Comparison of Intraocular Pressure and Pachymeter Measurements Obtained by Non Contact Tonometerpachymeter with Goldmann Applanation Tonometer and Corneal Topography

Sayfa
12–17
Yayın Tarihi
2022-10-18
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

Mean differences between devices were signifi cantly different in terms of IOP and CCT measurement (p

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Purpose: To compare the measurements of intraocular pressure (IOP) and central corneal thickness (CCT) by tono-pachymeter (CT-1P) with IOP measurements by Goldmann applanation tonometer (GAT), and CCT measurements by Pentacam.
📊
Temel Bulgular & Veri:Materials and Methods: Two hundred eighteen eyes of 109 patients were included in the study.
💡
Sonuç & Katkı:Mean differences between devices were signifi cantly different in terms of IOP and CCT measurement (p
🔬Tür:Araştırma Makalesi
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Alan Parametreleri

Abstract

Purpose: To compare the measurements of intraocular pressure (IOP) and central corneal thickness (CCT) by tono-pachymeter (CT-1P) with IOP measurements by Goldmann applanation tonometer (GAT), and CCT measurements by Pentacam. Materials and Methods: Two hundred eighteen eyes of 109 patients were included in the study. Intraocular pressure measurements were obtained by the CT-1P and GAT, whereas CCT measurements were obtained by CT-1P and Pentacam. Blant-Altman plots for agreement, and Pearson correlation and interclass correlation (ICC) for reliability between the measurements were analyzed. Results: The mean IOPCT-1P, IOPCT-1P adj., IOPGAT were measured as 18.2±4.32 mmHg, 17.75±4.05 mmHg and 16.9±4.16 mmHg, respectively. The mean CCT was measured as 554.8±37.3 μm and 543.6±35.9 μm with CT-1P and Pentacam, respectively. Mean differences between devices were signifi cantly different in terms of IOP and CCT measurement (p

🔖

Bu Makaleye Atıf Yap

MUTLU, F. M., KOYLU, M. T., AKINCIOGLU, D., YETER, D. Y., OGUZ, Y. G., YESILTAS, Y. S., GURDAL, F. (2022). Comparison of Intraocular Pressure and Pachymeter Measurements Obtained by Non Contact Tonometerpachymeter with Goldmann Applanation Tonometer and Corneal Topography. Glokom Katarakt, Cilt: 16 - Sayı: 1, 12–17. https://ajindex.com/articles/525747