YOĞUN BAKIM ÜNİTELERİNDE MEKANİK VENTİLASYON ÇIKIŞINDAN KAYNAKLANAN VİRÜS YAYILIMININ DEĞERLENDİRİLMESİ VE UV-C STERİLİZASYONUNUN ETKİSİ

EVALUATION OF VIRAL DISPERSION FROM MECHANICAL VENTILATION OUTLETS IN INTENSIVE CARE UNITS AND THE EFFECTIVENESS OF UV-C STERILIZATION

Elif Kağa(Afyonkarahisar Sağlık Bilimleri Üniversitesi, Tıbbi Hizmetler ve Teknikler Bölümü)0000-0002-2279-6105
Sadık Kağa(Afyon Kocatepe Üniversitesi, Biyomedikal Mühendisliği Bölümü)0000-0002-6303-7981
Ugur Fidan(Afyon Kocatepe Üniversitesi, Biyomedikal Mühendisliği Bölümü)0000-0003-0356-017X
Ahmet Murat Koyuncu(Afyonkarahisar Devlet Hastanesi, Genel Cerrahi AB.)0000-0002-4331-2704
Gizem Fatma Ergüner(Afyon Kocatepe Üniversitesi, Biyomedikal Mühendisliği Bölümü)0000-0002-6436-6891
Abdullah Yalçın(Afyonkarahisar Devlet Hastanesi, Anesteziyoloji ve Reanimasyon AD.)0000-0001-7732-8213
Sayfa
233–239
Yayın Tarihi
2025-07-16
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

BULGULAR: Aşılama testinden elde edilen verilere göre, ‘UV açık’ ve ‘UV kapalı’ modlarında kullanılan filtrelerden alınan örneklerde virüs üremesi görülmedi.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:AMAÇ: Bu çalışmanın amacı, yoğun bakım servislerinde, mekanik ventilasyon cihazlarından kaynaklı aerosol viral yayılımın test edilmesi, virus canlılık parametrelerinin incelenmesi ve UV-C sterilizasyonun etkisinin araştırılmasıdır.
📊
Temel Bulgular & Veri:GEREÇ VE YÖNTEM: COVID-19 tanısı almış 14 adet gönüllü hastanın bağlı olduğu mekanik ventilasyon cihazı çıkış hattına UV-C sterilizatör entegre edilmiştir.
💡
Sonuç & Katkı:BULGULAR: Aşılama testinden elde edilen verilere göre, ‘UV açık’ ve ‘UV kapalı’ modlarında kullanılan filtrelerden alınan örneklerde virüs üremesi görülmedi.
🔬Tür:Laboratuvar & Saha Deneyi
📍Kapsam:14 adet
⚙️Değişkenler:Yoğun Bakım · Ventilasyon · COVID-19

Özet

AMAÇ: Bu çalışmanın amacı, yoğun bakım servislerinde, mekanik ventilasyon cihazlarından kaynaklı aerosol viral yayılımın test edilmesi, virus canlılık parametrelerinin incelenmesi ve UV-C sterilizasyonun etkisinin araştırılmasıdır. GEREÇ VE YÖNTEM: COVID-19 tanısı almış 14 adet gönüllü hastanın bağlı olduğu mekanik ventilasyon cihazı çıkış hattına UV-C sterilizatör entegre edilmiştir. Hasta kaynaklı viral bulaşın tespiti için steril şartlarda filtre örnekleri toplanmıştır. Virolojik Teşhis Laboratuvarı’nda virüs inokülasyonu ve RT-PCR testi yapılmıştır. BULGULAR: Aşılama testinden elde edilen verilere göre, ‘UV açık’ ve ‘UV kapalı’ modlarında kullanılan filtrelerden alınan örneklerde virüs üremesi görülmedi. Bununla beraber hastaların sürüntü örneklerinde, PCR testine göre COVID-19 pozitif çıksa bile, mekanik ventilasyon çıkış hattından elde edilen örneklerde PCR sonucu negatif olarak tespit edilmiştir. SONUÇ: Çalışma kapsamında COVID-19 tanısı almış 14 gönüllü hastanın bağlı olduğu mekanik ventilasyon cihazının çıkış hattından aerosol virüs tespiti ve aktivite testleri yapıldı. Çalışmada mekanik ventilasyon alan hastaların solunumlarından aerosol virüsünün ortam havasına yayılma riskinin düşük olduğu belirlendi.

