İlk yerli üretim katlanabilir hidrofobik akrilik göz içi lensi Zaraccom F260 ile alınan sonuçlar: Karşılaştırmalı çalışma

The results of the first domestic manufactured foldable hydrophobic acrylic intraocular lens Zaraccom F260: A comparative study

Sayfa
90–94
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

Bulgular: Gruplar arasında yaş, cinsiyet, etkili fako zamanı, katarakt tipi, takip süresi, preoperatif ve postoperatif görme keskinliği açısından fark yoktu (p>0.05).

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Amaç: Türkiye’ de üretilen ilk hidrofobik akrilik katlanabilir göz içi lensi (GİL) olan Zaraccom F260 ile alınan sonuçları değerlendirmek ve Acrysof Natural SN60AT ile karşılaştırmak.
📊
Temel Bulgular & Veri:Gereç ve Yöntem: Prospektif randomize planlanan bu çalışmada, fakoemulsifikasyon yapılan ve hidrofobik akrilik GİL’leri Zaraccom F260 (Grup 1;50 göz) ile Acrysof Natural SN60AT (Grup 2;50 göz) yerleştirilen olgulara ait sonuçlar karşılaştırıldı.
💡
Sonuç & Katkı:Bulgular: Gruplar arasında yaş, cinsiyet, etkili fako zamanı, katarakt tipi, takip süresi, preoperatif ve postoperatif görme keskinliği açısından fark yoktu (p>0.05).
🔬Tür:Araştırma Makalesi
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Karşılaştırmalı çalışma · Katarakt çıkartılması · Kontakt lensler

Özet

Amaç: Türkiye’ de üretilen ilk hidrofobik akrilik katlanabilir göz içi lensi (GİL) olan Zaraccom F260 ile alınan sonuçları değerlendirmek ve Acrysof Natural SN60AT ile karşılaştırmak. Gereç ve Yöntem: Prospektif randomize planlanan bu çalışmada, fakoemulsifikasyon yapılan ve hidrofobik akrilik GİL’leri Zaraccom F260 (Grup 1;50 göz) ile Acrysof Natural SN60AT (Grup 2;50 göz) yerleştirilen olgulara ait sonuçlar karşılaştırıldı. GİL’leri 3.0 mm’lik kesiden enjeksiyon yöntemiyle veya 4.1 mm’lik kesiden katlama yöntemiyle yerleştirildi. Katarakt operasyonu sırasında görülen komplikasyonlar ve GİL yerleştirilmesi sırasında yaşanılan sorunlar kaydedildi. Takip muayeneleri 1, 7, 30 ve 90. günlerde gerçekleştirildi. Kontrollerde rutin muayenelere ek olarak kornea ödemi, ön kamarada hücre, fibrin reaksiyonu, ön kapsülde fibrozis, arka kapsül opasifikasyonu ve hedeflenen sonuç refraksiyon değerinden sapma değerlendirildi. Bulgular: Gruplar arasında yaş, cinsiyet, etkili fako zamanı, katarakt tipi, takip süresi, preoperatif ve postoperatif görme keskinliği açısından fark yoktu (p>0.05). GİL yerleştirme teknikleri benzerdi (p=0.539). İntraoperatif komplikasyon oluşması açısından fark yoktu (p=0.617). GİL’nin yerleştirilmesi sırasında gözlenen en önemli sorun Grup 1’de 9 gözde lens optiği üzerinde bir miktar materyal soyulmasına benzer birikintilerin görülmesi ve enjektör sistemi kullanılan 3 gözde de kartuşun patlamasıydı. Kornea ödemi ve ön kamarada hücre miktarı incelendiğinde gruplar arasında fark olmadığı izlendi (p>0.05). Her iki grupta ön kapsül fibrozisi ve arka kapsül opasifikasyonu gelişmesi benzerdi (p>0.05). Postoperatuar hedeflenen refraksiyon değerinden sapma miktarı araştırıldığında gruplar arasında fark olmadığı görüldü (p>0.05). Sonuç: Zaraccom F260’ın Acrysof Natural SN60AT ile karşılaştırılmasında benzer, sonuçlarla karşılaşılmış olup, birçok özellik açısından iki GİL arasında önemli bir fark saptanmamıştır. Hastalarda görsel yönden alınan sonuçlar son derece yüz güldürücü olup, erken dönemde lenslerin kapsüler biyouyumluluğu da iyi bulunmuştur. Lense özgü kartuş ya da implantasyon sistemlerinin geliştirilmesi gerektiği kanaatine varılmıştır.

