Journals / International Journal of Medical Biochemistry / 2022 / Cilt: 5 - Sayı: 2
Sigma metric evaluation with different TEa targets in clinical biochemistry
- Pages
- 120–124
- DOI
- —
Abstract
Objectives: The aim of this study was to evaluate the analytical performances of various clinical biochemistry analytes by the sigma metrics method according to different total allowable error (TEa) targets and to determine the causes of errors that lead to low sigma score by using Quality Goal Index (QGI). Methods: The study was carried out in the Central Laboratory of Bursa Karacabey State Hospital. Twelve analytes that were studied on the Roche Cobas c 501 autoanalyzer were included in the study. Internal (level 1 and 2) and external quality control data for the period March–August 2020 were obtained retrospectively. The TEa targets were obtained from the Clinical Laboratory Improvement of 2019 (CLIA 2019), biological variation database (BVD), Rili-BAEK, and Turkish data. QGI was calculated for analytes with sigma score