Effects of Pulsatile Flow on Phosphorylcholine Coated Oxygenator and Arterial Filter

Sayfa
793–799
Yayın Tarihi
2022-03-31
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

Coating of oxygenators is beneficial in case of surface biocompatibility and pulsatile perfusion develops lower amount of blood elements on arterial filter.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Objective: We aimed to compare the effects of pulsatile/nonpulsatile flow on phosphorylcholine coated (PC) oxygenator fibers, arterial filters by using scanning electron microscope (SEM).
📊
Temel Bulgular & Veri:Results: The mean fiber thickness from the axial images were calculated as 46.9m and 47.6 m at group P and NP respectively which is statistically insignificant.
💡
Sonuç & Katkı:Coating of oxygenators is beneficial in case of surface biocompatibility and pulsatile perfusion develops lower amount of blood elements on arterial filter.
🔬Tür:Araştırma Makalesi
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Pulsatil flow · Coated oxygenator · Arterial filter

Özet

Objective: We aimed to compare the effects of pulsatile/nonpulsatile flow on phosphorylcholine coated (PC) oxygenator fibers, arterial filters by using scanning electron microscope (SEM). Methods: Eleven patients were randomly divided into two groups, as; nonpulsatile and pulsatile flow groups (NP and P groups) by using PC oxygenators. The oxygenator fiber samples were examined under SEM to compare the thickness of absorbed blood proteins and amount of blood cells on the surface of oxygenators. Arterial filters were also analysed by SEM regarding the captured blood elements or particles. Results: The mean fiber thickness from the axial images were calculated as 46.9m and 47.6 m at group P and NP respectively which is statistically insignificant. Evaluation of the blood samples that were extracted from the arterial filter bring out higher amount of fibrin network and blood cells on fibers at group NP. Conclusion: We concluded that there is lesser amount of blood components on the fibers of arterial filter at pulsatile flow. Coating of oxygenators is beneficial in case of surface biocompatibility and pulsatile perfusion develops lower amount of blood elements on arterial filter.

Abstract

Objective: We aimed to compare the effects of pulsatile/nonpulsatile flow on phosphorylcholine coated (PC) oxygenator fibers, arterial filters by using scanning electron microscope (SEM). Methods: Eleven patients were randomly divided into two groups, as; nonpulsatile and pulsatile flow groups (NP and P groups) by using PC oxygenators. The oxygenator fiber samples were examined under SEM to compare the thickness of absorbed blood proteins and amount of blood cells on the surface of oxygenators. Arterial filters were also analysed by SEM regarding the captured blood elements or particles. Results: The mean fiber thickness from the axial images were calculated as 46.9 m and 47.6 m at group P and NP respectively which is statistically insignificant. Evaluation of the blood samples that were extracted from the arterial filter bring out higher amount of fibrin network and blood cells on fibers at group NP. Conclusion: We concluded that there is lesser amount of blood components on the fibers of arterial filter at pulsatile flow. Coating of oxygenators is beneficial in case of surface biocompatibility and pulsatile perfusion develops lower amount of blood elements on arterial filter.

🔖

Bu Makaleye Atıf Yap

KESKİN, G., ULUS, A. T., GÜRAY, T., ÜRPERMEZ, E., ÖZYALÇIN, S., HABERAL, O. E., KOCAKULAK, M. (2022). Effects of Pulsatile Flow on Phosphorylcholine Coated Oxygenator and Arterial Filter. Avrupa Bilim ve Teknoloji Dergisi, Sayı: 34, 793–799. https://ajindex.com/articles/470019

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (10)

  1. Abramov, D., Tamariz, M., Serrick, C. I., Sharp, E., Noel, D., Harwood, S.,. & Goldman, B. S. (2003). The influence of cardiopulmonary bypass flow characteristics on the clinical outcome of 1820 coronary bypass patients. The Canadian journal of cardiology, 19(3), 237-243. Aĝirbaşli, M. A., Song, J., Lei, F., Wang, S., Kunselman, A. R., Clark, J. B.,. & Ündar, A. (2014). Comparative effects of pulsatile and nonpulsatile flow on plasma fibrinolytic balance in pediatric patients undergoing cardiopulmonary bypass. Artificial organs, 38(1), 28-33.
  2. Alghamdi, A. A., & Latter, D. A. (2006). Pulsatile versus nonpulsatile cardiopulmonary bypass flow: an evidence‐based approach. Journal of cardiac surgery, 21(4), 347-354.
  3. Chiu, I. S., Chu, S. H., & Hung, C. R. (1984). Pulsatile flow during routine cardiopulmonary bypass. The Journal of Cardiovascular Surgery, 25(6), 530-536.
  4. De Leval, M., Hill, J. D., Mielke, H., Bramson, M. L., Smith, C., & Gerbode, F. (1972). Platelet kinetics during extracorporeal circulation. ASAIO Journal, 18(1), 355-357.
  5. De Somer, F., Van Belleghem, Y., Foubert, L., François, K., Dubrulle, F., De Wolf, D., & Van Nooten, G. (1960). In vivo evaluation of a phosphorylcholine coated cardiopulmonary bypass. Strategies for optimisation of paediatric cardiopulmonary bypass, 31(2), 241. Dunn, J., Kirsh, M. M., Harness, J., Carroll, M., Straker, J., & Sloan, H. (1974). Hemodynamic, metabolic, and hematologic effects of pulsatile cardiopulmonary bypass. The Journal of Thoracic and Cardiovascular Surgery, 68(1), 138-147.
  6. Frering, B., Philip, I., Dehoux, M., Rolland, C., Langlois, J. M., & Desmonts, J. M. (1994). Circulating cytokines in patients undergoing normothermic cardiopulmonary bypass. The Journal of thoracic and cardiovascular surgery, 108(4), 636-641.
  7. Gorbet, M. B., & Sefton, M. V. (2004). Biomaterial-associated thrombosis: roles of coagulation factors, complement, platelets and leukocytes. Biomaterials, 25(26), 5681-5703.
  8. Gourlay, T. (2001). Biomaterial development for cardiopulmonary bypass. Perfusion, 16(5), 381-390.
  9. Guan, Y., Palanzo, D., Kunselman, A., & Ündar, A. (2009). Evaluation of membrane oxygenators and reservoirs in terms of capturing gaseous microemboli and pressure drops. Artificial organs, 33(11), 1037-1043.
  10. Iwasaki, Y., Tojo, Y., Kurosaki, T., & Nakabayashi, N. (2003). Reduced adhesion of blood cells to biodegradable polymers by introducing phosphorylcholine moieties. Journal of Biomedical Materials Research Part A: An Offi