Journals / Türk Oftalmoloji Dergisi / 2021 / Cilt: 51 - Sayı: 5
The Surgeon’s Role in Inducing and Controlling Motion Errors During Intraocular Membrane Peeling Procedures
- Journal
- Türk Oftalmoloji Dergisi
- Pages
- 288–293
- DOI
- —
Abstract
Objectives: To understand the surgeon’s role in inducing and correcting movement inaccuracies during intraocular membrane peeling procedures. Materials and Methods: Optical sensors were used to record movement errors during actuation at the distal tip of 23-gauge pneumatic forceps both when the handle was handheld and when fixed with no human contact. Movements were also recorded at the proximal part of the forceps shaft (near the sclerotomy site) and compared to movement recorded at the distal end. The root mean square (RMS) and range values of the signals obtained from the sensors were calculated before and after applying high (7-13 Hz) and low (
Özet
Amaç: Göz içi membran soyma prosedürleri sırasında meydana gelen hareket hatalarının indükleme ve düzeltilmesinde cerrahın rolünü anlamak. Gereç ve Yöntem: 23-gauge pnömatik forsepsin distal ucundaki hareket hatalarını kayıt etmek için optik yansıtıcı sensörler kullanıldı. Hareket hataları, forseps sapı hem elde tutulduğunda hem de insan teması olmadan sabitlendiğinde kaydedildi. Ayrıca hareket hataları forseps şaftının hem proksimal kısmında (sklerotomi bölgesine yakın) hem de distal kısmında kaydedildi ve sonuçları karşılaştırıldı. Optik sensörlerden elde edilen sinyallarin karekök ortalaması (RMS) ve aralığı, yüksek (7-13 Hz) ve düşük (