Journals / Middle East Journal of Science / 2020 / Cilt: 6 - Sayı: 2
COMPARATIVE ASSESSMENT OF DOSE CALIBRATORS USED IN NUCLEAR MEDICINE
- Pages
- 44–56
- DOI
- —
Özet
Dose calibrators are used to measure the amount of radioactive to be given to the patient. It is necessary to determine the correct dose and measure the amount of radioactive material with the least possible error. To minimize these potential errors, quality control (QC) tests should be carried out periodically according to the United States Nuclear Regulatory Authority (NRC). ATOM LAB 400 (serial number:11070208 ) and 500 (serial number: 15091215) dose calibrators that are actively used in our clinic were used. The aim of this study is to compare with another recently calibrated dose calibrator to verify the dose calibrator that needs to be updated is working properly. QC tests were performed on both dose calibrators. Test results of the currently certified dose calibrator and ATOM LAB 400 dose calibrator whose certificate will be updated were found to be compatible with each other. The tests performed on both dose calibrators remained with the error limits. The calibration certificate of the ATOM LAB 400 calibrator has been updated in accordance with NRC protocol.