Clinical accuracy of non-contact infrared thermometer from umbilical region in children: A new side

Sayfa
180–186
Yayın Tarihi
2022-11-05
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

The easy application may lead noncontact measurements from umbilicus site to be the preferable method for health care providers, but agreement limits mentioned in this study should be considered.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Apa H, Gözmen S, Keskin-Gözmen Ş, Aslan F, Bayram N, Devrim İ.
📊
Temel Bulgular & Veri:Turk J Pediatr 2016; 58: 180-186.Measurement from axillary site with digital thermometer has been accepted as the most accurate method.
💡
Sonuç & Katkı:The easy application may lead noncontact measurements from umbilicus site to be the preferable method for health care providers, but agreement limits mentioned in this study should be considered.
🔬Tür:Araştırma Makalesi
📍Kapsam:2016 Verileri
⚙️Değişkenler:Alan Parametreleri

Abstract

Apa H, Gözmen S, Keskin-Gözmen Ş, Aslan F, Bayram N, Devrim İ. Clinical accuracy of non-contact infrared thermometer from umbilical region in children: A new side. Turk J Pediatr 2016; 58: 180-186.Measurement from axillary site with digital thermometer has been accepted as the most accurate method. But this method is time consuming. Tympanic and forehead measurements are often used but don't always seem to be more appropriate. Another site, umbilical region, could be an alternative site. This study aims to compare the measurements with axillary digital thermometer and non-contact infrared thermometers at sites from umbilicus and forehead to determine whether umbilical site could be used accurately in children. For each method, 2,048 measurements in total were performed. Using axillary method as gold standard, with a cut-off temperature of 38oC, the sensitivities and specificities, positive and negative predictive values of umbilical and forehead temperatures and area under the ROC curve were determined in non obese children. There was a significant positive correlation between axillary and umbilical temperatures with a correlation coefficient of 0.78. The average difference between the mean of both axillary and umbilical temperatures was -0.47 ± 0.65°C. The Bland-Altman plot showed good accuracy with only 2.5 % of the readings falling outside the 95% level of confidence. Umbilical measurements showed sensitivity of 71.7% and specificity of 95.8%. The area under the ROC curve was 0.93. The easy application may lead noncontact measurements from umbilicus site to be the preferable method for health care providers, but agreement limits mentioned in this study should be considered.

🔖

Bu Makaleye Atıf Yap

BAYRAM, N., APA, H., ASLAN, F., GÖZMEN, S., KESKİN-GÖZMEN, Ş., DEVRİM, İ. (2022). Clinical accuracy of non-contact infrared thermometer from umbilical region in children: A new side. Turkish Journal of Pediatrics, Cilt: 58 - Sayı: 2, 180–186. https://ajindex.com/articles/613194

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (15)

  1. Teran CG, Torrez-Llanos J, Teran-Miranda TE, et al. Clinical accuracy of a non-contact infrared skin thermometer in paediatric practice. Child Care Health Dev 2012; 38: 471-476
  2. Kara A, Devrim I, Cengiz AB, et al. Is the axilla the right site for temperature measurement in children by chemical thermometer? Turk J Pediatr 2009; 51: 325-327
  3. Mackowiak PA, Wasserman SS, Levine MM. A critical appraisal of 98.6 degrees F, the upper limit of the normal body temperature, and other legacies of Carl Reinhold August Wunderlich. JAMA 1992; 268: 1578- 1580
  4. Pocock G, Richards CD. Human Physiology The Basis of Medicine. New York, NY: Oxford University Press; 1999: 539-546
  5. Lorin M. Measurement of body temperature. Semin Pediatr Infect Dis 1993; 4: 4-8
  6. Fulbrook P. Core temperature measurements in adults: a literature review. J Adv Nurs 1993; 18: 1451-1460
  7. Apa H, Gozmen S, Bayram N, et al. Clinical accuracy of tympanic thermometer and noncontact infrared skin thermometer in pediatric practice. An alternative for axillary digital thermometer. Pediatr Emerg Care 2013; 29: 992-997
  8. Fischler MP, Reinhart WH. Fever: friend or enemy? Schweiz Med Wochenschr 1997; 127: 864-870
  9. Baraff LJ, Bass JW, Fleisher GR. Practice guideline for the management of infants and children 0 to 36 months of age with fever without source. Ann Emerg Med 1993; 22: 1198-1210
  10. Morley CJ, Hewson PH, Thornton AJ, et al. Axillary and rectal temperature measurements in infants. Arch Dis Child 1992; 67: 122-125
  11. Jensen BN, Jensen FS, Madsen SN, et al. Accuracy of digital tympanic, oral, axillary, and rectal thermometers compared with standard rectal mercury thermometers. Eur J Surg 2000; 166: 848-851
  12. Caravati EM, Erdman AR, Christianson G, et al. Elemental mercury exposure: an evidence-based consensus guideline for out-of-hospital management. Clin Toxicol (Philia) 2008; 46: 1-21
  13. Schmitz T, Bair N, Falk M, et al. A comparison of five methods of temperature measurements in febrile intensive care patients. Am J Crit Care 1995; 4: 286 -292
  14. Chiappini E, Sollai S, Longhi R, et al. Performance of non-contact infrared thermometer for detecting febrile children in hospital and ambulatory settings. J Clin Nurs 2011; 20: 1311-1318
  15. Neyzi O, Günöz H, Furman A, Bundak R, Gökçay G, Darendeliler F, Baş F. Türk çocuklarında vücut ağırlığı, boy uzunluğu, baş çevresi ve vücut kitle indeksi referans değerleri. Çocuk Sağlığı ve Hastalıkları Dergisi 2008; 51: 1-14 16. Blan