Journals / Türk Kardiyoloji Derneği Arşivi / 2021 / Cilt: 49 - Sayı: 2
A case report of an acute reversible increase in pacing impedance of a ventricular epicardial lead after catheter ablation of ventricular tachycardia
- Pages
- 162–166
- DOI
- —
Abstract
Summary– Pacemakers are lifesaving devices that are being implanted with various indications, such as sinus node disease, atrioventricular block, and cardiac resynchronization therapy. Impedance measurement is one of the integral tests by which electrical resistance in pacing lead is tested. In this paper, we report an interesting observation of sudden impedance rise after ventricular tachycardia ablation with transmural lesions, which subsequently normalized without any need for intervention. The clinical implication of our findings is that careful observation might be indicated instead of lead replacement in case of a sudden surge in epicardial lead impedance after endocardial ablation owing that the impedance surge might be reversible
Özet
Özet– Kalıcı kalp pilleri sinüs düğüm hastalığı, atriyoventriküler düğüm hastalığı ve kardiyak resenkronizasyon tedavisi gibi çeşitli endikasyonlarla implante edilmekteler. İmpedans ölçümü pil lead’lerin elektriksel rezistansını ölçen temel bir testir. Biz bu raporda transmural lezyonların oluşturulduğu bir ventriküler takikardi ablasyonu vakasında kendiğinden gerileyen akut impedans artışını sunuyoruz. Ventrikül lead’lerin insersiyo bölgesi yakınlarında yapılan transmural lezyon hedefleyen ablasyonlardan sonra akut impedans artışları görülebilir, ancak, geridönüşümlü olması sebebiyle herhangi müdahaleden önce klinik takip yapılması önerilebilir.