Validation of force production during quadriceps set exercise using a hand-held dynamometer: A comparative study with isokinetic dynamometry in healthy individuals

Esmanur Yıldırım(School of Physical Therapy and Rehabilitation, Kırşehir Ahi Evran University, Kırşehir)0009-0006-2987-0925
Abdulhamit Tayfur(School of Physical Therapy and Rehabilitation, Kırşehir Ahi Evran University, Kırşehir)0000-0001-7366-5212
Beyza Tayfur(School of Physical Therapy and Rehabilitation, Kırşehir Ahi Evran University, Kırşehir)0000-0003-4632-9541
Sayfa
—
Yayın Tarihi
2026-02-25
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

These findings indicate that the use of the hand-held dynamometer as a measurement method in the quadriceps set exercise position is valid and reliable, and it may serve as a viable alternative to isokinetic testing, particularly in the early postoperative phase when isokinetic strength assessment is often considered unsafe.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:The study aim was to compare lower extremity muscle strength measurements obtained during the quadriceps set exercise in healthy individuals.
📊
Temel Bulgular & Veri:Twenty-four healthy individuals (23.1±2.1 years, 13 females) were recruited.
💡
Sonuç & Katkı:These findings indicate that the use of the hand-held dynamometer as a measurement method in the quadriceps set exercise position is valid and reliable, and it may serve as a viable alternative to isokinetic testing, particularly in the early postoperative phase when isokinetic strength assessment is often considered unsafe.
🔬Tür:Research Article
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Hand-held dynamometer · isokinetic dynamometer · strength

Abstract

The study aim was to compare lower extremity muscle strength measurements obtained during the quadriceps set exercise in healthy individuals. Twenty-four healthy individuals (23.1±2.1 years, 13 females) were recruited. Lower extremity muscle strength measurements were obtained using a hand-held dynamometer during the quadriceps set exercise and using an isokinetic dynamometer with isometric testing at 30° of knee flexion and at 60°/s concentrically. Validity was assessed using Pearson correlation and Bland and Altman analyses, while within-session reliability was assessed using the Intraclass Correlation Coefficient (ICC). Also, standard error of measurement (SEM%), minimal detectable change (MDC95) and minimal important change (MIC95) at the 95% confidence interval were assessed. Hand-held dynamometer measurements during the quadriceps set exercise were strongly correlated with isometric measurements at 30° of knee flexion (r=0.598, p=0.002) and concentric measurements at 60°/s (r=0.535, p=0.007) on the isokinetic dynamometer. The hand-held dynamometer demonstrated a good within-session reliability (ICC=0.88, 95%CI:0.80–0.94), while the isometric measurements at 30° of knee flexion (ICC=0.91, 95%CI:0.85–0.95) and the isokinetic measurements at 60°/s (ICC=0.92, 95%CI:0.86–0.96) demonstrated an excellent within-session reliability. Across all measurements, SEM% values ranged between 9–11%. The MDC95 was calculated as 5.97 (95%CI:3.5–8.5) for the hand-held dynamometer, 34.25 (95%CI:16.9–51.8) for the isometric measurements at 30° of knee flexion, and 40.0 (95%CI:20.8–59.3) for the isokinetic measurements at 60°/s. The MIC95 values corresponded to an approximate change of 14% for the hand-held dynamometer and 19% for the isokinetic dynamometer relative to the mean muscle strength values of healthy participants. These findings indicate that the use of the hand-held dynamometer as a measurement method in the quadriceps set exercise position is valid and reliable, and it may serve as a viable alternative to isokinetic testing, particularly in the early postoperative phase when isokinetic strength assessment is often considered unsafe.

