Biplan ve Monoplan Serebral Anjiyografide Radyasyon Maruziyetinin Karşılaştırmalı Analizi: Tek Merkezli Retrospektif Çalışma

COMPARATIVE ANALYSIS OF RADIATION EXPOSURE IN BIPLANE VERSUS MONOPLANE CEREBRAL ANGIOGRAPHY: A SINGLE-CENTER RETROSPECTIVE STUDY

Hakan Ayyıldız(SAĞLIK BİLİMLERİ ÜNİVERSİTESİ, İSTANBUL BAŞAKŞEHİR ÇAM VE SAKURA ŞEHİR SAĞLIK UYGULAMA VE ARAŞTIRMA MERKEZİ, DAHİLİ TIP BİLİMLERİ BÖLÜMÜ, RADYOLOJİ ANABİLİM DALI)0000-0003-0244-7765
Mücahit Mustafa Karabulut(YOZGAT BOZOK ÜNİVERSİTESİ, TIP FAKÜLTESİ, DAHİLİ TIP BİLİMLERİ BÖLÜMÜ, RADYOLOJİ ANABİLİM DALI)
Berke Ersoy(İSTANBUL ÜNİVERSİTESİ, ONKOLOJİ ENSTİTÜSÜ)0000-0003-0161-1914
Mehmet Barburoğlu(İSTANBUL ÜNİVERSİTESİ, İSTANBUL TIP FAKÜLTESİ, DAHİLİ TIP BİLİMLERİ BÖLÜMÜ, RADYODİAGNOSTİK ANABİLİM DALI)0000-0002-8715-1893
Sayfa
199–204
Yayın Tarihi
2026-08-14
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

Bulgular: Monoplan ve biplan grupları arasında floroskopi süresi benzerdi (p = 0.736).

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Amaç: Tanısal serebral anjiyografi sırasında biplan ve monoplan dijital çıkarma anjiyografi (DSA) sistemleri arasında hastaların maruz kaldığı radyasyon dozunu karşılaştırmak ve modern çift düzlemli teknoloji ile sağlanan doz etkinliğini değerlendirmek.
📊
Temel Bulgular & Veri:Gereç ve Yöntem: Bu retrospektif tek merkezli çalışma kapsamında Kasım 2021 – Nisan 2022 arasında tanısal serebral DSA uygulanan ardışık 61 hasta değerlendirildi.
💡
Sonuç & Katkı:Bulgular: Monoplan ve biplan grupları arasında floroskopi süresi benzerdi (p = 0.736).
🔬Tür:Deneysel Tarla Denemesi
📍Kapsam:2021 · 61 hasta
⚙️Değişkenler:Radyasyon Dozu · Anjiyografi · Floroskopi

Özet

Amaç: Tanısal serebral anjiyografi sırasında biplan ve monoplan dijital çıkarma anjiyografi (DSA) sistemleri arasında hastaların maruz kaldığı radyasyon dozunu karşılaştırmak ve modern çift düzlemli teknoloji ile sağlanan doz etkinliğini değerlendirmek. Gereç ve Yöntem: Bu retrospektif tek merkezli çalışma kapsamında Kasım 2021 – Nisan 2022 arasında tanısal serebral DSA uygulanan ardışık 61 hasta değerlendirildi. İşlemler, Siemens Artis Zee monoplan sistemi (n = 31) ve Philips Azurion ClarityIQ biplan sistemi (n = 30) kullanılarak gerçekleştirildi. Her işlem için toplam floroskopi süresi, floroskopi doz-alan çarpımı (DAP), ekspojur DAP, toplam DAP ve kümülatif hava kerma değerleri kaydedildi. Gruplar bağımsız örneklem t-testi veya Mann–Whitney U testi ile karşılaştırıldı; p < 0.05 anlamlı kabul edildi. Bulgular: Monoplan ve biplan grupları arasında floroskopi süresi benzerdi (p = 0.736). Biplan sistem daha fazla sayıda görüntü elde etmesine rağmen (p < 0.001), anlamlı derecede daha düşük radyasyon dozu sağladı. Medyan ekspojur DAP ve toplam DAP değerleri biplan grupta yaklaşık %25–35 oranında azalmıştı ve toplam hava kerma belirgin olarak daha düşüktü (p < 0.001). Floroskopi DAP değeri biplan sistemde hafifçe yüksek bulundu (p = 0.008); bunun zaman zaman çift düzlem kullanımına bağlı olduğu düşünüldü. Sonuç: Daha fazla projeksiyon elde edilmesine rağmen biplan DSA sistemi, monoplan sisteme kıyasla anlamlı derecede daha düşük toplam radyasyon maruziyeti sağlamıştır. Bu bulgular, gelişmiş doz azaltma algoritmalarına (ClarityIQ) sahip modern biplan teknolojisinin tanısal serebral anjiyografide daha etkili doz yönetimi ve artmış hasta güvenliği sunduğunu göstermektedir.

