Evaluation of the Vision C erythrocyte sedimentation rate analyzer

Sayfa
172–177
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

Conclusion: The Vision C analyzer using a modified Westergren method met the precision and sample stability criteria, but the results were not fully compatible with the Westergren method at all ESR levels, and the difference was signifi cant at high ESR levels.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Objectives: The International Council for Standardization in Haematology recommends that new methods and instru ments used in medical laboratories be thoroughly tested with an analytical performance evaluation prior to imple mentation.
📊
Temel Bulgular & Veri:This study is an assessment of the analytical performance characteristics of the Vision C analyzer (Shenzhen YHLO Biotech Co., Ltd., Shenzhen, China), which uses a modified Westergren method to measure erythrocyte sedimen tation rate (ESR) and a comparison with the reference Westergren method.
💡
Sonuç & Katkı:Conclusion: The Vision C analyzer using a modified Westergren method met the precision and sample stability criteria, but the results were not fully compatible with the Westergren method at all ESR levels, and the difference was signifi cant at high ESR levels.
🔬Tür:Araştırma Makalesi
📍Kapsam:5 samples
⚙️Değişkenler:Alan Parametreleri

Özet

Objectives: The International Council for Standardization in Haematology recommends that new methods and instru ments used in medical laboratories be thoroughly tested with an analytical performance evaluation prior to imple mentation. This study is an assessment of the analytical performance characteristics of the Vision C analyzer (Shenzhen YHLO Biotech Co., Ltd., Shenzhen, China), which uses a modified Westergren method to measure erythrocyte sedimen tation rate (ESR) and a comparison with the reference Westergren method. Methods: The analytical performance of the Vision C was evaluated according to intra-run/inter-run precision, stabil ity, and compatibility with the gold standard Westergren method. In all, 173 patient samples were studied using both methods. The Westergren method levels were used to form 3 groups: 80 mm/hour. Results: The intra-run precision ranged from 4.93% to 18.18% for 5 samples at different levels. Inter-run precision for the normal level was 7.03% and 2.94% for the pathological level. Both the room temperature and refrigerated samples were stable at 4 hours. The overall bias between the Westergren method and the Vision C analyzer was -5.23 mm/hour (confidence interval [CI]: -6.66 to -3.79), the correlation coefficient was 0.948, and the Passing-Bablok regression equa tion was y=2.073+0.805x. The bias of the Vision C for ESR values 80 mm/hour. Conclusion: The Vision C analyzer using a modified Westergren method met the precision and sample stability criteria, but the results were not fully compatible with the Westergren method at all ESR levels, and the difference was signifi cant at high ESR levels.

🔖

Bu Makaleye Atıf Yap

ŞİRİN, F. B., İLHAN, İ., BÜYÜKBAYRAM, H. İ., DOĞUÇ, D. K. (2021). Evaluation of the Vision C erythrocyte sedimentation rate analyzer. International Journal of Medical Biochemistry, Cilt: 4 - Sayı: 3, 172–177. https://ajindex.com/articles/413636

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (14)

  1. Brigden ML. Clinical utility of the erythrocyte sedimentation rate. Am Fam Physician 1999;60(5):1443–50
  2. Aletaha D, Smolen JS. Diagnosis and management of rheumatoid arthritis: a review. JAMA 2018;320(13):1360–72
  3. Dasgupta B, Cimmino MA, Maradit-Kremers H, Schmidt WA, Schirmer M, Salvarani C, et al. Provisional criteria for polymyalgia rheumatica: a European league against rheumatism/ American college of rheumatology collaborative initiative. Ann Rheum Dis 2012;71(4):484–92
  4. Ness T, Bley TA, Schmidt WA, Lamprecht P. The diagnosis and treatment of giant cell arteritis. Dtsch Arztebl Int 2013;110(21):376–86
  5. Barrack R, Bhimani S, Blevins JL, Blevins K, Demetres M, Figgie M, et al. General assembly, diagnosis, laboratory test: proceedings of international consensus on orthopedic infections. J Arthroplasty 2019;34(2S):S187–95
  6. Eichenauer DA, Aleman BM, André M, Federico M, Hutchings M, Illidge T, et al. Hodgkin lymphoma: ESMO clinical practice guidelines for diagnosis, treatment, and follow-up. Ann Oncol 2018;29(4):19–29
  7. Westergren ALF. Studies of the suspension stability of the blood in pulmonary tuberculosis. Acta Medica Scandinavica 1921:54(1);247–82
  8. Reference method for the erythrocyte sedimentation rate (ESR) test on human blood. Br J Haematol 1973;24(5):671–3
  9. Jou JM, Lewis SM, Briggs C, Lee SH, De La Salle B, McFadden S; International Council for Standardization in Haematology. ICSH review of the measurement of the erythocyte sedimentation rate. Int J Lab Hematol 2011;33(2):125–32
  10. Clinical and Laboratory Standards Institute (CLSI). Procedures for the erythrocyte sedimentation rate test; approved standard-fifth edition. CLSI document H02-A5. Wayne, Pennslyvania: Clinical and Laboratory Standards Institute; 2011
  11. Kratz A, Plebani M, Peng M, Lee YK, McCafferty R, Machin SJ. International Council for Standardization in Haematology (ICSH). ICSH recommendations for modified and alternate methods measuring the erythrocyte sedimentation rate. Int J Lab Hematol 2017;39(5):448–57
  12. Manley RW. The effect of room temperature on erythrocyte sedimentation rate and its correction. J Clin Pathol 1957;10(4):354–6
  13. Lapic I, Piva E, Spolaore F, Musso G, Tosato F, Pelloso M, et al. Ves-Matic CUBE 200: is modified Westergren method for erythrocyte sedimentation rate a valid alternative to the gold standard? J Clin Pathol 2019;72(10):716–9
  14. Lapić I, Piva E, Spolaore F, Tosato F, Pelloso M, Plebani M. Automated measurement of t