🔖

Bu Makaleye Atıf Yap

Yıldırım, E., Tayfur, A., Tayfur, B. (2026). Validation of force production during quadriceps set exercise using a hand-held dynamometer: A comparative study with isokinetic dynamometry in healthy individuals. Turkish Journal of Kinesiology, Volume: 12 Issue: 1. https://doi.org/10.31459/turkjkin.1868322

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (36)

  1. Aerts, F., Sheets, H., Anderson, C., Bussie, N., Hoskins, R., Maninga, A., & Novak, E. (2025). Reliability and agreement of hand-held dynamometry using three standard rater test positions. Int J Sports Phys Ther, 20(2), 243–252. doi: 10.26603/001c.128286
  2. Bland, J. M., & Altman, D. G. (2010). Statistical methods for assessing agreement between two methods of clinical measurement. International Journal of Nursing Studies, 47(8), 931-936. doi: 10.1016/j.ijnurstu.2009.10.001
  3. Briggs, K. K., Lysholm, J., Tegner, Y., Rodkey, W. G., Kocher, M. S., & Steadman, J. R. (2009). The reliability, validity, and responsiveness of the Lysholm score and Tegner activity scale for anterior cruciate ligament injuries of the knee: 25 years later. Am J Sports Med, 37(5), 890-897. doi: 10.1177/0363546508330143
  4. Bujang, M. A., & Baharum, N. (2017). A simplified guide to determination of sample size requirements for estimating the value of intraclass correlation coefficient: a review. Arch Orofac Sci, 12(1), 1-11.
  5. Cavanaugh, J. T., & Powers, M. (2017). ACL rehabilitation progression: Where are we now? Curr Rev Musculoskelet Med, 10(3), 289-296. doi: 10.1007/s12178-017-9426-3
  6. Cohen, J. (1988). Statistical power analysis for the behavioral sciences (2nd ed.). Hillsdale. NJ Lawrence Earlbaum Associates.
  7. de Vet, H. C., Terwee, C. B., Ostelo, R. W., Beckerman, H., Knol, D. L., & Bouter, L. M. (2006). Minimal changes in health status questionnaires: distinction between minimally detectable change and minimally important change. Health Qual Life Outcomes, 4(1), 54. doi: 10.1186/1477-7525-4-54
  8. Deones, V. L., Wiley, S. C., & Worrell, T. (1994). Assessment of quadriceps muscle performance by a hand-held dynamometer and an isokinetic dynamometer. J Orthop Sports Phys Ther, 20(6), 296-301. doi: 10.2519/jospt.1994.20.6.296
  9. Eymir, M., Unver, B., & Karatosun, V. (2024). Reliability, concurrent validity, and minimal detectable change of the hand-held dynamometry for the assessment of knee muscle strength in patients with revision total knee arthroplasty. Arch Phys Med Rehabil, 105(1), 34-39. doi: 10.1016/j.apmr.2023.05.002
  10. Heijne, A., & Werner, S. (2007). Early versus late start of open kinetic chain quadriceps exercises after ACL reconstruction with patellar tendon or hamstring grafts: a prospective randomized outcome study. Knee Surg Sports Traumatol Arthrosc, 15(4), 402-414. doi: 10.1007/s00167-006-0246-z
  11. Hirano, M., Katoh, M., Gomi, M., & Arai, S. (2020). Validity and reliability of isometric knee extension muscle strength measurements using a belt-stabilized hand-held dynamometer: a comparison with the measurement using an isokinetic dynamometer in a sitting posture. J Phys Ther Sci, 32(2), 120-124. doi: 10.1589/jpts.32.120
  12. Jewiss, D., Ostman, C., & Smart, N. (2017). Open versus closed kinetic chain exercises following an anterior cruciate ligament reconstruction: a systematic review and meta-analysis. J Sports Med (Hindawi Publ Corp), 2017, 4721548. doi: 10.1155/2017/4721548
