Journals / Türk Kardiyoloji Derneği Arşivi / 2015 / Cilt: 43 - Sayı: 3
MitraClip cihazının beklenmeyen komplikasyonu: Cihaz kanatlarından birinin sol atriyumda yönlendirici kateter içerisinde sıkışması
- Pages
- 281–283
- DOI
- —
Abstract
Summary Percutaneous edge-to-edge mitral valve repair using the MitraClip (Abbot Vascular, USA) system is a promising technique for mitral regurgitation treatment in select high-risk surgical patients. Although the safety and efficacy of the technique have been demonstrated, a few complications of the MitraClip device have been reported. In this report, we present a rare complication that recently occurred during the performance of a MitraClip procedure in a patient with severe functional mitral regurgitation. One MitraClip arm got stuck inside the guide catheter in the left atrium and a decision was made to discontinue percutaneous intervention because the problem could not be resolved.
Özet
Perkütan uç uca mitral kapak tamiri yöntemi olarak MitraClip (Abbot Vascular, USA) sistemi, yüksek riskli seçilmiş hastalarda mitral yetersizliği tedavisi için umut vaad etmektedir. Bu tekniğin etkinliği ve güvenilirliği gösterilmiş olmakla birlikte, az sayıda da olsa MitraClip cihaz komplikasyonu bildirilmiştir. Biz bu yazıda ileri derece fonksiyonel mitral yetersizliği nedeniyle MitraClip işlemi yapılırken ortaya çıkan nadir bir komplikasyonu sunuyoruz. MitraClip cihazının kanatlarından biri yönlendirici kateter içerisinde sıkıştı ve problemin sol atriyumda çözülememesi nedeniyle işleme son verme kararı aldık.