Abstract

OBJECTIVE: This study aimed to test aerosol-based viral dispersion originating from mechanical ventilation devices in intensive care units, examine virus viability parameters, and investigate the effect of UV-C sterilization. MATERIAL AND METHODS: A UV-C sterilizer was integrated into the mechanical ventilation device outlet line connected to 14 volunteer patients diagnosed with COVID-19. Filter samples were collected under sterile conditions to detect patient-related viral contamination. Viral inoculation and RT-PCR tests were performed in the Virological Diagnostic Laboratory. RESULTS: According to the data from the inoculation test, no virus growth occurred in the samples obtained from the filters used in the ‘UV on’ and ‘UV off’ modes. However, even if COVID-19 was positive according to the PCR test in the patients' swab samples, the PCR result was detected as negative in the samples obtained from the outlet line of the mechanical ventilation. CONCLUSIONS: Within the scope of the study, aerosol virus detection and activity tests were performed from the outlet line of the mechanical ventilation device to which 14 volunteer patients diagnosed with COVID-19 were connected. The findings suggest that the risk of aerosolized virus spreading into the ambient air from patients receiving mechanical ventilation is low.

🔖

Bu Makaleye Atıf Yap

Kağa, E., Kağa, S., Fidan, U., Koyuncu, A. M., Ergüner, G. F., Yalçın, A. (2025). YOĞUN BAKIM ÜNİTELERİNDE MEKANİK VENTİLASYON ÇIKIŞINDAN KAYNAKLANAN VİRÜS YAYILIMININ DEĞERLENDİRİLMESİ VE UV-C STERİLİZASYONUNUN ETKİSİ. Kocatepe Tıp Dergisi, Cilt: 26 Sayı: 3, 233–239. https://doi.org/10.18229/kocatepetip.1637686

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (33)

