Dergiler / Glokom Katarakt / 2017 / Cilt: 12 - Sayı: 2

Tonometri-Pakimetri Kombine Cihazının (Topcon CT-1P) Merkezi Kornea Kalınlık Ölçümlerinin Ultrason Pakimetri ve Optik Düşük Koherens Refl ektometri ile Karşılaştırılması

Comparison of Central Corneal Thickness Measurements of Tonometry- Pachymetry Combined Device (Topcon CT-1P) with Optical Low-Coherence Refl ectometry and Ultrasonic Pachymetry

Sayfa
93–97
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

Ölçümlerin birbiri ile olan uyumu bağımlı örneklem t testi ve Bland-Altman grafi kleri ile değerlendirildi.Bulgular: Yetmiş bir gönüllünün 36'sı kadın 35'i erkek olup ortalama yaşları 35,4±8,3 idi.Ortalama sferik eşdeğeri -1.15±0.67 D olarak saptandı.Ortalama MKK değerleriLenStar ile 533,4±35,8 ?m, Topcon CT-1P ile 542,9±36,6 ?m, ultrason pakimetri ile 532,4±36,2 ?m olarak saptandı.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Amaç: Tonometri-pakimetri kombine cihazlarından Topcon CT-1P'nin merkezi kornea kalınlığı (MKK) ölçümlerini optik düşük koherens refl ektometri (LenStar) ve ultrasonik pakimetri ile karşılaştırmak.Gereç ve Yöntem: Sağlıklı gönüllülerin tam oftalmolojik muayenesi yapıldıktan sonra aynı seansta sırasıyla LenStar LS900, Topcon CT-1P ve ultrason pakimetri (PacScan 300A) ile MKK ölçümleri yapıldı.
📊
Temel Bulgular & Veri:Ölçümlerin birbiri ile olan uyumu bağımlı örneklem t testi ve Bland-Altman grafi kleri ile değerlendirildi.Bulgular: Yetmiş bir gönüllünün 36'sı kadın 35'i erkek olup ortalama yaşları 35,4±8,3 idi.Ortalama sferik eşdeğeri -1.15±0.67 D olarak saptandı.Ortalama MKK değerleriLenStar ile 533,4±35,8 ?m, Topcon CT-1P ile 542,9±36,6 ?m, ultrason pakimetri ile 532,4±36,2 ?m olarak saptandı.
💡
Sonuç & Katkı:Bu durum glokom taramalarında ve özellikle ileri evre glokom hastalarının takibinde önemli olabilir.
🔬Tür:Uygulamalı Anket & Alan Araştırması
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Alan Parametreleri

Özet

Amaç: Tonometri-pakimetri kombine cihazlarından Topcon CT-1P'nin merkezi kornea kalınlığı (MKK) ölçümlerini optik düşük koherens refl ektometri (LenStar) ve ultrasonik pakimetri ile karşılaştırmak.Gereç ve Yöntem: Sağlıklı gönüllülerin tam oftalmolojik muayenesi yapıldıktan sonra aynı seansta sırasıyla LenStar LS900, Topcon CT-1P ve ultrason pakimetri (PacScan 300A) ile MKK ölçümleri yapıldı. Ölçümlerin birbiri ile olan uyumu bağımlı örneklem t testi ve Bland-Altman grafi kleri ile değerlendirildi.Bulgular: Yetmiş bir gönüllünün 36'sı kadın 35'i erkek olup ortalama yaşları 35,4±8,3 idi.Ortalama sferik eşdeğeri -1.15±0.67 D olarak saptandı.Ortalama MKK değerleriLenStar ile 533,4±35,8 ?m, Topcon CT-1P ile 542,9±36,6 ?m, ultrason pakimetri ile 532,4±36,2 ?m olarak saptandı. Topcon CT-1P ile MKK diğer iki yönteme göre anlamlı olarak daha yüksek ölçülmektedir (ikisi için p<0,001). Ayrıca Topcon CT1-P ile diğer iki cihazın ölçümleri arasında korelasyon mevcut değildir (LenStar; R=0,023 p=0,456; ultrason pakimetri; R=0.010 p=0,762). LenStar ve ultrason pakimetri ile elde edilen ölçümler ise birbirine oldukça yakındır (p=0,875) ve iki cihazın ölçümleri arasında korelasyon mevcuttur (R=0.600 p=0,001).Sonuç: Topcon CT-1P ile elde edilen MKK ölçümleri hem ultrason pakimetri hem de LenStar ölçümlerine göre yaklaşık 10 ?m daha yüksek çıkmaktadır. Dolayısıyla cihaz tarafından sunulan düzeltilmiş GİB gerçek GİB'na göre yaklaşık 1 mmHg daha düşük olabilir. Bu durum glokom taramalarında ve özellikle ileri evre glokom hastalarının takibinde önemli olabilir.

