Erroneously high troponin measurement caused by fibrin clot: Two cases

Sayfa
208–210
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

The laboratory staff were trained on preanalytical errors, and the use of lithium heparin tubes was implemented in the laboratory as an additional means to eliminate the problem of fibrin interference.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Acute myocardial infarction is the most common cause of morbidity and mortality in the world.
📊
Temel Bulgular & Veri:In our hospital, high-sensitivity tro ponin-I (hs-TnI) was routinely analyzed using a serum separator tube (reference value: female <15.5 pg/mL, male <34.2 pg/mL).
💡
Sonuç & Katkı:The laboratory staff were trained on preanalytical errors, and the use of lithium heparin tubes was implemented in the laboratory as an additional means to eliminate the problem of fibrin interference.
🔬Tür:Araştırma Makalesi
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Alan Parametreleri

Özet

Acute myocardial infarction is the most common cause of morbidity and mortality in the world. Cardiac troponin mea surements play a key role in the diagnosis. However, preanalytical errors as a result of the presence of fibrin or inter ference due to conditions such as heterophile antibody positivity may cause erroneously high results. Such errors may result in invasive procedures, such as angiography, which may add unnecessary risk. In our hospital, high-sensitivity tro ponin-I (hs-TnI) was routinely analyzed using a serum separator tube (reference value: female <15.5 pg/mL, male <34.2 pg/mL). This report describes the cases of 2 patients with a false initial troponin measurement: a 19-year-old male pa tient and a 55-year-old female patient. The hs-TnI value of the male patient was initially measured as 55.5 pg/mL. After the analysis, it was noted that the sample contained fibrin. The sample was centrifuged again and the TnI result was 1.8 pg/mL. Similarly, the TnI result of the female patient was first measured as 90.2 pg/mL. When it was observed that there was fibrin present, the sample was recentrifuged. The revised result was 2.4 pg/mL. The laboratory staff were trained on preanalytical errors, and the use of lithium heparin tubes was implemented in the laboratory as an additional means to eliminate the problem of fibrin interference.

🔖

Bu Makaleye Atıf Yap

YAVUZ, H. B. (2021). Erroneously high troponin measurement caused by fibrin clot: Two cases. International Journal of Medical Biochemistry, Cilt: 4 - Sayı: 3, 208–210. https://ajindex.com/articles/413639

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (15)

  1. Thygesen K, Alpert JS, Jaffe AS, Simoons ML, Chaitman BR, White HD, et al; Joint ESC/ACCF/AHA/WHF Task Force for the Universal Definition of Myocardial Infarction. Third universal definition of myocardial infarction. Circulation 2012;126(16):2020–35
  2. Cullen L, Greenslade J, Merollini K, Graves N, Hammett CJ, Hawkins T, et al. Cost and outcomes of assessing patients with chest pain in an Australian emergency department. Med J Aust 2015;202(8):427–32
  3. Ward G, Simpson A, Boscato L, Hickman PE. The investigation of interferences in immunoassay. Clin Biochem 2017;50(18):1306–11
  4. Wu AHB. Cardiac markers. 2nd ed. New York: Springer Science & Business Media; 2003
  5. Xu RY, Zhu XF, Yang Y, Ye P. High-sensitive cardiac troponin. TJ Geriatr Cardiol 2013;10(1):102–9
  6. Christenson RH, Peacock WF, Apple FS, Limkakeng AT Jr, Nowak RM, McCord J, et al. Trial design for assessing analytical and clinical performance of high-sensitivity cardiac troponin I assays in the United States: The HIGH-US study. Contemp Clin Trials Commun 2019;14:100337
  7. Nosanchuk JS. False increases of troponin I attributed to incomplete separation of serum. Clin Chem 1999;45(5):714
  8. Gerhardt W, Nordin G, Herbert AK, Burzell BL, Isaksson A, Gustavsson E, et al. Troponin T and I assays show decreased concentrations in heparin plasma compared with serum: lower recoveries in early than in late phases of myocardial injury. Clin Chem 2000;46(6 Pt 1):817–21
  9. Dominici R, Infusino I, Valente C, Moraschinelli I, Franzini C. Plasma or serum samples: measurements of cardiac troponin T and of other analytes compared. Clin Chem Lab Med 2004;42(8):945–51
  10. Penttilä K, Koukkunen H, Halinen M, Punnonen K, Pyörälä K, Rantanen T, et al. Serum and plasma as alternative sample types in analysis of cardiac markers in the clinical routine. Scand J Clin Lab Invest 2002;62(7):553–60
  11. Hawkins RC. Laboratory turnaround time. Clin Biochem Rev 2007;28(4):179–94
  12. Bowen RA, Remaley AT. Interferences from blood collection tube components on clinical chemistry assays. Biochem Med (Zagreb) 2014;24(1):31–44
  13. Sharif W, Stretton A, Cornes M, Shetty C. Frequency of latent fibrin in centrifuged serum specimens. CCLM 2019;57(4):12– 4
  14. Franeková J, Bláha M, Bělohoubek J, Kotrbatá M, Sečník P Jr, Kubíček Z, et al. A clinical and laboratory approach used to elucidate discordant results of high-sensitivity troponin T and troponin I. Clin Chim Acta 2015;446:128–31
  15. Yaman SK, Yaman H, Yavuz HB, Kan