Dergiler / İstanbul Medical Journal / 2019 / Cilt: 20 - Sayı: 3

F-18 FDG PET/BT Görüntüleme Yönteminin Tetkik Sürecinde ve Sonrasında Çevresel Radyasyon Güvenliğinin Değerlendirilmesi

The Evaluation of External Dose Rate Measurements of Patients During and After F-18 FDG PET/CT Imaging and Appropriate Discharge Time from PET/CT Department

Sayfa
188–192
Yayın Tarihi
2022-10-19
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

Bulgular: Hastalara IV yol ile 291±62 megabecquerel (MBq) (Aralık: 226-440 MBq) radyoaktif madde enjekte edilmiş ve nükleer tıp kliniğinde enjeksiyon sonrası 70,3±13,8 dk (Aralık: 53 dk-100 dk) beklemiştir.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Amaç: Nükleer tıpta pozitron emisyon tomografi/bilgisayarlı tomografi (PET/BT) görüntüleme yönteminde kullanılan F-18 radyoizotopu konvansiyonel nükleer tıpta tanı amacıyla kullanılan diğer radyoizotoplara göre daha yüksek foton enerjisine sahiptir.
📊
Temel Bulgular & Veri:Çalışmamızda rutin uygulamada yaygın olarak kullanılan F-18 fluorodeoksiglukoz (FDG), PET/BT çekimi için kullanılan F-18 FDG radyoizotopunun enjeksiyonundan sonra hastadan çevreye salınan radyasyonun doz hızının ölçülmesi ve çevresel radyasyon maruziyetinin değerlendirilmesi amaçlanmıştır.
💡
Sonuç & Katkı:Bulgular: Hastalara IV yol ile 291±62 megabecquerel (MBq) (Aralık: 226-440 MBq) radyoaktif madde enjekte edilmiş ve nükleer tıp kliniğinde enjeksiyon sonrası 70,3±13,8 dk (Aralık: 53 dk-100 dk) beklemiştir.
🔬Tür:Deneysel Tarla Denemesi
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Alan Parametreleri

Özet

Amaç: Nükleer tıpta pozitron emisyon tomografi/bilgisayarlı tomografi (PET/BT) görüntüleme yönteminde kullanılan F-18 radyoizotopu konvansiyonel nükleer tıpta tanı amacıyla kullanılan diğer radyoizotoplara göre daha yüksek foton enerjisine sahiptir. Çalışmamızda rutin uygulamada yaygın olarak kullanılan F-18 fluorodeoksiglukoz (FDG), PET/BT çekimi için kullanılan F-18 FDG radyoizotopunun enjeksiyonundan sonra hastadan çevreye salınan radyasyonun doz hızının ölçülmesi ve çevresel radyasyon maruziyetinin değerlendirilmesi amaçlanmıştır. Yöntemler: Çalışmamıza, Nükleer Tıp PET/BT ünitesine başvuran toplam 29 (E/K=9/20) hasta dahil edilmiştir. Hastalara intravenöz (IV) olarak F-18 FDG enjekte edildikten hemen sonra, çekim öncesi ve taburcu edilmeden önce 0 m, 0,5 m, 1 m, 1,5 m ve 2 m mesafeden toraks düzeyinden Geiger-Müller sayacı kullanılarak doz hızı ölçümleri yapılmıştır. Her bir ölçümün ortalama ve standart sapma değerleri hesaplanmış, ayrıca ölçümler arasında istatistiksel olarak anlamlı farklılık olup olmadığı değerlendirilmiştir. Bulgular: Hastalara IV yol ile 291±62 megabecquerel (MBq) (Aralık: 226-440 MBq) radyoaktif madde enjekte edilmiş ve nükleer tıp kliniğinde enjeksiyon sonrası 70,3±13,8 dk (Aralık: 53 dk-100 dk) beklemiştir. Bir m mesafeden alınan radyasyon doz hızı ölçümleri, radyoaktif madde enjeksiyonu sonrası 39,7±11,9 μSv/sa, çekim esnasında 21,1±4,5 μSv/sa, taburcu anında 14,3±2,9 μSv/sa olarak hesaplanmıştır. Sonuç: F-18’in foton enerjisi 511 kilo elektron volt olmasına karşın nispeten kısa yarı ömürlü olmasından dolayı F-18 FDG tanı amacı ile kullanılabilen ideal bir radyofarmasötiktir. PET/ BT çekimi sonlandığında hastaların 1 m mesafedeki doz hızı ölçümlerinin, Türkiye Atom Enerjisi Kurumu mevzuatları da göz önünde bulundurulduğunda, çevresel radyasyon güvenliği açısından bir tehlike arz etmediği gösterilmiştir. Bu nedenle hastalar F-18 FDG PET/BT görüntülemesi tamamlandıktan sonra gerekli radyasyon güvenlik önlemleri anlatılarak güvenli bir şekilde taburcu edilebilmektedir.

