The Effects of Corneal Edema on the Intraocular Pressure Measurements

Sayfa
285–292
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

However, there is still no gold standard method for IOP measurement in edematous corneas.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:We aimed to evaluate the possible effects of corneal edema on intraocular pressure (IOP) measurements obtained with three different tonometers on the first day after cataract surgery.
📊
Temel Bulgular & Veri:Sixty-four eyes of the 64 patients underwent phacoemulsification and IOP measurements using Goldmann applanation tonometer (GAT), Tono-Pen applanation tonometer (TAT), and Pascal digital contour tonometer (DCT) the first day after surgery.
💡
Sonuç & Katkı:However, there is still no gold standard method for IOP measurement in edematous corneas.
🔬Tür:Araştırma Makalesi
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Alan Parametreleri

Özet

We aimed to evaluate the possible effects of corneal edema on intraocular pressure (IOP) measurements obtained with three different tonometers on the first day after cataract surgery. Sixty-four eyes of the 64 patients underwent phacoemulsification and IOP measurements using Goldmann applanation tonometer (GAT), Tono-Pen applanation tonometer (TAT), and Pascal digital contour tonometer (DCT) the first day after surgery. Central corneal thickness (CCT) measurements were used to quantify the postoperative corneal edema. Intraclass correlation coefficient and Bland-Altman plot were used to assess the compatibility of IOP measurements among three method pairs. TAT and DCT IOP measurements were found to be compatible with each other (p=0.141) in the corneal edema group. The GAT and TAT IOP measurements in the eyes with corneal edema were inversely proportional to the CCT values (p=0.011 and p=0.033, respectively), but the DCT relationship was not statistically significant (p=0.44). The lack of a significant relationship between DCT and CCT in edematous corneas may suggest that DCT, as in normal corneas, can perform CCT-independent measurements. However, there is still no gold standard method for IOP measurement in edematous corneas.

🔖

Bu Makaleye Atıf Yap

SEVEN, E., TEKİN, S., BATUR, M., YAVUZER, B., YAŞAR, T. (2020). The Effects of Corneal Edema on the Intraocular Pressure Measurements. Eastern Journal of Medicine, Cilt: 25 - Sayı: 2, 285–292. https://ajindex.com/articles/158314

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (14)

  1. Cook JA, Botello AP, Elders A, et al. Systematic review of the agreement of tonometers with Goldmann applanation tonometry. Ophthalmology 2012; 119: 1552-1557. IOf S. Ophthalmic Instruments—Tonometers: ISO8612: 2001.
  2. Geneva, Switzerland: International Organization for Standardization. 2001.
  3. Galgauskas S, Strupaite R, Strelkauskaite E, Asoklis R. Comparison of intraocular pressure measurements with different contact tonometers in young healthy persons. Int J Ophthalmol 2016; 9: 76-80.
  4. Gunvant P, O'Leary DJ, Baskaran M, Broadway DC, Watkins RJ, Vijaya L. Evaluation of tonometric correction factors. J Glaucoma 2005; 14: 337-343.
  5. Schneider E, Grehn F. Intraocular pressure measurement-comparison of dynamic contour tonometry and goldmann applanation tonometry. J Glaucoma 2006; 15: 2-6.
  6. Kohlhaas M, Boehm AG, Spoerl E, Pursten A, Grein HJ, Pillunat LE. Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry. Arch Ophthalmol 2006; 124: 471-476.
  7. Hessemer V, Rössler R, Jacobi KW. Tono-Pen, a new tonometer. International Ophthalmology 1989; 13: 51-56.
  8. Tonnu PA, Ho T, Newson T, et al. The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non- contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry. Br J Ophthalmol 2005; 89: 851-854.
  9. Dohadwala AA, Munger R, Damji KF. Positive correlation between Tono-Pen intraocular pressure and central corneal thickness. Ophthalmology 1998; 105: 1849-1854.
  10. Bhan A, Browning AC, Shah S, Hamilton R, Dave D, Dua HS. Effect of corneal thickness on intraocular pressure measurements with the pneumotonometer, Goldmann applanation tonometer, and Tono-Pen. Investigative ophthalmology & visual science 2002; 43: 1389-1392.
  11. Kanngiesser HE, Kniestedt C, Robert YC. Dynamic contour tonometry: presentation of a new tonometer. J Glaucoma 2005; 14: 344-350.
  12. Liu J, Roberts CJ. Influence of corneal biomechanical properties on intraocular pressure measurement: quantitative analysis. J Cataract Refract Surg 2005; 31: 146-155.
  13. Hamilton KE, Pye DC, Kao L, Pham N, Tran AQ. The effect of corneal edema on dynamic contour and goldmann tonometry. Optom Vis Sci 2008; 85: 451-456.
  14. Herr A, Remky A, Hirsch T, Rennings C, Plange N. Tonometry in corneal edema after cataract surgery: dynamic contour tonometry versus Goldmann applanation tonometry. Clin Ophthalmol 2013; 7: 815-859. Stelzer R. Relationship of the corneal applanations to intraocular pressure in enucleat