Dergiler / Bulletin of Biotechnology / 2025 / Volume: 6 Issue: 2

Comparative Study of Metal Oxide Nanoparticles for Enhanced Amperometric Glucose Detection in Real Samples Using a Polypyrrole-Based Biosensor

Gül Ozyılmaz(MUSTAFA KEMAL ÜNİVERSİTESİ)0000-0002-2373-6219
Rağibe Hülya Akyürekoğlu(MUSTAFA KEMAL UNIVERSITY)
Sayfa
—
Yayın Tarihi
2025-12-29
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

Consequently, it can be confidently asserted that the developed amperometric glucose biosensor exhibits significant potential and robustness for the precise and reliable determination of glucose concentrations in complex real-sample matrices.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:In this study, we developed a highly sensitive amperometric glucose biosensor based on a nanocomposite of electrochemically synthesized polypyrrole and metal oxide nanoparticles.
📊
Temel Bulgular & Veri:A comparative analysis of the current response to varying glucose concentrations was conducted for enzyme electrodes modified with metal oxide nanoparticles (NPs) at concentrations of 0.1, 0.5, and 2.0 mg/mL.
💡
Sonuç & Katkı:Consequently, it can be confidently asserted that the developed amperometric glucose biosensor exhibits significant potential and robustness for the precise and reliable determination of glucose concentrations in complex real-sample matrices.
🔬Tür:Modelleme & Sayısal Simülasyon
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Amperometrik biosensor · Glucose Oxidase · Polypyrrole

Abstract

In this study, we developed a highly sensitive amperometric glucose biosensor based on a nanocomposite of electrochemically synthesized polypyrrole and metal oxide nanoparticles. A platinum electrode was coated by PPy film by cyclic voltammetry, which served as a host matrix for the immobilization of GOD using a chitosan-glutaraldehyde cross-linking system. We conducted a comparative study to evaluate the signal-enhancing effects of incorporating ZnO, TiO₂, and zinc titanate (ZnTiO₃) nanoparticles into the immobilization matrix. A comparative analysis of the current response to varying glucose concentrations was conducted for enzyme electrodes modified with metal oxide nanoparticles (NPs) at concentrations of 0.1, 0.5, and 2.0 mg/mL. The results unequivocally demonstrated that across all three concentrations tested, the electrode modified with ZnO NPs (PPy/ZnO/GOD) yielded a substantially superior amperometric response compared to those incorporating other metal oxides. This pronounced enhancement in current signal underscores the exceptional efficacy of ZnO nanoparticles in facilitating electron transfer within the biosensor matrix. Subsequent to this optimization, the ZnO NP-modified biosensor was deployed for the quantitative determination of glucose in a series of real-world samples, namely apple juice, grape juice, a fizzy beverage, and honey. To rigorously validate the accuracy and reliability of the developed biosensor, its performance was benchmarked against a commercially available spectrophotometric glucose assay kit. The analytical data obtained from both methods exhibited a strong correlation, confirming the high degree of accuracy of the fabricated biosensor. Consequently, it can be confidently asserted that the developed amperometric glucose biosensor exhibits significant potential and robustness for the precise and reliable determination of glucose concentrations in complex real-sample matrices.

🔖

Bu Makaleye Atıf Yap

Ozyılmaz, G., Akyürekoğlu, R. H. (2025). Comparative Study of Metal Oxide Nanoparticles for Enhanced Amperometric Glucose Detection in Real Samples Using a Polypyrrole-Based Biosensor. Bulletin of Biotechnology, Volume: 6 Issue: 2. https://doi.org/10.51539/biotech.1787438

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (18)

