Dergiler / Clinical and Experimental Health Sciences / 2021 / Cilt: 11 - Sayı: 2

The Effect of Various Canal Contents on the Accuracy of Two Electronic Apex Locators in Detecting Different Size of Root Perforations

Sayfa
258–262
Yayın Tarihi
2021-06-30
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

Conclusion: Within the limitation of this study, intracanal conditions did not affect the measurements of DentaPort ZX in both perforation sizes, whereas in 0.5 mm perforation size, GRM measurements were affected by intracanal conditions..

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Objective: This study was to investigate the effect of different canal contents on the accuracy of Gold Reciproc Motor (GRM; VDW, Munich, Germany) and DentaPort ZX (Morita Co, Kyoto, Japan) in the determination of artificial root perforations.
📊
Temel Bulgular & Veri:The crowns of the teeth were removed and the root lengths were standardized to be 14±1 mm.
💡
Sonuç & Katkı:Conclusion: Within the limitation of this study, intracanal conditions did not affect the measurements of DentaPort ZX in both perforation sizes, whereas in 0.5 mm perforation size, GRM measurements were affected by intracanal conditions..
🔬Tür:Araştırma Makalesi
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Blood · DentaPort ZX · GRM

Abstract

Objective: This study was to investigate the effect of different canal contents on the accuracy of Gold Reciproc Motor (GRM; VDW, Munich, Germany) and DentaPort ZX (Morita Co, Kyoto, Japan) in the determination of artificial root perforations. Methods: Forty mandibular premolar teeth were included in this study. The crowns of the teeth were removed and the root lengths were standardized to be 14±1 mm. Roots were divided into 2 groups (n=20). Artificial root perforations were created 0.5±0.1mm and 1±0.1 mm in size respectively. The actual lengths (AL) up to the perforation areas were measured under the stereomicroscope. Electronic measurements (EL) were obtained by GRM and DentaPort ZX in dry conditions and the presence of NaOCl (5.25%), EDTA (17%) and blood. The difference was calculated by subtracting the ALs from the ELs. This difference was positive when the measurement was longer than the AL and negative when the measurement was shorter. The Friedman and Wilcoxon signed-rank tests were used to analyze the data (p˂0.05). Results: There were no significant differences amongst the different intracanal conditions for both apex locators in teeth with perforation of 1 mm. In the teeth with a perforation of 0.5 mm DentaPort ZX measurements were not affected by intracanal conditions and the most accurate measurement was obtained when the canal was dry with GRM. Conclusion: Within the limitation of this study, intracanal conditions did not affect the measurements of DentaPort ZX in both perforation sizes, whereas in 0.5 mm perforation size, GRM measurements were affected by intracanal conditions..

🔖

Bu Makaleye Atıf Yap

DOĞAN, T., AYDIN, Z. U., ALTUNBAŞ, D. (2021). The Effect of Various Canal Contents on the Accuracy of Two Electronic Apex Locators in Detecting Different Size of Root Perforations. Clinical and Experimental Health Sciences, Cilt: 11 - Sayı: 2, 258–262. https://ajindex.com/articles/481289

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (13)

  1. Nazari Moghaddam K, Nazari S, Shakeri L, Honardar K, Mirmotalebi F. In vitro detection of simulated apical root perforation with two electronic apex locators. Iran Endod J 2010;5 (1):23-6
  2. Marroquin BB, Fernandez CC, Schmidtmann I, Willershausen B, Goldberg F. Accuracy of electronic apex locators to detect root canal perforations with inserted metallic posts: an ex vivo study. Head Face Med 2014;10 (1):57
  3. Kaufman AY, Fuss Z, Keila S, Waxenberg S. Reliability of different electronic apex locators to detect root perforations in vitro. Int Endod J 1997;30 (6):403-7
  4. Siew K, Lee AH, Cheung GS. Treatment Outcome of Repaired Root Perforation: A Systematic Review and Meta-analysis. J Endod 2015;41 (11):1795-804
  5. Altunbas D, Kustarci A, Toyoglu M. The Influence of Various Irrigants on the Accuracy of 2 Electronic Apex Locators in Locating Simulated Root Perforations. J Endod 2017;43 (3):439-42
  6. D'Assuncao FL, Sousa JC, Felinto KC, de Medeiros TC, Leite DT, de Lucena RB et al. Accuracy and repeatability of 3 apex locators in locating root canal perforations: an ex vivo study. J Endod 2014;40 (8):1241-4
  7. Ustun Y, Aslan T, Sekerci AE, Sagsen B. Evaluation of the Reliability of Cone-beam Computed Tomography Scanning and Electronic Apex Locator Measurements in Working Length Determination of Teeth with Large Periapical Lesions. J Endod 2016;42 (9):1334-7
  8. Angwaravong O, Panitvisai P. Accuracy of an electronic apex locator in primary teeth with root resorption. Int Endod J 2009;42 (2):115-21
  9. Duran-Sindreu F, Gomes S, Stober E, Mercade M, Jane L, Roig M. In vivo evaluation of the iPex and Root ZX electronic apex locators using various irrigants. Int Endod J 2013;46(8):769-74
  10. Duran-Sindreu F, Stöber E, Mercadé M, Vera J, Garcia M, Bueno R et al. Comparison of in vivo and in vitro readings when testing the accuracy of the Root ZX apex locator. J Endod 2012;38 (2):236-9
  11. Altunbas D, Kustarci A, Arslan D, Er K, Kocak S. Comparison of various current electronic apex locators to determine the working length using the clearing technique. Niger J Clin Pract 2015;18 (3):359-63
  12. Erdemir A, Eldeniz AU, Ari H, Belli S, Esener T. The influence of irrigating solutions on the accuracy of the electronic apex locator facility in the Tri Auto ZX handpiece. Int Endod J 2007;40 (5):391-7
  13. Goldberg F, De Silvio AC, Manfre S, Nastri N. In vitro measurement accuracy of an electronic apex locator in teeth with simulated apical root resorption. J End