A test and simulation device for Doppler-based fetal heart rate monitoring

Sayfa
1187–1194
Yayın Tarihi
2015-08-13
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

This device has been tested and its reliability has been proved.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:The Doppler effect is the preferred technique in fetal heart rate (FHR) monitoring devices.
📊
Temel Bulgular & Veri:This device generates a similar Doppler frequency shift of fetal cardiac activity including the heart's wall and valve motions.
💡
Sonuç & Katkı:This device has been tested and its reliability has been proved.
🔬Tür:Araştırma Makalesi
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Doppler effect · fetal heart simulation · fetal heart rate

Özet

The Doppler effect is the preferred technique in fetal heart rate (FHR) monitoring devices. The main objective of the recent studies on the Doppler FHR has been to improve the accuracy. On the other hand, a reliable fetal heart simulator becomes essential for testing Doppler FHR monitoring devices. The motivation of this study is to design a reliable system that will be used to test Doppler FHR monitors. This device generates a similar Doppler frequency shift of fetal cardiac activity including the heart's wall and valve motions. The components of this system are basically a signal generator using a microcontroller and a modified relay. The relay is the most important and studied component that affects the Doppler frequency shift. Thus, the relay's contact distance and length as well as its closing and opening velocities have been modified to produce an appropriate Doppler shift to fetal cardiac motion. This device has been tested and its reliability has been proved.

Abstract

The Doppler effect is the preferred technique in fetal heart rate (FHR) monitoring devices. The main objective of the recent studies on the Doppler FHR has been to improve the accuracy. On the other hand, a reliable fetal heart simulator becomes essential for testing Doppler FHR monitoring devices. The motivation of this study is to design a reliable system that will be used to test Doppler FHR monitors. This device generates a similar Doppler frequency shift of fetal cardiac activity including the heart's wall and valve motions. The components of this system are basically a signal generator using a microcontroller and a modified relay. The relay is the most important and studied component that affects the Doppler frequency shift. Thus, the relay's contact distance and length as well as its closing and opening velocities have been modified to produce an appropriate Doppler shift to fetal cardiac motion. This device has been tested and its reliability has been proved.

🔖

Bu Makaleye Atıf Yap

MERT, A., SEZDİ, M., AKAN, A. (2015). A test and simulation device for Doppler-based fetal heart rate monitoring. Turkish Journal of Electrical Engineering and Computer Science, Cilt: 23 - Sayı: 4, 1187–1194. https://ajindex.com/articles/187515