Discontinuous moving shot technique for conformal thermal ablation in an ex vivo porcine liver model

Sayfa
418–423
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

This unique ablation technique may be suitable for conformal thermal ablation.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:PURPOSE We aimed to determine the ablation characteristics of discontinuous moving shot technique (DMST) in microwave ablation (MWA), radiofrequency ablation (RFA) and laser ablation (LA), and analyze the differences compared with fixed electrode technique (FET) in an ex vivo porcine liver model.METHODSFET was defined as the ablation needle remaining fixed during ablation.
📊
Temel Bulgular & Veri:Four moving distances (0.5 cm, 0.75 cm, 1 cm and 2 cm) were used in DMST.
💡
Sonuç & Katkı:This unique ablation technique may be suitable for conformal thermal ablation.
🔬Tür:Modelleme & Sayısal Simülasyon
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Alan Parametreleri

Özet

PURPOSE We aimed to determine the ablation characteristics of discontinuous moving shot technique (DMST) in microwave ablation (MWA), radiofrequency ablation (RFA) and laser ablation (LA), and analyze the differences compared with fixed electrode technique (FET) in an ex vivo porcine liver model.METHODSFET was defined as the ablation needle remaining fixed during ablation. In DMST, ablation needle moved backward for a fixed distance twice along the long axis during ablation. Four moving distances (0.5 cm, 0.75 cm, 1 cm and 2 cm) were used in DMST. Long-axis diameter (LAD) and short-axis diameter (SAD) of ablation zones were measured. The ratio of LAD/SAD was calculated.RESULTSThe shape and size of ablation zones were different between DMST and FET. Compared with FET, DMST could achieve greater LAD when the moving distance became long enough. In MWA with DMST, SAD decreased with the extension of moving distance and finally became smaller than the SAD in FET. While in LA and RFA, the change of moving distance did not affect SAD significantly.CONCLUSIONIn MWA, RFA and LA, the characteristics of ablation zone of DMST were different from that of FET. This unique ablation technique may be suitable for conformal thermal ablation.

🔖

Bu Makaleye Atıf Yap

MA, G., CHEN, H., XU, J., PAN, H., YU, M., WANG, Y., XIE, H., ZHOU, W., WANG, S. (2021). Discontinuous moving shot technique for conformal thermal ablation in an ex vivo porcine liver model. Diagnostic and Interventional Radiology, Cilt: 27 - Sayı: 3, 418–423. https://ajindex.com/articles/345540

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (13)

  1. Liang P, Dong B, Yu X, et al. Prognostic factors for survival in patients with hepatocellular carcinoma after percutaneous microwave ablation. Radiology 2005
  2. 299–307. [Crossref] 2. Feng K, Yan J, Li X, et al. A randomized controlled trial of radiofrequency ablation and surgical resection in the treatment of small hepatocellular carcinoma. J Hepatol 2012
  3. 794–802. [Crossref] 3. Chen MS, Li JQ, Zheng Y, et al. A prospective randomized trial comparing percutaneous local ablative therapy and partial hepatectomy for small hepatocellular carcinoma. Ann Surg 2006
  4. 321–328. [Crossref] 4. Gervais DA, Arellano RS. Percutaneous tumor ablation for hepatocellular carcinoma. AJR Am J Roentgenol 2011
  5. 789–794. [Crossref] 5. Goldberg SN, Gazelle GS, Mueller PR. Thermal ablation therapy for focal malignancy: A unified approach to underlying principles, techniques, and diagnostic imaging guidance. AJR Am J Roentgenol 2000
  6. 323–331. [Crossref] 6. Yu J, Liang P, Yu X, Liu F, Chen L, Wang Y. A comparison of microwave ablation and bipolar radiofrequency ablation both with an internally cooled probe: Results in ex vivo and in vivo porcine livers. Eur J Radiol 2011
  7. 124–130. [Crossref] 7. Ha EJ, Baek JH, Lee JH. Moving-shot versus fixed electrode techniques for radiofrequency ablation: Comparison in an ex-vivo bovine liver tissue model. Korean J Radiol 2014
  8. 836– 843. [Crossref] 8. Rhim H, Goldberg SN, Dodd GD, 3rd, et al. Essential techniques for successful radio-frequency thermal ablation of malignant hepatic tumors. Radiographics 2001
  9. S17–39. [Crossref] 9. Baek JH, Kim YS, Sung JY, Choi H, Lee JH. Locoregional control of metastatic well-differentiated thyroid cancer by ultrasound-guided radiofrequency ablation. AJR Am J Roentgenol 2011
  10. W331–336. [Crossref] 10. Baek JH, Kim YS, Lee D, Huh JY, Lee JH. Benign predominantly solid thyroid nodules: Prospective study of efficacy of sonographically guided radiofrequency ablation versus control condition. AJR Am J Roentgenol 2010
  11. 1137–1142. [Crossref] 11. Lee JH, Kim YS, Lee D, Choi H, Yoo H, Baek JH. Radiofrequency ablation (RFA) of benign thyroid nodules in patients with incompletely resolved clinical problems after ethanol ablation (EA). World J Surg 2010
  12. 1488–1493. [Crossref] 12. Zhang X, Ge Y, Ren P, Liu J, Chen G. Horner syndrome as a complication after thyroid microwave ablation: Case report and brief literature review. Medicine 2018
  13. e11884. [Crossref] 13. Kim C, Lee JH, Choi YJ