Tiroid ve paratiroidin sintigrafik görüntülenmesinde pinhol kolimatör ile çekim hatalarının değerlendirilmesi

Pages
29–32
DOI
—

Abstract

In this study acquisition mistakes that can be observed during thyroid and parathyroid imaging were quantified. We used standart phantoms that situmulate extensions and flexions of the patients neck during the acquisition with a pinhole collimator. The localization mistakes that can be occured while the phantom was angled with 5°C, 10°C and 25°C both for extension and flexion positions. At the same positions, localization mistakes were evaluated during the parathyroid adenoma imaging. According to the standard imaging of parathyroid and thyroid, a 10 cm distance between the collimator and phantom surface was kept. When the phantom was in an angle of 25°C, localization mistakes of 8 mm in thyroid imaging and 14 mm in parathyroid imaging were observed.

Özet

Bu çalışmada tiroid ve paratiroid sintigrafisi çekimlerinde olası görüntüleme hataları kantifiye edilmiştir. Hasta boynunun öne ya da arkaya eğimlenerek pinhol kolimatör ile görüntülenmesi konumlarını simule eden standart fantomlar kullanıldı. Tiroid fantomunun yatayla 5°C, 10°C ve 25°C'lik açılarla öne ve arkaya eğimlenmesi durumlarında oluşabilen lokalizasyon yanılgıları belirlendi. Aynı açılandırmalar ile görüntülerdeki paratiroid adenomlarına ait lokalizasyon yanılgıları belirlendi. Klinik çekim protokolüne uygun olarak, fantom yüzeyi ile kolimatör ağzı arasındaki yükseklik 10 cm ve açılandırma 25°C iken tiroid görüntülerinde 8 mm, paratiroid görüntülerinde 14 mm'lik lokalizasyon yanılgıları tespit edildi.

Tiroid ve paratiroidin sintigrafik görüntülenmesinde pinhol kolimatör ile çekim hatalarının değerlendirilmesi — AJIndex