Journals / Türk Kardiyoloji Derneği Arşivi / 2012 / Cilt: 40 - Sayı: 1
Cardiac pacemakers designed for magnetic resonance environment
- Pages
- 69–75
- DOI
- —
Abstract
Electromagnetic fields associated with magnetic resonance imaging (MRI) can cause serious functional disturbances in patients with cardiac pacemak- ers. Thus, an implanted pacemaker is generally consid- ered a contraindication for an MRI scan. Studies have shown that MRI indication may emerge in about 17% of patients within one year of pacemaker implantation, and up to 75% of these patients may require MRI during the lifetime of their device. Introduction of pacemaker systems designed specifically for the MRI environment is an important development, providing pacemaker patients with access to this important diagnostic modality. In this article, the current status of MR conditional pacemakers is discussed, together with clinical applications.
Özet
Manyetik rezonans (MR) görüntüleme sırasında oluşan elektromanyetik alanlar, kalp pili fonksiyonlarında ciddi ve tehlikeli bozukluklara yol açabilir. Bu nedenle, kalp pili bulunan hastalarda MR görüntüleme genellikle kontrendike olarak kabul edilmektedir. Çalışmalar, kalp pili takılan hastaların %17sinde bir yıl içinde MR görüntüleme endikasyonu oluştuğunu ve bu oranın takılmış cihazların süresi bitmeden %75e kadar çıkabileceğini göstermiştir. Son yıllarda, MR ortamı için geliştirilmiş pillerin klinik uygulamaya girmesi, kalp pili bulunan hastaların MR görüntülemeden yararlanmaları için atılmış önemli bir adımdır. Bu makalede MR ortamı için geliştirilmiş kalp pilleri ve klinik uygulamalarının bugünkü durumu değerlendirilmiştir.