HbA2 ölçümü için BioRad D-10TM ve Tosoh HLC 723 G8 HPLC sistemlerinin karşılaştırılması

Comparison of the BioRad D-10TM and Tosoh HLC 723 G8 HPLC instruments for the detection of HbA2

Güzin Aykal(Antalya Eğitim ve Araştırma Hastanesi, Biyokimya Lab, Antalya)
Fazıla Atakan Erkal(Antalya Halk Sağlığı Laboratuvarı, Biyokimya Lab, Antalya)
Ayşenur Yeğin(Antalya Eğitim ve Araştırma Hastanesi, Biyokimya Lab, Antalya)
Esin Eren(Antalya Eğitim ve Araştırma Hastanesi, Biyokimya Lab, Antalya)
Necat Yılmaz(Antalya Eğitim ve Araştırma Hastanesi, Biyokimya Lab, Antalya)
Sayfa
245–252
Yayın Tarihi
2016-09-01
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

Bulgular: Analizler sonucunda Tosoh HLC 723 G8 cihazında örneklerin 14’ü %46.66 ve BioRad D-10TM cihazında örneklerin dokuz’u %30 β -talasemi taşıyıcısı olarak saptandı.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Amaç: Özellikle Akdeniz bölgesinde önemli bir halk sağlığı sorunu olan hemoglobin bozuklukları taramasında hemoglobin sub tiplerini ayırt etmek amacıyla en yaygın kullanılan yöntem HPLC High Performance Liquid Chromatography ’dir.
📊
Temel Bulgular & Veri:Bu çalışmanın amacı; D-10TM Hemoglobin testingsystem BioRad,Hercules,USA marka HPLC cihazı ile Tosoh HLC 723 G8 TosohBioscience, Japan HPLC cihazını hemoglobinopati tanı ve taraması açısından önemli olan HbA2 parametresi ölçümü bakımından karşılaştırılmasıdır.
💡
Sonuç & Katkı:Bulgular: Analizler sonucunda Tosoh HLC 723 G8 cihazında örneklerin 14’ü %46.66 ve BioRad D-10TM cihazında örneklerin dokuz’u %30 β -talasemi taşıyıcısı olarak saptandı.
🔬Tür:Laboratuvar & Saha Deneyi
📍Kapsam:2016 Verileri
⚙️Değişkenler:hemoglobinopati · HPLC · HbA2

Özet

Amaç: Özellikle Akdeniz bölgesinde önemli bir halk sağlığı sorunu olan hemoglobin bozuklukları taramasında hemoglobin sub tiplerini ayırt etmek amacıyla en yaygın kullanılan yöntem HPLC High Performance Liquid Chromatography ’dir. Pek çok marka ve model HPLC sistemleri ile hemoglobinopati taraması yapılmakta ancak sistemlerin birbirleri ile uyumu bilinmemektedir. Bu çalışmanın amacı; D-10TM Hemoglobin testingsystem BioRad,Hercules,USA marka HPLC cihazı ile Tosoh HLC 723 G8 TosohBioscience, Japan HPLC cihazını hemoglobinopati tanı ve taraması açısından önemli olan HbA2 parametresi ölçümü bakımından karşılaştırılmasıdır. Yöntemler: Çalışmaya Antalya Eğitim ve Araştırma Hastanesi Merkez Laboratuvarı Biyokimya Bölümüne 1-31 Ocak 2016 tarihleri arasında talasemi taraması için başvuran 12’si erkek, 18’i kadın toplam 30 kişi alındı. Her iki cihazda tek seferde çalışılan hemoglobin A2 HbA2 değerlerinin farkları için bağımlı iki grup ve iki ölçüm fark testleri, istatistiksel uyum ve ilişki analizi için iki-yönlü rastgele sınıf içi korelasyon katsayısı SKK , Cohen’s kappa katsayısı, korelasyon katsayısı r ve Bland Altman Analizi yapıldı. Sonuçlar uygun grafiklerle sunuldu. Bulgular: Analizler sonucunda Tosoh HLC 723 G8 cihazında örneklerin 14’ü %46.66 ve BioRad D-10TM cihazında örneklerin dokuz’u %30 β -talasemi taşıyıcısı olarak saptandı. Her iki cihazda ölçülen HbA2 değerleri arasında anlamlı fark saptandı z= -4.487; p