  1. Albanese D, Sannini A, Malvano F, Pilloton R, Di Matteo (2014) Optimisation of glucose biosensors based on sol-gel entrapment and Prussian Blue-modified screen-printed electrodes for real food analysis. Food Anal. Methods, 7: 1002-1008 doi: 10.1007/s12161-013-9705-6
  2. Ayenimo JG, Adeloju SB (2017) Amperometric detection of glucose in fruit juices with polypyrrole-based biosensor with an integrated permselective layer for exclusion of interferences. Fodd Chem, 229: 127-135 doi: 10.1016/j.foodchem.2017.01.138
  3. Belluzo MS, Ribone ME, Lagier CM (2008) Assembling amperometric biosensors for clinical diagnostic. Sensors (Basel), 8(3):1366-1399. doi: 10.3390/s8031366
  4. Chauhan N, Pundir CS (2012) An amperometric acetylcholinesterase sensor based on Fe3O4 nanoparticle/multi-walled carbon nanotube-modified ITO-coated glass plate for the detection of pesticides. Electrochim Acta, 67: 79-86. doi: 10.1016/j.electacta.2012.02.012
  5. Chahrour KM, Ooi PC, Nazeer AA, Al-Hajji LA, Jubu PR, Dee CF, Ahmadipour M, Hamzah AA (2023). CuO/Cu/rGO nanocomposite anodic titania nanotubes for boosted non-enzymatic glucose biosensors. New J Chem, 47(16): 7890–7902. doi: 10.1039/D3NJ00666B
  6. Hu F, Chen S, Wang C, Yuan R, Chai Y, Xiang Y, Wang C. (2011). ZnO nanoparticle and multiwalled carbon nanotubes for glucose oxidase direct electron transfer and electrocatalytic activity investigation. J Mol Catal B: Enzym, 72:298-304. doi: 10.1016/j.molcatb.2011.07.005
  7. Kausaite-Minkstimiene A, Krikstaponyte A, Stasyuk N, Gayda G, Ramanaviciene A (2025). An Amperometric Enzyme–Nanozyme Biosensor for Glucose Detection. Biosensors, 15: 545. doi: 10.3390/bios15080545
  8. Lin YH, Sivakumar C, Balraj B, Murugesan G, Nagarajan SK, Ho MS (2023) Ag-Decorated Vertically Aligned ZnO Nanorods for Non-Enzymatic Glucose Sensor Applications. Nanomaterials, 13: 754. doi: 10.3390/nano13040754
  9. Myndrul V, Coy E, Babayevska N, Zahorodna V, Balitskyi V, Baginskiy I, Gogotsi O, Bechelany M, Giardi MT, Iatsunskyi I. (2022). MXene nanoflakes decorating ZnO tetrapods for enhanced performance of skin-attachable stretchable enzymatic electrochemical glucose sensor. Biosens Bioelectron, 207: 114141. doi: 10.1016/j.bios.2022.114141
  10. Ozdemir C, Yeni F, Odaci D, Timur S. (2010) Electrochemical glucose biosensing by pyranose oxidase immobilized in gold nanoparticle-polyaniline/AgCl/gelatin nanocomposite matrix, Food Chem, 119(1): 380-385 doi: 10.1016/j.foodchem.2009.05.087
  11. Ozyilmaz G, Ozyilmaz AT, Can F. (2011). Glucose Oxidase- Polypyrrole Electrodes Synthesized In P-Toluenesulfonic Acid And Sodium p-Toluenesulfonate, Appl Biochem Microbiol 47:217-225
  12. Ozyilmaz G, Ozyilmaz AT, Akyürekoğlu RH. (2017) Poly (N-Methylpyrrole)-Chitosan layers for Glucose Oxidase İmmobilization for Amperometric Glucose Biosensor Design, Natural Eng Sci, 2(3):123-134. doi: 10.28978/nesciences.354825
  13. Ozyilmaz G, Ozyilmaz AT, Ağçam S. (2018) Using Response Surface Methodology for Amperometric Glucose Biosensor Construction, Natural Eng Sci. 3(1):1-15 doi: 10.28978/nesciences.379311
  14. Ozyilmaz G, Ozyilmaz AT, Bayram E, Akyürekoğlu RH (2019). Amperometric glucose biosensor based on homopolymer-chitosan double layered glucose oxidase electrode modified with zinc oxide nanoparticles. Acta Chimica Slovenica, 66: 950-957
  15. Raicopol M, Pruna A, Damian C, Pilan L. (2013). Functionalized single-walled carbon nanotubes/polypyrrole composites for amperometric glucose biosensors, Nanoscale Res Lett, 8, 1-8
  16. Tang W, Li L, Zeng X, (2015). A glucose biosensor based on the synergistic action of nanometer-sized TiO2 and polyaniline, Talanta, 131: 417–423
  17. Yağiz E, Ozyilmaz G, Ozyilmaz AT (2022) Optimization of Graphite-mineral oil ratio with response surface methodology in glucose oxidase based carbon paste electrode design, Natural Eng Sci, 7: 22-23. doi: 10.28978/nesciences.1098655
  18. Yağiz E, Ozyilmaz G, Ozyilmaz AT (2024) Response Surface Methodology use in construction of polianiline polianiline-coated carbon paste electrode electrode–based biosensor: Modification and characterization. Biotechnol Appl Biochem 71:147–161. doi: 10.1002/bab.2528