Evaluation of Detection Limits with Wavelength Dispersive X-Ray Fluorescence Spectrometry

Sayfa
38–47
Yayın Tarihi
2017-11-30
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

It is concluded that WDXRF spectrometry is a suitable technique for the determination of detection limits.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:The influence of analyte mass concentration on the determination of detection limits has been investigated by using WDXRF (Wavelength dispersive X­ ray fluorescence) spectrometry.
📊
Temel Bulgular & Veri:From this results corresponding to the interval of 5­25 times the best value of the detection limits of analytical concentration values are found to be the ideal values.
💡
Sonuç & Katkı:It is concluded that WDXRF spectrometry is a suitable technique for the determination of detection limits.
🔬Tür:Araştırma Makalesi
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:analyte · detection limits · WDXRF

Abstract

The influence of analyte mass concentration on the determination of detection limits has been investigated by using WDXRF (Wavelength dispersive X­ ray fluorescence) spectrometry. It is aimed to obtain the analyte mass concentration that is necessary to find the best detection limit values. Ti, V, Cr, Mn, Fe, Ni, Cu, Zn, As, Y, Zr, Nb, Mo, Ag, Cd, Sb, Nd, Gd, Er elements and binary and triple combination of these elements have been used. The samples have been selected considering the absorption edges due to the importance of the effect of matrix absorption and enhancement. The elements which have absorption edges that are close to each other have been used observe the matrix absorption and enhancement. The results of this study are given in tables and graphs. From this results corresponding to the interval of 5­25 times the best value of the detection limits of analytical concentration values are found to be the ideal values. In this study, the experimental apparatus and sample preparation procedures used are presented. It is concluded that WDXRF spectrometry is a suitable technique for the determination of detection limits.

🔖