  13. Kai-Ming Chan, N. M., Pirkko Korkia, Raymond C. T. Li. (1996). Principles and Practice of Isokinetics in Sports Medicine and Rehabilitation. Williams & Wilkins Asia-Pacific.
  14. Kondo, Y., Yoshida, Y., Iioka, T., Kataoka, H., Sakamoto, J., Hirase, T., Honda, Y., Nawata, A., & Okita, M. (2022). Short-term effects of isometric quadriceps muscle exercise with auditory and visual feedback on pain, physical function, and performance after total knee arthroplasty: a randomized controlled trial. J Knee Surg, 35(8), 922-931. doi: 10.1055/s-0040-1721035
  15. Koo, T. K., & Li, M. Y. (2016). A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med, 15(2), 155-163. doi: 10.1016/j.jcm.2016.02.012
  16. Korakakis, V., Kotsifaki, A., Korakaki, A., Karanasios, S., & Whiteley, R. (2021). Current perspectives and clinical practice of physiotherapists on assessment, rehabilitation, and return to sport criteria after anterior cruciate ligament injury and reconstruction. An online survey of 538 physiotherapists. Phys Ther Sport, 52, 103-114. doi: 10.1016/j.ptsp.2021.08.012
  17. Kosti, C., Tsoukos, A., Pelekis, I., Paschalis, V., Terzis, G., & Bogdanis, G. C. (2025). Isokinetic vs. hand-held dynamometry for assessing knee flexor and extensor strength in athletes: evaluating a low-cost alternative across the range of motion. J Hum Kinet, doi: 10.5114/jhk/210346
  18. Kotsifaki, R., Korakakis, V., King, E., Barbosa, O., Maree, D., Pantouveris, M., Bjerregaard, A., Luomajoki, J., Wilhelmsen, J., & Whiteley, R. (2023). Aspetar clinical practice guideline on rehabilitation after anterior cruciate ligament reconstruction. Br J Sports Med, 57(9), 500-514. doi: 10.1136/bjsports-2022-106158
  19. Kuwahara, K., Makino, Y., Kato, T., Fukuda, A., & Asada, K. (2025). Reliability of a portable device for measuring lower-limb muscle strength in patients who underwent total knee arthroplasty. J Phys Ther Sci, 37(12), 613–618. doi: 10.1589/jpts.37.613
  20. Martin, H., Yule, V., Syddall, H., Dennison, E., Cooper, C., & Aihie Sayer, A. (2006). Is hand-held dynamometry useful for the measurement of quadriceps strength in older people? A comparison with the gold standard Biodex dynamometry. Gerontology, 52(3), 154-159. doi: 10.1159/000091824
  21. Mentiplay, B. F., Perraton, L. G., Bower, K. J., Adair, B., Pua, Y.-H., Williams, G. P., McGaw, R., & Clark, R. A. (2015). Assessment of lower limb muscle strength and power using hand-held and fixed dynamometry: a reliability and validity study. PLoS One, 10(10), e0140822. doi: 10.1371/journal.pone.0140822
  22. Muff, G., Dufour, S., Meyer, A., Severac, F., Favret, F., Geny, B., Lecocq, J., & Isner-Horobeti, M.-E. (2016). Comparative assessment of knee extensor and flexor muscle strength measured using a hand-held vs. isokinetic dynamometer. J Phys Ther Sci, 28(9), 2445-2451. doi: 10.1589/jpts.28.2445
  23. Nurdan, P., Bugdayci, D., Sabirli, F., Ozel, S., & Ersoy, S. (2007). Knee Injury and Osteoarthritis Outcome Score: Reliability and Validation of the Turkish Version. Turkiye Klinikleri Journal of Medical Sciences, 27, 350-356.
  24. Pasco, J. A., Stuart, A. L., Holloway-Kew, K. L., Tembo, M. C., Sui, S. X., Anderson, K. B., Hyde, N. K., Williams, L. J., & Kotowicz, M. A. (2020). Lower-limb muscle strength: normative data from an observational population-based study. BMC Musculoskelet Disord, 21(1), 89. https://doi.org/10.1186/s12891-020-3098-7