  1. Li X, Lester D, Rosengarten G, et al. A spatiotemporally resolved infection risk model for airborne transmission of COVID-19 variants in indoor spaces. Science of the Total Environment. 2022;812:152592.
  2. Fraser V J, Johnson K, Primack J, et al. Evaluation of rooms with negative pressure ventilation used for respiratory isolation in seven midwestern hospitals. Infection Control & Hospital Epidemiology. 1993;14(11):623-8.
  3. Vincent J L, Rello J, Marshall J, et al. International study of the prevalence and outcomes of infection in intensive care units. Jama. 2009;302(21):2323-9.
  4. Vogelaers D, De Bels D, Forêt F, et al. ANTHICUS Study Investigators. Patterns of antimicrobial therapy in severe nosocomial infections: empiric choices, proportion of appropriate therapy, and adaptation rates a multicentre, observational survey in critically ill patients. International Journal of Antimicrobial Agents. 2010;35(4): 75-81.
  5. Grasselli G, Zangrillo A, Zanella A, et al. Baseline characteristics and outcomes of 1591 patients infected with SARS-CoV-2 admitted to ICUs of the Lombardy Region, Italy. Jama. 2020;323(16):1574-81.
  6. Keane S, Martin-Loeches I. Host-pathogen interaction during mechanical ventilation: systemic or compartmentalized response? Critical Care. 2019;23:134.
  7. Tran K, Cimon K, Severn M, et al. Aerosol generating procedures and risk of transmission of acute respiratory infections to healthcare workers: a systematic review. PloS one. 2012;7(4):35797.
  8. Heuer J F, Crozier T A, Howard G, et al. Can breathing circuit filters help prevent the spread of influenza A (H1N1) virus from intubated patients?. GMS hygiene and infection control. 2013;8(1):Doc09.
  9. Rhee C, Baker M A, Klompas M. Prevention of SARS-CoV-2 and respiratory viral infections in healthcare settings: current and emerging concepts. Current Opinion in Infectious Diseases. 2022;35(4):353-62.
  10. Casini B, Tuvo B, Cristina M L, et al. Evaluation of an ultraviolet C (UVC) light-emitting device for disinfection of high touch surfaces in hospital critical areas. International journal of environmental research and public health. 2019;16(19):3572.
  11. Yin R, Dai T, Avci P, et al. Light based anti-infectives: ultraviolet C irradiation, photodynamic therapy, blue light, and beyond. Current Opinion in Pharmacology. 2013;13(5):731-62.
  12. Kowalski W. Ultraviolet germicidal irradiation handbook: UVGI for air and surface disinfection.New York: Springer science & Business Media; 2010.
  13. Van Doremalen N, Bushmaker T, Morris D H, et al. Aerosol and surface stability of SARS-CoV-2 as compared with SARS-CoV-1. New England Journal of Medicine. 2020;382(16):1564-7.
  14. Chang K. Scientists consider indoor ultraviolet light to zap coronavirus in the air. The New York Times Company. 2020.
  15. Kocamemi B A, Kurt H, Hacıoglu S, Yaralı C, Saatci A M, Pakdemirli B. First data-set on SARS-CoV-2 detection for Istanbul wastewaters in Turkey. MedRxiv. 2020;5(6):1-11.
  16. Corman V M, Landt O, Kaiser M, et al. Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR. Eurosurveillance. 2020;25(3):2000045.
  17. Beaussier M, Vanoli E, Zadegan F, et al. Aerodynamic analysis of hospital ventilation according to seasonal variations. A simulation approach to prevent airborne viral transmission pathway during Covid-19 pandemic. Environment international. 2022;158:106872.
  18. Goodfellow H D, Kosonen R, editors. Industrial Ventilation Design Guidebook: Volume 1: Fundamentals. Academic Press. 2020.
  19. Meng L, Qiu H, Wan L et al. Intubation and ventilation amid the COVID-19 outbreak: Wuhan’s experience. Anesthesiology. 2020;10-97.
  20. Renee LC, Matos I, Kevin K. Er al. COVID-19 PRACTICE MANAGEMENT GUIDE Clinical Management of COVID-19, 2020:3-23.
  21. Hamner L. High SARS-CoV-2 attack rate following exposure at a choir practice Skagit County, Washington, March 2020. MMWR. Morbidity and mortality weekly report. 2020;69.
  22. Morawska L, Cao J. Airborne transmission of SARS-CoV-2: The world should face the reality. Environment international. 2020;139:105730.
  23. WHO Considerations for quarantine of individuals in the context of containment for coronavirus disease (COVID-19). March 2020.
  24. U.S. PUBLIC HEALTH SERVICE COMMISSIONED CORPS. Optimizing Ventilator Use during the COVID-19 Pandemic. March 2020.
  25. Vasilyev AI, Kostyuchenko SV, Kudryavtsev NN, Sobur D A, Sokolov DV. UV disinfection technologies for water, air and surface treatment. Light Eng. 2018;26(1):25-31.
  26. Casini B, Tuvo B, Cristina ML, et al. Evaluation of an ultraviolet C (UVC) light-emitting device for disinfection of high touch surfaces in hospital critical areas. International Journal of Environmental Research and Public Health. 2019;16(19):3572.
  27. Wurtmann EJ, Wolin SL. RNA under attack: cellular handling of RNA damage. Critical Reviews in Biochemistry and Molecular Biology. 2009;44(1):34-49.
  28. Chiappa F, Frascella B, Vigezzi GP, et al. The efficacy of ultraviolet light-emitting technology against coronaviruses: a systematic review. J Hosp Infect. 2021;114:63-78.
  29. Trevisan A, Piovesan S, Leonardi A, et al. Unusual high exposure to ultraviolet‐C radiation. Photochemistry and Photobiology. 2006;82(4):1077-9.
  30. Lednicky JA, Lauzard M, Fan ZH, et al. Viable SARS-CoV-2 in the air of a hospital room with COVID-19 patients. International Journal of Infectious Diseases. 2020;100:476-82.
  31. Su WL, Hung PP, Lin CP, et al. Masks and closed-loop ventilators prevent environmental contamination by COVID-19 patients in negative-pressure environments. Journal of Microbiology, Immunology and Infection. 2021;54(1):81-4.
  32. Altındis M, Aktepe O C, Cetinkaya Z. Determination if Methicillin Resistant Staphylococcus Aureus in Intensive. Kocatepe Tıp Dergisi. 2006;7: 43-6.
  33. Ahn J Y, An S, Sohn Y, et al. Environmental contamination in the isolation rooms of COVID-19 patients with severe pneumonia requiring mechanical ventilation or high-flow oxygen therapy. Journal of Hospital Infection. 2020;106(3):570-6.