Postoperative biometry in pseudophakic eyes

Psödofakik gözlerde biyometrik değerlendirme

Pages
234–236
DOI
—

Abstract

Preopretive and postoperative biometry was performed in 100 eyes which underwent cataract extraction with posterior chamber intraocular lens implantation. The mean preoperative axial length was 23.40±0.91mm and the mean postoperative axial length was 23.15±1.12 mm. This difference was attributed to va¬ried ultrasound speed in phakic and pseudophakic eyes rather than axial length shortening due to surgery. The mean postoperative anterior chamber depth (ACD) was 4.01+0.51 mm. The mean postope¬rative ACD was shallower by 0.27-1.14 mm than pre¬sumed values with various intraocular lens models. We concluded that presumed postoperative anterior chamber depth on the lens prospectus should be re¬garded with suspicion.

Özet

Arka kamara intraoküler lens implantasyonu yapılan 100 olguda preoperatif ve postoperatif biyometrik ölçümler yapılarak sonuçlar değerlendirildi. Ortalama preoperatif aksiyel uzunluk 23.40+0.91 mm, ortalama postoperatif aksiyel uzunluk 23.15+1.12 mm idi. Preoperatifve postoperatif aksiyel uzunluklar arasındaki bu .farklılık, cerrahiye bağlı glob boyutu değişiminden daha çok ultrason hızının fakik ve psödofakik gözlerde farklı olmasına bağlandı. Ortalama postoperatif ön kamara derinliği 4.01±0.51 mm idi. Kullanılan in¬traoküler lens modeline göre, ortalama postoperatif ön kamara de¬rinliği tahmini değerlerden 0.27 ila 1.14mm daha sığ bulundu. Üre¬tici firmaların önerdiği tahmini postoperatif ön kamara derinliği değerlerinin şüphe ile karşılanması gerektiği kanısına varıldı.