  25. Roos, E. M., Roos, H. P., Lohmander, L. S., Ekdahl, C., & Beynnon, B. D. (1998). Knee Injury and Osteoarthritis Outcome Score (KOOS) - Development of a self-administered outcome measure. J Orthop Sports Phys Ther, 28(2), 88-96. doi: 10.2519/jospt.1998.28.2.88
  26. Roy, M. A., & Doherty, T. J. (2004). Reliability of hand-held dynamometry in assessment of knee extensor strength after hip fracture. Am J Phys Med Rehabil, 83(11), 813-818. doi: 10.1097/01.phm.0000143405.17932.78
  27. Stark, T., Walker, B., Phillips, J. K., Fejer, R., & Beck, R. (2011). Hand-held dynamometry correlation with the gold standard isokinetic dynamometry: a systematic review. PM R, 3(5), 472-479. doi: 10.1016/j.pmrj.2010.10.025
  28. Sullivan, S. J., Chesley, A., Hebert, G., McFaull, S., & Scullion, D. (1988). The validity and reliability of hand-held dynamometry in assessing isometric external rotator performance. J Orthop Sports Phys Ther, 10(6), 213-217. doi: 10.2519/jospt.1988.10.6.213
  29. Suzuki, M., Yamada, S., Inamura, A., Omori, Y., Kirimoto, H., Sugimura, S., & Miyamoto, M. (2009). Reliability and validity of measurements of knee extension strength obtained from nursing home residents with dementia. Am J Phys Med Rehabil, 88(11), 924-933. doi: 10.1097/PHM.0b013e3181ae1003
  30. Tayfur, A., Aslan, M., Canlı, M., Gürses, Ö. A., Çakmak, M. F., Yetiş, M., & Ceylan, İ. (2025). Validity, reliability and responsiveness of the Turkish Version of the Munich Knee Questionnaire. Rehabilitation, 64(4), 231-237. doi: 10.1055/a-2618-2483
  31. Tayfur, A., Kuyulu Haksal, P., & Ergun, N. (2025). Equivalence of online and paper administration of Identification of Functional Ankle Instability (IdFAI) scale. Journal of Exercise Therapy and Rehabilitation, 12(2), 136-145. doi: 10.15437/jetr.1652052
  32. Tayfur, A., Özyurt, F., Büyükperdah, B. N., Genç, B. E., Rio, E. K., Murphy, M. C., & Tayfur, B. (2025). Construct validity, reliability and interpretability of the Turkish Version of the TENDINopathy Severity Assessment–Achilles (TENDINS-A) Questionnaire. Foot Ankle Surg, S1268-7731(25)00266-8. doi: 10.1016/j.fas.2025.11.009
  33. Tayfur, B., Johnson, A. K., & Palmieri‐Smith, R. M. (2025). Compromised quadriceps and hamstring force control, not maximal strength, is associated with gait biomechanics at 9 months following anterior cruciate ligament reconstruction. Knee Surg Sports Traumatol Arthrosc, 33(12), 4466-4474. doi: 10.1002/ksa.70129
  34. Topal, Y., Kınıklı, G. İ., Bozgeyik, S., & Deniz, H. G. (2023). Relationship of lower extremity muscle strength, range of hip motion and subtalar angle with dynamic balance. Turk J Physiother Rehabil, 34(1), 55-63. doi: 10.21653/tjpr.1016732
  35. Trybulski, R., Wieckowski, J., Muracki, J., Matuszczyk, F., Galeziok, K., Wilk, M., & Kuzdzal, A. (2025). Reliability and reproducibility of the Kinvent K-push dynamometer for assessing quadriceps strength and force development in athletes and untrained individuals. Front Physiol, 16, 1573748. doi: 10.3389/fphys.2025.1573748
  36. van Melick, N., van Cingel, R. E., Brooijmans, F., Neeter, C., van Tienen, T., Hullegie, W., & Nijhuis-van der Sanden, M. W. (2016). Evidence-based clinical practice update: practice guidelines for anterior cruciate ligament rehabilitation based on a systematic review and multidisciplinary consensus. Br J Sports Med, 50(24), 1506-1515. doi: 10.1136/bjsports-2015-095898