QUANTITATIVE TOTAL IRON ANALYSIS USING XRD - FLUORESCENCE RADIATION INTENSITY

Sayfa
64–64
Yayın Tarihi
1990-04-01
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

In addition, another important result of this study is that the radiation intensity have direct relation with the total iron content in the sample, regardless of iron combination with other elements.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Fluorescence radiation is being known as "unwanted incident", while XRD determination, and various methods arc in use to solve this secondary radiation.
📊
Temel Bulgular & Veri:It takes only 40 seconds to find the quantity of total iron in the analysed sample with the suggested method.
💡
Sonuç & Katkı:In addition, another important result of this study is that the radiation intensity have direct relation with the total iron content in the sample, regardless of iron combination with other elements.
🔬Tür:Araştırma Makalesi
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Alan Parametreleri

Abstract

Fluorescence radiation is being known as "unwanted incident", while XRD determination, and various methods arc in use to solve this secondary radiation. In this study, fluorescence radiation is especially created in order to show the possibility of quantitative analysis with the help of this secondary radiation. Some selected standards, which have different quantity of iron content, were chosen and Cutarget was intentionaly used during XRD determination, in order to cause fluorescence radiation in different intensity. Finally the positive relation was detected between the total iron content and fluorescence radiation intensity. It takes only 40 seconds to find the quantity of total iron in the analysed sample with the suggested method. In addition, another important result of this study is that the radiation intensity have direct relation with the total iron content in the sample, regardless of iron combination with other elements.

🔖

Bu Makaleye Atıf Yap

Aydal, D. (1990). QUANTITATIVE TOTAL IRON ANALYSIS USING XRD - FLUORESCENCE RADIATION INTENSITY. Maden Tetkik ve Arama Dergisi, Cilt: 111 Sayı: 111, 64–64. https://ajindex.com/articles/1120312

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (8)

  1. Graham, A.R., 1969, Quantitative determination of hexagonal and monoclinic pyrrotite by X—ray diffraction: Can.Mineralogist, 10,4-24.
  2. Gulbrandsen, R.A., 1969, A method of X-ray analysis for determining the ratio of calcite to dolomite in mineral mixture: US, Geol.Surv.Bull., 1111-D, 147-152.
  3. Gunatilaka, H.A. ve Till, R. 1971, A precise and accurate method for the quantitative determination of carbonate minerals by X—ray diffraction using a spiking technique: Mineral Mag., 38, 481-487.
  4. Hutchinson, C.S., 1976, Laboratory handbook of Petrografic techniques.: A Wiley-lnterscience Publication, John Wiley and Sons, Nevv York, London, Sydney, Toronto.
  5. Johnson, W. ve Andrevvs, K.W., 1962, X-ray examination of aluminosilicatcs:Trans.Brit.Ceram.Soc., 61, 724.
  6. Klug, H.P. ve Alexandre, L., 1973, X-ray diffraction procedures: John Wiley and Sons Inc., Nevv York, London.
  7. Norrish, K. ve Taylor, R.M., 1962, Quantitative analysis by X-ray diffraction: Clay Minerals Bull., 5, 98-109.
  8. Schoen, R., 1962, Semi-quantitative analysis of chlorites by X—ray diffraction: Am.Mineralogist, 47, 1384.