Abstract

Purpose: To compare patient radiation exposure between biplane and monoplane digital subtraction angiography (DSA) systems during diagnostic cerebral angiography and evaluate the dose-efficiency of modern dual-plane technology. Material-Methods: This retrospective single-center study included 61 consecutive patients who underwent diagnostic cerebral DSA between November 2021 and April 2022. Examinations were performed using a Siemens Artis Zee monoplane system (n = 31) and a Philips Azurion ClarityIQ biplane system (n = 30). For each procedure, total fluoroscopy time, fluoroscopy dose-area product (DAP), exposure DAP, total DAP, and cumulative air kerma were recorded. Groups were compared using independent-samples t-tests or Mann–Whitney U tests, with p < 0.05 considered statistically significant. Results: Fluoroscopy time was comparable between the monoplane and biplane groups (p = 0.736). The biplane system produced a significantly greater number of image acquisitions (p < 0.001) yet achieved markedly lower radiation doses. Median exposure DAP and total DAP were reduced by approximately 25–35%, and total air kerma was significantly lower in the biplane group (p < 0.001). Fluoroscopy DAP was slightly higher for the biplane system (p = 0.008), likely reflecting occasional dual-plane use during complex catheter navigation. Conclusion: Despite acquiring more projections, the biplane DSA system demonstrated substantially lower total radiation exposure compared with the monoplane system. These results indicate that modern biplane technology with advanced dose-reduction algorithms (ClarityIQ) offers improved dose efficiency and enhanced patient safety in diagnostic cerebral angiography.

🔖

Bu Makaleye Atıf Yap

Ayyıldız, H., Karabulut, M. M., Ersoy, B., Barburoğlu, M. (2026). Biplan ve Monoplan Serebral Anjiyografide Radyasyon Maruziyetinin Karşılaştırmalı Analizi: Tek Merkezli Retrospektif Çalışma. İstanbul Tıp Fakültesi Dergisi, Cilt: 89 Sayı: 3, 199–204. https://ajindex.com/articles/1113270

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (16)

  1. Crummy AB, Strother CM, Mistretta CA. The history of digital subtraction angiography. J Vasc Interv Radiol 2018;29(8):1138-41. google scholar
  2. Summerlin D, Willis J, Boggs R, Johnson LM, Porter KK. Radiation dose reduction opportunities in vascular imaging. Tomography 2022;8(5):2618-2638. google scholar
  3. Suh SH. Striving to improve image quality and minimize radiation dose in neurointerventional procedures. Neurointervention 2025;20(2):49-51. google scholar
  4. Stephenson TF. Non-isocentric applications of biplane digital subtraction angiography. Comput Radiol 1985;9(6):399-404. google scholar
  5. Sadick V, Trinh L, Fernandes G, Pau P, Spiteri M, Vu LA, et al. Does biplane imaging reduce contrast load, procedural and screening time compared to single-plane imaging in routine diagnostic coronary angiography? Radiographer 2008;55(1):20-4. google scholar
  6. Andres C, Agulla M, Torres R, Perez-Garcia H, del Castillo A, Miguel D, et al. Patient doses in a biplane angiography system. Physica Medica 2016;32(3):307 google scholar
  7. van Asch CJ, Velthuis BK, Rinkel GJ, Algra A, de Kort GA, Witkamp TD, et al. Diagnostic yield and accuracy of CT angiography, MR angiography, and digital subtraction angiography for detection of macrovascular causes of intracerebral haemorrhage: prospective, multicentre cohort study. BMJ 2015;351:h5762. google scholar
  8. Nickoloff EL, Lu ZF, Dutta AK, So JC. Radiation dose descriptors: BERT, COD, DAP, and other strange creatures. Radiographics 2008;28(5):1439-50. google scholar
  9. The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP 2007;37(2-4):1-332. google scholar
  10. Huda W. Kerma-area product in diagnostic radiology. AJR Am J Roentgenol 2014;203(6):W565-9. google scholar
  11. Hamada N, Fujimichi Y. Classification of radiation effects for dose limitation purposes: history, current situation and future prospects. J Radiat Res 2014;55(4):629-40. google scholar
  12. Vañó E, Miller DL, Martin CJ, Rehani MM, Kang K, Rosenstein M, et al. ICRP publication 135: Diagnostic reference levels in medical imaging. Ann ICRP 2017;46(1):1-144. google scholar
  13. Johnson, Coley SJ, Kyrion J, Taylor WJ. Comparing the performance of mono- and biplane fluoroscopy systems in diagnostic and interventional neuroangiography using the dose- area product. Neuroradiology 2001;43(9):728-34. google scholar
  14. Guenego A, Mosimann PJ, Pereira VM, Zuber K, Lotterie JA, Tomas D, et al. Comparison of mono versus biplane performance and factors associated with higher radiation doses and contrast exposure during cerebrovascular mechanical thrombectomy, an international multi-centers study. Journal of Neuroradiology 2019;46(2):64-5. google scholar
  15. Roccatagliata L, Presilla S, Pravatà E, Cianfoni A. Radiation dose to the operator during fluoroscopically guided spine procedures. Neuroradiology 2017;59(9):885-92. google scholar
  16. Ortenzia O, Trojani V, Bertolini M, Nitrosi A, Iori M, Ghetti C. Radiation dose reduction and static image quality assessment using a channelized hotelling observer on an angiography system upgraded with clarity IQ. Biomed Phys Eng Express 2020;6(2):02500 google scholar