Dergiler / Joint diseases and related surgery / 2020 / Cilt: 31 - Sayı: 2

A novel method to assess intraarticular screw penetration into the joint space

Sayfa
218–222
Yayın Tarihi
2022-10-19
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

Penetration was confirmed by dissection of the joint.Results: There was a significant decrease in electrical resistance when screw penetration occurred (p

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Objectives: This study aims to propose a novel method to detect articular penetration of screws by relying on their electrical conductivity properties and control the validity of this method.Materials and methods: In this ex vivo study, conducted between June 2017 and August 2017, we used five fresh sheep shoulder joints.
📊
Temel Bulgular & Veri:When a sudden decrease in conduction resistance was detected, fluoroscopic images were obtained in the anteroposterior (AP) and lateral directions.
💡
Sonuç & Katkı:Penetration was confirmed by dissection of the joint.Results: There was a significant decrease in electrical resistance when screw penetration occurred (p
🔬Tür:Araştırma Makalesi
📍Kapsam:2017 Verileri
⚙️Değişkenler:Alan Parametreleri

Abstract

Objectives: This study aims to propose a novel method to detect articular penetration of screws by relying on their electrical conductivity properties and control the validity of this method.Materials and methods: In this ex vivo study, conducted between June 2017 and August 2017, we used five fresh sheep shoulder joints. First, the shoulder joint space was filled with saline solution. An insulated cannula was placed in the joint capsule, and a conductive wire was introduced into the joint via this cannula. A single titanium screw was inserted from the tuberculum majus into the posteroinferior quadrant of the humeral head under fluoroscopic observation. Conductivity was continuously measured using a digital multimeter. When a sudden decrease in conduction resistance was detected, fluoroscopic images were obtained in the anteroposterior (AP) and lateral directions. These images were assessed for penetration by a blinded surgeon. Penetration was confirmed by dissection of the joint.Results: There was a significant decrease in electrical resistance when screw penetration occurred (p

🔖

Bu Makaleye Atıf Yap

CEYLAN, H., ERDEN, T., KAPICIOĞLU, M., KÜÇÜKDURMAZ, F. (2022). A novel method to assess intraarticular screw penetration into the joint space. Joint diseases and related surgery, Cilt: 31 - Sayı: 2, 218–222. https://ajindex.com/articles/559287

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (14)

  1. Aksu N, Göğüş A, Kara AN, Işiklar ZU. Complications encountered in proximal humerus fractures treated with locking plate fixation. Acta Orthop Traumatol Turc 2010;44:89-96
  2. Egol KA, Sugi MT, Ong CC, Montero N, Davidovitch R, Zuckerman JD. Fracture site augmentation with calcium phosphate cement reduces screw penetration after open reduction-internal fixation of proximal humeral fractures. J Shoulder Elbow Surg 2012;21:741-8
  3. Hepp P, Theopold J, Voigt C, Engel T, Josten C, Lill H. The surgical approach for locking plate osteosynthesis of displaced proximal humeral fractures influences the functional outcome. J Shoulder Elbow Surg 2008;17:21-8
  4. Owsley KC, Gorczyca JT. Fracture displacement and screw cutout after open reduction and locked plate fixation of proximal humeral fractures [corrected]. J Bone Joint Surg [Am] 2008;90:233-40
  5. Wang Q, Zhu Y, Liu Y, Wang L, Chen Y. Correlation between classification and secondary screw penetration in proximal humeral fractures. PLoS One 2017;12:e0183164
  6. Lehman WB, Grant A, Rose D, Pugh J, Norman A. A method of evaluating possible pin penetration in slipped capital femoral epiphysis using a cannulated internal fixation device. Clin Orthop Relat Res 1984;186:65-70
  7. Noordeen MH, Lavy CB, Briggs TW, Roos MF. Unrecognised joint penetration in treatment of femoral neck fractures. J Bone Joint Surg [Br] 1993;75:448-9
  8. Anglen JO, DiPasquale T. The reliability of detecting screw penetration of the acetabulum by intraoperative auscultation. J Orthop Trauma 1994;8:404-8
  9. Spross C, Jost B, Rahm S, Winklhofer S, Erhardt J, Benninger E. How many radiographs are needed to detect angular stable head screw cut outs of the proximal humerus - a cadaver study. Injury 2014;45:1557-63
  10. Atik OŞ. Every new technique either conservative or surgical is good? Eklem Hastalik Cerrahisi 2019;30:183-4
  11. Hernigou P, Besnard P. Articular penetration is more likely in Garden-I fractures of the hip. J Bone Joint Surg [Br] 1997;79:285 - 8
  12. Wera JC, Seligson D, Riehl JT. Medial malleolus screws: out in one view and out. Eur J Orthop Surg Traumatol 2015;25:1189-93
  13. Weil YA, Beyth S, Abu-Ahmad A, Khoury A, Liebergall M, Safran O. Use of three-dimensional fluoroscopy to determine intra-articular screw penetration in proximal humeral fracture model. J Shoulder Elbow Surg 2014;23:1150-5
  14. Hernigou P, Germany W. Unrecognized shoulder joint penetration during fixation of proximal fractures of the humerus. Acta Orthop