Abstract

Purpose: To evaluate the results of the first foldable hydrophobic acrylic intraocular lens (IOL) Zaraccom F260 manufactured in Turkey and compare the results with Acrysof Natural SN60AT. Material and Method: In this prospective randomized study, the results of the cases who had phacoemulsification surgery and implantation of hydrophobic acrylic IOLs Zaraccom F260 (Group 1;50 eyes) and Acrysof Natural SN60AT (Group 2;50 eyes) were compared. The IOLs were implanted either by injection method through 3.0 mm incision or by folding method through 4.1 mm incision. Complications which occured during the cataract surgery and the problems occured while implanting the IOLs were recorded. The follow-up examinations were performed at days 1, 7, 30 and 90. Corneal edema, cellular reaction in the anterior chamber, fibrin reaction, fibrosis of the anterior capsule, posterior capsule opacification and deviation from the target refraction values were all evaluated other than the routine follow-up examinations. Results: There was not difference between the groups according to age, sex, effective phaco time, cataract type, follow-up period, preoperative and postoperative visual acuities (p>0.05). The IOL implantation techniques were similar (p=0.539). There was not difference according to intraoperative complication occurence rates (p=0.617). The most important problem seen while implanting the IOLs in Group 1 was that; in 9 eyes there was a material on the optic of the IOLs which looked like an exfoliated material and the cartridge was deformed in 3 eyes of the cases where the injection method was used. When corneal edema and cell amount in the anterior chamber were evaluated, no difference was seen between groups (p>0.05). Anterior capsule fibrosis and posterior capsule opacification formations were similar in both groups (p>0.05). Deviation amount from the postoperatively targeted refraction values was not different between groups (p>0.05). Conclusion: Similar results were seen while comparing Zaraccom F260 and Acrysof Natural SN60AT and according to many characteristics, no important difference was found between the IOLs. The visual results were exceedingly well and the capsular biocompatibility of the lenses was very good in the early postoperative period. It was also concluded that special cartrigdes or implantation systems for the lens must be developed.

🔖

Bu Makaleye Atıf Yap

Takmaz, T., CAN, İ., BİLKAY, H., AŞIK, Ş. (2007). İlk yerli üretim katlanabilir hidrofobik akrilik göz içi lensi Zaraccom F260 ile alınan sonuçlar: Karşılaştırmalı çalışma. MN Oftalmoloji, Cilt: 14 - Sayı: 2, 90–94. https://ajindex.com/articles/295937

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (13)

  1. Barbour W, Saika S, Miyamoto T, Ohnishi Y. Biological com patibility of polymethyl methacrylate, hydrophilic acrylic and hydrophobic acrylic intraocular lenses. Ophthalmic Res 2005; 37:255-61
  2. Amon M, Menapace R, Radax U, Freyler H. In vivo study of cell reactions on poly(methyl methacrylate) intraocular lenses with different surface properties. J Cataract Refract Surg 1996; 22:825-9
  3. Abela-Formanek C, Amon M, Schild G, Schauersberger J, He inze G, Kruger A. Uveal and capsular biocompatibility of hydrophilic acrylic, hydrophobic acrylic, and silicone intraocu lar lenses. J Cataract Refract Surg 2002; 28:50-61
  4. Tehrani MTBiek HB, Wolters B, Pakula T, Wolf E. Material pro perties of various intraocular lenses in an experimental study. Ophthalmologica 2004; 218:57-63
  5. Kugelberg M, Wejde G, Jayaram H, Zetterström C. Posterior capsule opacification after implantation of a hydrophilic or a hydrophobic acrylic intraocular lens: One-year follow-up. J Ca taract Refract Surg 2006; 32:1627-31
  6. Werner L, Pandey SK, Escobar-Gomez M, Visessook N, Peng Q, Apple DJ. Anterior capsule opasification; a histopathologi cal study comparing different IOL styles. Ophthalmology 2000; 107:463-71
  7. Tognetto D, Toto L, Ballone E, Ravalico G. Biocompatibility of hydrophilic intraocular lenses. J Cataract Refract Surg 2002; 28:644-51
  8. Miyake K, Ota I, Miyake S, Maekubo K. Correlation between in traocular lens hydrophilicity and anterior capsule opasification and aqueous flare. J Cataract Refract Surg 1996;22:764-9
  9. Scaramuzza A, FernandoGT, Crayford BB. Posterior capsule opacification and lens epithelial cell layer formation: Hydro view hydrogel versus AcrySof acrylic intraocular lenses. J Ca taract Refract Surg 2001; 27: 1047-54
  10. Amon M, Menapace R. In vivo documentation of cellular re actions on lens surfaces for assessing the biocompatibility of different intraocular implants. Eye 1994; 8:649-56
  11. Schauersberger J, Amon M, Kruger A, Abela C, Schild G, Ko lodaschna J. Comparison of the biocompatibility of 2 foldable intraocular lenses with sharp optic edges. J Cataract Refract Surg 2001; 27:1579-85
  12. Ravalico G, Tognetto D, Palomba MA, et al. Capsulorhexis size and posterior capsule opasification. J Cataract Refract Surg 1996; 22:98-103
  13. Ursell PG, Spalton DJ, Pande MV, Hollick EJ, Barman S, Boyce J, Tilling K. Relationship between intraocular lens bio materials and posterior capsule opasification. J Cataract Ref ract Surg 1998; 24:352-60. 14.