Journals / MN Oftalmoloji / 1998 / Cilt: 5 - Sayı: 2

Reliability of Keeler Pulsair 2000 non-contact tonometer in measuring the intraocular pressure in the pediatric age group

Pediatrik yaş grubunda göz içi basıncının ölçümünde Keeler Pulsair 2000 non-kontakt tonometresinin güvenirliliği

Pages
181–182
DOI
—

Abstract

This study is designed to determine the availability and reliability of Pulsair noncontact tonometers(NKT) in measuring the intraocular pressure(IOP) in the pédiatrie age group. 100 eyes of 50 cases aged between 6-8 years were included in this study. The intraocular pressures were measured by Pulsair noncontact tonometers, Goldmann applanation tonometer and Perkins hand-held applanation tonometer. The mean intraocular pressures values were found 13.92 $\pm$ 1.60 mmHg, 14.08±1.54 mmHg, and 13.79 $\pm$ 1.75 mmHg respectively. No statistically difference was determined in the three groups(p>0.05). It has been concluded that Pulsair noncontact tonometer is an available and reliable method in measuring the intraocular pressure in the pédiatrie age group.

Özet

Bu çalışma pediatrik yaş grubunda Pulsair nonkontakt tonometrenin göz içi basmcı(GİB)'nı ölçmedeki kullanılabilirliliğini ve güvenirliliğini saptamak amacıyla yapıldı. Çalışmaya yaşları 6 ile 8 arasında değişen 50 olgunun 100 gözü alındı. Göz içi basınçları Pulsair nonkontakt tonometre(NKT), Goldmann applanasyon tonometresi ve Perkins el applanasyon tonometresi ile ölçüldü. Ortalama göz içi basınçları sırasıyla; 13.92±1.60mmHg, 14.08±1.54mmHg ve 13.79±1.75mmHg olarak bulundu. Her üç yöntem arasında istatistiksel olarak anlamlı bir fark saptanmadı(p>0.05). Bu sonuçla pediatrik yaş grubunda göziçi basıncını ölçmede Pulsair nonkontakt tonometrenin kullanılabilir ve güvenli bir yöntem olduğu düşünüldü.