Comparison of Primus Ultra2 Analyzer and Tosoh HLC-723 G8 HPLC Systems in Thalassemia Screening

Talasemi Taramasında Primus Ultra2 Analyzer ve Tosoh HLC-723 G8 HPLC Sistemlerinin Karşılaştırılması

Pages
99–104
DOI
—

Abstract

Objective: HPLC (High Performance Liquid Chromatography) is the most commonly used method to differentiate the subtypes of hemoglobin in screening for hemoglobinopathy. Today, different models and trademarks of HPLC equipment have been used in screening for hemoglobinopathy. However, the agreement of the results of the analyses performed by these types of equipment are unknown. In this study, the aim was to investigate the agreement between the results of the analyses of hemoglobinopathy screening performed in two sets of HPLC equipment with the Primus Ultra2 analyzer and the analyzer Tosoh HLC 723 G8 trademarks. Materials and Methods: A total of 40 individuals, 22 males and 18 females, presented to the laboratory between March 15 and April 15, 2013 for thalassemia screening were included in this prospective study. Statistical correlations of hemoglobin A2 (HbA2) values, analyzed in each analyzer as a single analysis from each sample were evaluated using student t test, two-way random intraclass correlation coefficient (ICC), concordance correlation coefficient, correlation coefficient (r) and regression analysis. Results: As a result of the analysis, 6 of the samples (33.7%) were detected as -thalassemia carrier, 10 (36.2%) were diagnosed as anemic and 24 (30%) were detected to be normal. HbA2 values were not significantly different between the two HPLC equipments (t= -1.269 p= 0.208). Correlation coefficients for HbA2 values were found to be as r= 0.987, ICC=0.97 and CCC= 0.93. Regression analysis for HbA2 values were found to be as Y= -0.0605+1.1613X. Conclusion: According to the regression, correlation and other statistics analysis, agreement between the results of HbA2 values measured in the two analyzers was found to be sufficient.

Özet

Amaç: Hemoglobinopati taraması için hemoglobin alt tiplerini ayırmada HPLC (High Performance Liquid Chromatography) yaygın kullanılan bir yöntemdir. Günümüzde değişik marka ve modelde HPLC sistemleri hemoglobinopati taraması için kullanılmakta fakat bu cihazların birbiri ile olan uyumu bilinmemektedir. Bu amaçla, biz bu çalışmada Primus Ultra2 analyzer marka HPLC cihazı ile Tosoh HLC 723 G8 marka HPLC cihazını hemoglobin A2 (HbA2) bakımından birbirleri ile olan uyumu araştırdık. Gereç ve Yöntem: Laboratuvarımıza 15 Mart -15 Nisan 2013 tarihleri arasında talasemi taraması için başvuran 22si erkek, 18i kadın toplam 40 kişi prospektif olarak çalışmaya alındı. Her iki cihazda tek seferde çalışılan HbA2 değerleri için istatistiksel uyum, student t test, korelasyon katsayısı (r), regresyon analizi, iki-yönlü rasgele sınıfiçi korelasyon katsayısı (SKK) ve Uyum korelasyon katsayısı (UKK) kullanılarak değerlendirildi. Bulgular: Analiz sonucunda örneklerin 6sı (%33,7) -talasemi taşıyıcısı, 10u (%36.2) anemik ve 24ü (%30) normal olarak saptandı. Her iki cihazda ölçülen HbA2 değerleri arasında anlamlı farklılık yoktu (t= -1.269 p= 0.208). HbA2 değerleri için korelasyon katsayısı r= 0.987, SKK=0.97, UKK= 0.93 ve regresyon analizinde ise Y = -0.0605 + 1.1613X olarak bulundu. Sonuç: Korelasyon regresyon ve diğer istatistiksel analizlere göre iki cihazın HbA2 değerleri arasındaki uyum iyi olarak bulundu.