An effective and safe alternative to epicardial pacemaker placement for permanent pacemaker implantation in a patient with mechanical tricuspid valve: stimulation of the left ventricle through the coronary sinus
- Pages
- 244–247
- DOI
- —
Abstract
Implantation of a transvenous endocardial pacemaker is contraindicated in patients with a mechanical tricuspid valve. An epicardial left ventricular pacemaker lead was placed by a transvenous route through the coronary sinus into the lateral cardiac vein in a 58-year-old woman with mechanical aortic, mitral, and tricuspid valves, for permanent pacing due to chronic atrial fibrillation with a slow ventricular rate accompanied by syncope. This lead was then connected to a singlechamber pacemaker. The patient had no problem in the following three months. Placement of an epicardial pacing lead through the coronary sinus provides a safe and effective pacing in patients with a mechanical tricuspid valve, thus obviating major cardiac surgery
Özet
Mekanik triküspit kapak takılmış hastalara transvenöz endokardiyal pil yerleştirilmesi kontrendikedir. Aort, mitral ve triküspit kapakları değiştirilmiş 58 yaşında bir kadın hastaya, bayılma yakınmasına eşlik eden yavaş ventrikül yanıtlı atriyal fibrilasyon nedeniyle kalıcı pil uyarısı uygulamak için transvenöz yolla, koroner sinüs lateral kardiyak vene epikardiyal sol ventrikül kalp pili elektrodu yerleştirildi. Bu elektrot standart tek odacıklı kalp piline bağlandı. Üç aylık takibi sırasında hastada herhangi bir sorun yaşanmadı. Mekanik triküspit kapağı olan hastalarda, büyük bir cerrahi işleme gerek kalmadan, koroner sinüs dallarına epikardiyal olarak elektrot yerleştirmek ile güvenli ve etkili pil uyarısı elde edilmiş olur.