Journals / MN Oftalmoloji / 1999 / Cilt: 6 - Sayı: 2
Coating the Goldmann applanation tonometer prism with nylon stretch film
- Journal
- MN Oftalmoloji
- Pages
- 195–196
- DOI
- —
Abstract
PURPOSE: In our study, we investigated the effect of coating the Goldmann applanation tonometer prism with nylon strech film on pressure readings. METHOD: Intraocular pressure measurements were determined on right and then left eyes of 33 patients with Goldmann applanation tonometer for three times. Intraocular pressure readings were repeated with strech film coated prism on the same patients. RESULTS: Mean intraocular pressure was 18.215.0 mmHg for right eyes, 17.3±5.0mmHg for left eyes. With film coated prism, mean intraocular pressure for right eyes was 18±5.1 mmHg, for left eyes 17.5±5.2 mmHg. Coating applanation with strech film did not develop any statistical difference or intraocular pressure values (p=0.58, 0.52). CONCLUSION: Coating applanation prism tip with nylon strech film may help in preventing transmission of infectious agents without any affect on applanation intraocular pressure readings.
Özet
AMAÇ: Goldmann aplanasyon tonometresi prizmasının naylon streç film tabakası ile kaplanmasının ölçümlere etkisi incelendi. YÖNTEM: Göz içi basınç değerleri 33 hastanın önce sağ daha sonra sol gözlerinde Goldmann aplanasyon tonometresi ile üç kez saptandı. Aynı hastalarda streç filmle kaplı prizmayla ölçümler tekrarlandı. BULGULAR: Sağ göz içi basınçları ortalama 18.2±5.0 mmHg, sol gözde ise 17.3±5.0 mmHg bulundu. Film kaplı prizma ile elde edilen göziçi basıncı değerleri sağ göz için 18.0±5.1 mmHg sol göz için ise 17.5±5.2 mmHg olarak ölçüldü. Elde edilen değerler her iki göz için istatistiksel farklılık yaratmadı (p=0.58, 0.52). SONUÇ: Aplanasyon prizmasının naylon streç film tabakası ile kaplanması ölçüm sonuçlarını etkilemeden enfeksiyon etkenlerinin bulaşmasını engelleyebilir.