Transcatheter valve implantation into descending aorta due to the insufficient guide wire support

Yetersiz kılavuz tel desteğine bağlı inen aorta transkate ter kapak implantasyonu

Sayfa
363–365
Yayın Tarihi
2022-11-05
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

Sonrasındabaşka bir transkateter aort kapak (23 mm EdwardsSAPIEN XT) aort kapak seviyesine başarılı bir şekildeyerleştirildi.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Transkateter aort kapak implantasyonu (TAVI) ciddiaort kapak darlığı olup ameliyat edilemeyen hastalardacerrahiye alternatif bir tedavi seçeneğidir.
📊
Temel Bulgular & Veri:Ekokardiyografide hesaplanan aort kapakalanı 0.8 cm2 ile birlikte ortalama transaortik gradyan55 mmHg idi.
💡
Sonuç & Katkı:Sonrasındabaşka bir transkateter aort kapak (23 mm EdwardsSAPIEN XT) aort kapak seviyesine başarılı bir şekildeyerleştirildi.
🔬Tür:Araştırma Makalesi
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Alan Parametreleri

Özet

Transkateter aort kapak implantasyonu (TAVI) ciddiaort kapak darlığı olup ameliyat edilemeyen hastalardacerrahiye alternatif bir tedavi seçeneğidir. GünümüzdeTAVI kullanımı giderek artsa dahi işlemin komplikasyonriski halen yüksektir. Seksen altı yaşında ciddi aortdarlığı olan bir kadın hasta kliniğimize anjina yakınmasıile başvurdu. Ekokardiyografide hesaplanan aort kapakalanı 0.8 cm2 ile birlikte ortalama transaortik gradyan55 mmHg idi. Cerrahi risk skorları yüksek olduğu için(Logistic EuroSCORE= %31.21), hastaya transfemoralyol ile TAVI planlandı. Amplatz Super Stiff PTFE kaplıkılavuz tel (7 cm Bentson-Tip) sol ventriküle yerleştirildi.Aort köküne balon ile predilatasyon yapıldı. 23 mmEdwards SAPIEN XT kapak aorta ilerletildi; ancakkılavuz telin sol ventrikülden geri gelmesi nedeniyle,kapak aort kapak seviyesine yerleştirilemedi. Transkateteraort kapak çekilerek inen aorta yerleştirildi. Sonrasındabaşka bir transkateter aort kapak (23 mm EdwardsSAPIEN XT) aort kapak seviyesine başarılı bir şekildeyerleştirildi.

Abstract

Transcatheter aortic valve implantation (TAVI) is analternative therapy to surgical aortic valve replacement ininoperable patients with severe aortic stenosis. Despite theincreasing use of TAVI currently, the potential risk for life-threatening complications is still regarded to be high. An86-year-old female patient was admitted to our clinic withangina. On echocardiography, the mean transaortic gradientwas 55 mmHg with a calculated aortic valve area of 0.8 cm2.Due to high surgical risk scores (Logistic EuroSCORE=31.21%), the patient was scheduled for TAVI via transfemoralroute. An Amplatz Super Stiff PTFE-coated Guide wire (7 cmBentson-Type) was positioned correctly into the left ventricle.Predilatation was made into the aortic root by balloon.A 23 mm Edwards SAPIEN XT valve was advanced to theaorta, however, the valve was unable to be placed at the aorticvalve level due to the retraction of the guide wire from theleft ventricle. The transcatheter aortic valve was withdrawnand implanted to the descending aorta. Subsequently, anothertranscatheter aortic valve (23 mm Edwards SAPIEN XT) wasimplanted at the aortic valve level successfully.

🔖

Bu Makaleye Atıf Yap

GÜL, M., USLU, N., ÖZTÜRK, D., ASLAN, S., ÇELİK, Ö. (2022). Transcatheter valve implantation into descending aorta due to the insufficient guide wire support. Türk Göğüs Kalp Damar Cerrahisi Dergisi, Cilt: 23 - Sayı: 2, 363–365. https://ajindex.com/articles/624576

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (4)

  1. Holmes DR Jr, Mack MJ, Kaul S, Agnihotri A, Alexander KP, Bailey SR, et al. 2012 ACCF/AATS/SCAI/STS expert consensus document on transcatheter aortic valve replacement. J Am Coll Cardiol 2012;59:1200-54
  2. Madhwal S, Rajagopal V, Bhatt DL, Bajzer CT, Whitlow P, Kapadia SR. Predictors of difficult carotid stenting as determined by aortic arch angiography. J Invasive Cardiol 2008;20:200-4
  3. Généreux P, Head SJ, Van Mieghem NM, Kodali S, Kirtane AJ, Xu K, et al. Clinical outcomes after transcatheter aortic valve replacement using valve academic research consortium definitions: a weighted meta-analysis of 3,519 patients from 16 studies. J Am Coll Cardiol 2012;59:2317-26
  4. Safari Pre-Shaped TAVI Guidewire. Available from: http:// www.bostonscientific.com/en US/products/guidewires/ safari.html. [Accessed: September 17, 2014]