 0.2 dB/cm. MHz, respectively) of a new TMM were agreed with the IEC 61685 standards. Furthermore, the velocity percentages error were decreased with increase in the Doppler angle (the lowest % error (3%) it was at 53◦). The gelatin from bovine skin was a proper material to be added to KC to enhance the strength of TMM during for long-term utilize and storage of high-flow of blood mimicking Fluid (BMF). This wall-less flow phantom will be a suitable instrument for examining in-vitro research studies.

🔖

Bu Makaleye Atıf Yap

AMMAR, A. O., MATJAFRİ, M., SUARDİ, N., OQLAT, M. A., OQLAT, A. A., ABDELRAHMAN, M. A., FARHAT, O., AHMAD, M. S., ALKHATEB, B. N., GEMANAM, S. J., SHALBİ, S. M., ABDALRHEEM, R., SHİPLİ, M., MARASHDEH, M. (2018). Characterization and Construction of a Robust and Elastic Wall-Less Flow Phantom for High Pressure Flow Rate Using Doppler Ultrasound Applications. Natural and Engineering Sciences, Cilt: 3 - Sayı: 3, 359–377. https://ajindex.com/articles/793221

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (12)

  1. Bishop, C., Powell, S., & Rutt, D. (1986). Transcranial Doppler measurement of middle cerebral artery blood flow velocity: a validation study. Stroke, 17(5), 913-915.
  2. Blake, J. R., Meagher, S., Fraser, K. H., Easson, W. J., & Hoskins, P. R. (2008). A method to estimate wall shear rate with a clinical ultrasound scanner. Ultrasound in medicine & biology, 34(5), 760-774.
  3. Brands, P. J., Hoeks, A. P., Hofstra, L., & Reneman, R. S. (1995). A noninvasive method to estimate wall shear rate using ultrasound. Ultrasound in medicine & biology, 21(2), 171-185.
  4. Browne, J. E. (2014). A review of Doppler ultrasound quality assurance protocols and test devices. Physica Medica, 30(7), 742-751.
  5. Chen, J.-H., Pu, Y.-S., Liu, S.-P., & Chiu, T.-Y. (1993). Renal hemodynamics in patients with obstructive uropathy evaluated by duplex Doppler sonography. The Journal of urology, 150(1), 18-21.
  6. Commission, I. E. (2001). Ultrasonics—Flow measurement systems: Flow test object. Draft IEC, 1685. Doucette, J. W., Corl, P. D., Payne, H. M., Flynn, A. E., Goto, M., Nassi, M., & Segal, J. (1992). Validation of a Doppler guide wire for intravascular measurement of coronary artery flow velocity. Circulation, 85(5), 1899-1911.
  7. Evans, D. H. (2000). Doppler signal analysis. Ultrasound in medicine & biology, 26, S13-S15. Gerhard-Herman, M., Gardin, J. M., Jaff, M., Mohler, E., Roman, M., & Naqvi, T. Z. (2006). Guidelines for noninvasive vascular laboratory testing: a report from the American Society of Echocardiography and the Society for Vascular Medicine and Biology. Vascular Medicine, 11(3), 183-200.
  8. Grant, E. G., Benson, C. B., Moneta, G. L., Alexandrov, A. V., Baker, J. D., Bluth, E. I.,. Hertzberg, B. S. (2003). Carotid artery stenosis: gray-scale and Doppler US diagnosis—Society of Radiologists in Ultrasound Consensus Conference. Radiology, 229(2), 340-346.
  9. Hoskins, P. (1999). A comparison of single-and dual-beam methods for maximum velocity estimation. Ultrasound in medicine & biology, 25(4), 583-592.
  10. Hoskins, P. (2002). Ultrasound techniques for measurement of blood flow and tissue motion. Biorheology, 39(3, 4), 451-459.
  11. Hoskins, P. R. (2008). Simulation and validation of arterial ultrasound imaging and blood flow. Ultrasound in Medicine and Biology, 34(5), 693-717.
  12. Hoskins, P. R., Soldan, M., Fortune, S., Inglis, S., Anderson, T., & Plevris, J. (2010). Validation of endoscopic ultrasound measured flow rate in the azygos vein using a flow phantom. Ultrasound in medicine & biology, 3