Journals / Cerrahpaşa Tıp Dergisi / 2003 / Cilt: 34 Sayı: 1

EKSTERNAL RADYOTERAPİ CİHAZLARINA AİT DOZİMETRİK PARAMETRELERİN ÖLÇÜM VE FABRİKA DEĞERLERİNİN KARŞILAŞTIRMASI

Pages
1–1
Publication Date
2014-07-26
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

After the comparison of isodose distributions, for 6 MV, 1 cm difference was found between the measured and manufacturers. atlases.Conclusion.- Regarding, it was seen the dosimetric parameters that there were differences in percent depth doses, penumbras, and isodose distributions derived from high energy photon beams between measured data and the ones in the manifacturers. atlases, therefore before starting the treatment with these machines, all dosimetric parameters have to be measured.* Anahtar Kelimeler: Yüksek enerjili fotonlar, yüzde derin doz, penumbra, izodoz eğrileri, ışın kalitesi* Key Words: High energy photon beams, percentage depth dose, penumbra, isodose, beam quality

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Background and Design.- High energy photon beams which are produced by linear accelerators and Co60 treatment machines are used to treat deep seated tumors in radiotherapy.
📊
Key Findings & Data:Measurements were performed on a Saturne-42 linear accelerator with photon energies of 6, 15 MV, Orion linear accelerator with 4 MV photon energy, and Co60 treatment machine with 1.25 MeV photon energy .Beam quality measurements made in an RW3 solid water phantom using cylindric ionization chamber and PTW electrometer .
💡
Conclusion & Impact:After the comparison of isodose distributions, for 6 MV, 1 cm difference was found between the measured and manufacturers. atlases.Conclusion.- Regarding, it was seen the dosimetric parameters that there were differences in percent depth doses, penumbras, and isodose distributions derived from high energy photon beams between measured data and the ones in the manifacturers. atlases, therefore before starting the treatment with these machines, all dosimetric parameters have to be measured.* Anahtar Kelimeler: Yüksek enerjili fotonlar, yüzde derin doz, penumbra, izodoz eğrileri, ışın kalitesi* Key Words: High energy photon beams, percentage depth dose, penumbra, isodose, beam quality
🔬Design:Experimental Field Trial
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Özet

Background and Design.- High energy photon beams which are produced by linear accelerators and Co60 treatment machines are used to treat deep seated tumors in radiotherapy. In this study, parameters, the percent depth doses, dose profiles, penumbras, isodose distributions and beam quality are compared to those given in the manufacturers. atlases. Measurements were performed on a Saturne-42 linear accelerator with photon energies of 6, 15 MV, Orion linear accelerator with 4 MV photon energy, and Co60 treatment machine with 1.25 MeV photon energy .Beam quality measurements made in an RW3 solid water phantom using cylindric ionization chamber and PTW electrometer . Percent depth doses, dose profiles, penumbras, and isodose distribution measurements made in a water phantom.Results.- When the results were compared, the maximum difference in percent depth doses between our measurements and the manufactures. atlases was 5.2 % for 6 MV. At the depth of dose maximum the difference in penumbras found to be within 2 mm. After the comparison of isodose distributions, for 6 MV, 1 cm difference was found between the measured and manufacturers. atlases.Conclusion.- Regarding, it was seen the dosimetric parameters that there were differences in percent depth doses, penumbras, and isodose distributions derived from high energy photon beams between measured data and the ones in the manifacturers. atlases, therefore before starting the treatment with these machines, all dosimetric parameters have to be measured.* Anahtar Kelimeler: Yüksek enerjili fotonlar, yüzde derin doz, penumbra, izodoz eğrileri, ışın kalitesi* Key Words: High energy photon beams, percentage depth dose, penumbra, isodose, beam quality

🔖

Cite This Article

DEMİR, B., KOCA, A., GÜNHAN, B., KARAÇAM, S., ÖBER, A. (2014). EKSTERNAL RADYOTERAPİ CİHAZLARINA AİT DOZİMETRİK PARAMETRELERİN ÖLÇÜM VE FABRİKA DEĞERLERİNİN KARŞILAŞTIRMASI. Cerrahpaşa Tıp Dergisi, Cilt: 34 Sayı: 1, 1–1. https://ajindex.com/articles/984090

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...