Abstract

Aim: To compare the central corneal thickness measurements of Topcon CT-1P, a tonometry-pachymetry combined device, with optical low-coherence refl ectometry (LenStar) and ultrasonic pachymetry.Material and Methods: After the full ophthalmologic examination of healthy volunteers, CCT measurements were taken with LenStar LS900, Topcon SA-1P and ultrasound pachymetry in the same session. The correlation of measurements with each other were evaluated with paired samples t-test and Bland-Altman graphs.Results: Thirty six percent of 71 volunteers were female and 35% were male and the average age was 35.4±8,3 years. The mean spherical equivalent was d etected as -1.15±0.67 D. The mean CCT values measured by LenStar, Topcon SA-1P and ultrasonic pachymetry were 533,4±35.8 ?m, 542,9±36,6 ?m, 532,4±36,2 ?m, respectively. CCT measured with Topcon SA-1P was signifi cantly higher than the other two methods (p<0.001,in both cases). Also, there was no correlation between the measurements of Topcon CT1-P and other two devices. (LenStar; R=0,023 p=0,456, ultrasonic pachymetry; R=-0.010 p=0,762). The measurements obtained with LenStar and ultrasonic pachymetry were pretty close to each other (p:0,875). And there was a correlation between the measurements of the two devices (R=0.600, p=0.001).Conclusion: The mean CCT measured by Topcon SA-1P was approximately 10 ?m higher than with both ultrasound pachymetry and LenStar. Hence, the corrected IOP presented by the device might be approximately 1 mm Hg lower than the true IOP. This condition can be important for glaucoma screening and especially in follow-up of patients with advanced stage glaucoma.

🔖

Bu Makaleye Atıf Yap

KOÇ, M., TEKİN, K., YETKİN, E., ÇITIRIK, M., YILMAZBAŞ, P. (2017). Tonometri-Pakimetri Kombine Cihazının (Topcon CT-1P) Merkezi Kornea Kalınlık Ölçümlerinin Ultrason Pakimetri ve Optik Düşük Koherens Refl ektometri ile Karşılaştırılması. Glokom Katarakt, Cilt: 12 - Sayı: 2, 93–97. https://ajindex.com/articles/192604

Kaynakça (12)

  1. Herndon LW. Measuring intraocular pressure-adjustments for corneal thickness and new Technologies. Curr Opin Ophthalmol 2006;17:115-9.
  2. Price FW, Koller DL, Price MO. Central corneal pachymetry in patients undergoing laser in situ keratomileusis. Ophthalmology 1999;106:2216-20.
  3. Beutelspacher SC, Serbecic N, Scheuerle AF. Assessment of cent- ral corneal thickness using OCT, ultrasound, optical-low coheren- ce refl ectometry and Scheimpfl ug pachymetry. Eur J Ophthalmol 2011;21:132-7.
  4. Garcia-Resua C, Pena-Verdeal H, Minones M et al. Reliability of the non-contact tono-pachymeter Tonopachy NT-530P in healthy eyes Clin Exp Optom 2013;96:286-94.
  5. Fujimura F, Kamiya K, Fujiwara K et al. Repeatability and repro- ducibility of measurment using a NT-530P noncontact tono/ pachymeter and correlation of central corneal thickness with intraocular pressure. Biomed Res Int. 2013;2013:370592. doi: 10.1155/2013/370592. Lee YG, Kim JH, Kim NR et al Comprasion between Tonopachy and other tonometric and pachymetric devices. Optom Vis Sci 2011;88-843-9.
  6. Dervişoğulları MS, Totan Y, Yuvacı İ ve ark. Comprasion of central corneal thickness measurment of Nidek Al-Scan, Gali- lei G4 dual Scheimpfl ug analyzer and Cirrus HD-OCT. Glo-Kat 2015;10:287-292.
  7. Muscat S, McKay N, Parks S et al. Repeatability and reproducibi- lity of corneal thickness measurment by optical coherence tomog- raphy.Invest Ophthalmol 2002;43:1791-5.
  8. Schiano LD, Lombardo M, Tranchina L et al. Repeatability of intra-colar pressure and central corneal thickness measurment provided by a non-contact method of tonometry and pachymetry. Graefes Arch Clin Exp Ophthalmol 2011;249:429-34.
  9. Wells M, Wu N, Kokkinakis J et al. Correlation of central cor- neal thickness measurments using Topcon TRK- P, Zeiss Visante AS-OCT and DGH pachmate 55 handheld ultrasonic pachymeter. Clin Exp Optom 2013;96:385-7.
  10. Huang J, Liao N, Savini G et al. Measurment of corneal thickness with optical low-coherence refl ectometry and ultrasound pachy- metry in normal and post-femtosecond laser in situ keratomiluesis eyes. Cornea 2015;34:204-8.
  11. Şimşek A, Bilak Ş, Güler M ve ark. Comprasion of central corneal thickness measurments obtained by RTVue OCT, Lenstar, Sirus Topography and ultrasound pachymetry in healthy subjects. Semin Ophthalmol 2015;21:1-6.
  12. Bayhan HA, Bayhan SA, Muhafız E ve ark. Comprasion of an- terior segment parameters with optical low coherence refl ectome- ter and combine Schimpfl ug-placido disc topographer. Glo-Kat 2013;8:78-82.