Abstract

Objective: HPLC High Performance Liquid Chromatography is the most commonly used method to differentiate the subtypes of hemoglobin in screening for hemoglobinopathy which is a major public health care problem especially in Mediterranean region. Today, different models and trademarks of HPLC equipment have been used in screening for hemoglobinopathy. However, the agreement of the results of the analyses performed by these types of equipment are unknown. In this study, the aim was to investigate the agreement between the results of the analyses of HbA2 values with D-10TM Hemoglobin testing system BioRad,Hercules,USA and Tosoh HLC 723 G8 TosohBioscience, Japan trademark. Methods: A total of 30 individuals, 12 males and 18 females, applied to the Antalya Training and Research Hospital laboratory between January 1 and January 31, 2016 for thalassemia screening were included to this study. Statistical correlations of hemoglobin A2 HbA2 values, analyzed in each analyzer as a single analysis from each sample, were evaluated using two-way random intraclass correlation coefficient ICC , Cohen’s kappa coefficient, correlation coefficient r and Bland Altman analysis. Results are presented with graphics. Results: As a result of the analysis, 14 of the samples 46.66% were detected as – thalassemia carrier with Tosoh HLC 723 G8 system and nine of the samples 30% were detected as – thalassemia carrier with BioRad D-10TM system. There was significant differance in HbA2 values measured with both systems z= -4.487;p

🔖

Bu Makaleye Atıf Yap

Aykal, G., Erkal, F. A., Yeğin, A., Eren, E., Yılmaz, N. (2016). HbA2 ölçümü için BioRad D-10TM ve Tosoh HLC 723 G8 HPLC sistemlerinin karşılaştırılması. Türk Hijyen ve Deneysel Biyoloji Dergisi, Cilt: 73 Sayı: 3, 245–252. https://ajindex.com/articles/1322026

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (16)

  1. Alphabetical hemoglobin variants list. Hemoglobin, 1996; 20(3): 313–35.
  2. Gosselin RC, CarlinAC, Dwyre DM. Comparison of the BioRad Variant and Primus Ultra2 high-pressure liquid chromatography (HPLC) instruments for the detection of variant hemoglobins. Int Jnl Lab Hem, 2011; 33: 159–67
  3. Sönmez Ç, Kaymak AÖ, Güntaş G.Halk sağlığı problemi olan talasemilerde laboratuvar. Turk Hij Den Biyol Derg, 2014; 71(4): 221–28.
  4. Ellidağ HY, Eren E, Tosun SA, Yılmaz N. Talasemi taramasında Primus Ultra2 Analyzer ve Tosoh HLC723 G8 HPLC sistemlerinin karşılaştırılması. Türk Klinik Biyokimya Derg, 2013; 11(3): 99-104.
  5. Clarke GM, Higgins TN. Laboratory investigation of hemoglobinopathies and thalassemias: Review and update. Clin Chem, 2000; 46: 1284–90.
  6. Rund D, Rachmilewitz E. B-Thalassemia. N Engl J Med, 2005; 353: 1135-46.
  7. Çatak B, Sütlü S, Kılınç S, Badıllıoğlu O, Canatan D. Burdur’da ilköğretim 8. sınıflarda β-talasemi taşıyıcılık sıklığı. Cumhuriyet Med J, 2013; 35: 193-8.
  8. Paleari R, Gulbis B, Cotton F, Mosca A. Inter laboratory comparison of current high-performance methods for HbA2. Int Jnl Lab Hem, 2012; 34: 362–8
  9. Greene DN, Pyle AL, Chang JS, Hoke C, Lorey T. Comparison of Sebia Capillarys Flex capillary electrophoresis with the Bio Rad Variant II high pressure liquid chromatography in the evaluation of hemoglobinopathies. Clin Chim Acta, 2012; 413: 1232–8
  10. Anonymous. The laboratory diagnosis of haemoglobinopathies, Br J Haematol, 1998; 101(4): 783-92.
  11. Lawrence I, Lin K. A concordance correlation coefficient to evaluate reproducibility. Biometrics, 1989; 45(1): 255-68.
  12. Statistical Calculators, Lin'sConcordance: http:// www.niwa.co.nz/online-services/statisticalcalculators/concordance.(31.1.2016)
  13. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet, 1986; I: 307-10.
  14. Viera AJ, Garrett JM. Understanding inter observer agreement: the kappa statistic. Family Medicine, 2005; 37(6): 360-63.
  15. Fleiss JL. The design and analysis of clinical experiments. New York, John Wiley and Sons, 1986: 1-32.
  16. A Good Example. http://www.medical.siemens. com/siemens/nl_NLDIAG/gg_diag_FBAs/files/ EP_Evaluator/Webinars/02-EE9-Mods-FAll_2010. pdf (15.1.2016)