Journals / Selçuk Tıp Dergisi / 2019 / Cilt: 35 - Sayı: 2
Hybrid Extraction of Aortic Arcus Embolizated Atrial Septal Defect Occluder Device: One Nail Drives Out Another
- Journal
- Selçuk Tıp Dergisi
- Pages
- 133–136
- DOI
- —
Abstract
Atrial septal defect (ASD) is also responsible for 30-33% of adult defects. Transcatheter treatment is the first-line treatment for secundum ASD treatment, If defect is appropriate . A 36-year-old ASD patient that had a successful transcatheter closure. The occluder embolization became, within the first 24 hours after procedure, to the aortic arcus. It is known that the devices with embolism are taken by snare and biopsy catheter, but differently, this case involving surgical removal after retraction to femoral artery by another ASD occluder device. Hybrid approach should always be mindful when removing embolized device with high success rate, using different methods.
Özet
Atriyal septal defekt(ASD) erişkinlerdeki defektlerin de % 30-33’ünden sorumludur. Sekundum ASD tedavisinde transkateter tedavi, uygun defektler için birinci basamak tedavi olarak yer almaktadır. 36 yaşında sekundum ASD tanısı olan hastanın başarılı transkateter kapatma işlemi sonrası ilk 24 saat içinde aortik arka embolize olduğunu fark ettik. Embolize olan cihazların snare ve biyopsi kateteri ile alınması bilinmekle birlikte, farklı olarak femoral artere başka bir ASD okluder cihaz ile çekilmesi sonrası buradan cerrahi ile alınmasını içeren bir vaka sunduk. Farklı yöntemler kullanıp başarı şansını yükseltmek için embolize olan cihaz çıkartılmasında hibrit yaklaşım her zaman akılda olmalıdır.