Abstract

Introduction: F-18 radioisotope has higher photon energywhen compared with other radioisotopes that are preferredin conventional imaging procedures in nuclear medicine. Theaim of this study was to measure external radiation dose ratefrom patients injected with F-18 fluorodeoxyglucose (FDG)and to determine appropriate discharge time after F-18 FDGpositron emission tomography/computed tomography (PET/CT)procedure.Methods: Our study included 29 (M/F=9/20) patients who werereferred to nuclear medicine department for F-18 FDG PET/CTimaging. Dose rates were measured at a distance of 0, 0.5, 1,1.5 and 2 m from patients via Geiger-Müller probe at 3 differenttime points: 1) Just after the intravenous injection of F-18 FDG,2) before PET/CT imaging and 3) just before discharge of patientsfrom PET/CT department. Statistical analysis was performed toevaluate the differences between measurements.Results: Patients were injected 291±62 megabecquerel (MBq)(range: 226-440 MBq) F-18 FDG via IV route. The patients stayedin nuclear medicine department for 70.3±13.8 min (range:53 min-100 min) after injection of F-18 FDG. Radiation dosevelocity measurements taken at a distance of 1 meter were39.7±11.9 μSv/h just after the injection, 21.1±4.5 μSv/h beforePET/CT examination and 14.3±2.9 μSv/h just before discharge.Conclusion: Although F-18 has high photon energy, its halflifeis relatively short and this makes F-18 FDG an idealradiopharmaceutical for PET/CT imaging. When the PET/CT scanis terminated, the dose rate measurements of the patients ata distance of 1 m are appropriate for discharge according toTurkish Atomic Energy Authority regulations. Therefore, patientscan be safely discharged by explaining the necessary radiationsafety measures after completing F-18 FDG PET/CT imaging.

🔖

Bu Makaleye Atıf Yap

GÜNDOĞAN, C., ERGÜL, N., GÜNEŞ, B. Y., EREZ, Ö. (2022). F-18 FDG PET/BT Görüntüleme Yönteminin Tetkik Sürecinde ve Sonrasında Çevresel Radyasyon Güvenliğinin Değerlendirilmesi. İstanbul Medical Journal, Cilt: 20 - Sayı: 3, 188–192. https://ajindex.com/articles/551947

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (12)

  1. Gunes BY, Onsel C, Sonmezoglu K, Ozaras R, Halac M, Tabak F, et al. Diagnostic value of F-18 FDG PET/CT in patients with spondylodiscitis: Is dual time point imaging time worthy? Diagn Microbiol Infect Dis 2016; 85: 381-5
  2. Ergul N, Kadioglu H, Yildiz S, Yucel SB, Gucin Z, Erdogan EB, et al. Assessment of multifocality and axillary nodal involvement in early-stage breast cancer patients using 18F-FDG PET/CT compared to contrast-enhanced and diffusionweighted magnetic resonance imaging and sentinel node biopsy. Acta Radiol 2015; 56: 917-23
  3. Ergül N, Cermik TF. FDG-PET or PET/CT in Fever of unknown origin: The diagnostic role of underlying primary disease. Int J Mol Imaging 2011; 2011: 318051
  4. Akaike G, Itani M, Shah H, Ahuja J, Yilmaz Gunes B, Assaker R, et al. PET/CT in the diagnosis and workup of sarcoidosis: Focus on atypical manifestations. Radiographics 2018; 38: 1536-49
  5. Boellaard R, Delgado-Bolton R, Oyen WJ, Giammarile F, Tatsch K, Eschner W, et al. FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0. Eur J Nucl Med Mol Imaging 2015; 42: 328-54
  6. Atci IB, Yilmaz H, Antar V, Ozdemir NG, Baran O, Sadillioglu S, et al. What do we know about ALARA? Is our knowledge sufficient about radiation safety? J Neurosurg Sci 2017; 61: 597-602
  7. Siegel JA, Sacks B, Pennington CW, Welsh JS. Dose optimization to minimize radiation risk for children undergoing CT and nuclear medicine imaging is misguided and detrimental. J Nucl Med 2017; 58: 865-8
  8. Chen MK, Menard DH 3rd, Cheng DW. Determining the minimal required radioactivity of 18F-FDG for reliable semiquantification in PET/CT imaging: A Phantom Study. J Nucl Med Technol 2016; 44: 26-30
  9. Aras F, Kıraç S, Ayan A, Değer M, Demir M, Dönmez S, et al. Radiation safety process for the patient and people around during diagnostic radionuclide studies. Nucl Med Semin 2016; 3: 172-175
  10. Türkiye Atom Enerjisi Kurumu Kanunu. Kanun No: 2690. R.G: 13.7.1982/17753
  11. TAEK (Türkiye Atom Enerjisi Kurumu), Radyasyon Güvenliği Yönetmeliği Ek-1 12. TAEK (Türkiye Atom Enerjisi Kurumu), Radyasyon Güvenliği Yönetmeliği Ek-4 13. TAEK (Türkiye Atom Enerjisi Kurumu), Radyasyon Güvenliği Yönetmeliği Ek-6 14. Yeyin N, Abuqbeitah M, Çavdar İ, Sönmezoğlu K, Demir M. The evaluation of dose rate measurements and discharge time after radionuclide therapy. J Ist Faculty Med 2017; 80: 1
  12. TAEK (Türkiye Atom Enerjisi Kurumu), Radyasyon Güvenliği Yönetmeliği Ek-5 16. European Commission Radiation Prot