Journals / Uludağ Üniversitesi Mühendislik Fakültesi Dergisi / 2018 / Cilt: 23 - Sayı: 3

EFFECT OF DENTAL IMPLANT DIMENSIONS ON FATIGUE BEHAVIOUR: A NUMERICAL APPROACH

Dental İmplant Ölçülerinin Yorulma Davranışına Etkisi: Sayısal Bir Yaklaşım

Pages
249–260
Publication Date
2022-11-04
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

The parameters with the most effect on the implant screw fatigue were pitch and the height ofthe tooth.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:In this study, the effect of the implant shape on fatigue behavior was investigated with thefinite element method.
📊
Key Findings & Data:The implant material used was Ti6Al4V.
💡
Conclusion & Impact:The parameters with the most effect on the implant screw fatigue were pitch and the height ofthe tooth.
🔬Design:Modeling & Simulation
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

In this study, the effect of the implant shape on fatigue behavior was investigated with thefinite element method. The implant material used was Ti6Al4V. The examination focus on the effect onthe implant life with respect to changes in the crest and root dimension, thread depth, thread pitch,implant length, implant diameter, chamfer length, chamfer radius, and groove length to implant lengthratio. Models were subjected to fatigue analysis according to ISO 14801 standard with ANSYS finiteelement software. The mean stress correction theory is chosen in fatigue life solutions. As a result, theparameters seen to have the most effect on dental implant fatigue behavior were the implant length anddiameter. The parameters with the most effect on the implant screw fatigue were pitch and the height ofthe tooth.

Özet

Bu çalışmada implant şeklinin yorulma davranışına etkisi sonlu elemanlar metoduyla araştırılmıştır. İmplant malzemesi olarak Ti6Al4V kullanılmıştır. Çalışmada diş dibi kalınlığı, diş üstü kalınlığı, diş yüksekliği, diş hatvesi, implant boyu, implant çapı, pah boyu, pahın yarıçapı ve oyuk boyu değerlerinin değişiminin implant ömrüne etkisi araştırılmıştır. Modeller ISO 14801 standardında belirtilen kriterlere göre ANSYS programında yorulma analizine tabi tutulmuştur. Sonuçta dental implantların yorulma davranışını en çok etkileyen boyut değerlerinin implant çapı ve implant boyu olduğu görülmüştür. İmplant vidası yorulma ömrü üzerindeki en etkili parametreler hatve ve diş yüksekliğidir.

🔖

Cite This Article

KAMAN, M. O., TOPKAYA, H. (2022). EFFECT OF DENTAL IMPLANT DIMENSIONS ON FATIGUE BEHAVIOUR: A NUMERICAL APPROACH. Uludağ Üniversitesi Mühendislik Fakültesi Dergisi, Cilt: 23 - Sayı: 3, 249–260. https://ajindex.com/articles/643766

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (2)

  1. Ao J., Li T., Liu Y., Ding Y., Wu G., Hua K., Kong L. Optimal design of thread height and width on an immediately loaded cylinder implant: A finite element analysis, Computers in Biology and Medicine 2010; 40: 681-686 doi:10.1016/j.compbiomed.2009.10.007. Ausiello P., Franciosa P., Martorelli M.,Watts D.C. Effects of thread features in osseo-integrated titanium implants using a statistics-based finite element method. Dental Materials 2012; 28: 919-927. doi: 10.1016/j.dental.2012.04.035 Barbier L., Vander Sloten J., Krzesinski G., Schepers E., Van Der Perre G. Finite element analysis of non-axial versus axial loading of oral implants in the mandible of the dog. Journal of Oral Rehabilitation.1998;25:847-858.doi: 10.1046/j.13652842.1998.00318.x Chun H-J., S.-Y. Cheong S-Y., Han J-H., Heo S-J., Chung J-P., Rhyu I-C., Choı Y-C., Baik H-K., Ku Y. ve Kım M-H. Evaluation of design parameters of osseointegrated dental implants using finite element analysis. Journal of Oral Rehabilitation 2002; 29: 565–574. doi: 10.1046/j.1365-2842.2002.00891.x Daas, M., Dubois, G., Bonnet, A. S., Lipinski, P., Rignon-Bret, C. A complete finite element 224 model of a mandibular implant retained overdenture with two implants: Comparison between rigid and 225 resilient attachment configurations. Medical Engineering & Physics 2008; 30: 218-225. doi: 10.1016/j.medengphy.2007.02.005 Dilek O., Tezulas E., Dincel M. Required minimum primary stability and torque values for immediate loading of mini dentalimplants: an experimentalstudy in nonviable bovine femoral bone. Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology, and Endodontology 2008; 105: 20-27. doi: 10.1016/j.tripleo.2007.10.003. Dittmer S., Dittmer M.P., Kohorst P., Jendras M., Borchers L., Stiesch M. Effect of implant–abutment connection design on load bearing capacity and failure mode of implants. Journal of Prosthodontics 2011; 20: 510-516. doi: 10.1111/j.1532-849X.2011.00758.x Djebbar N., Serier B., Bachir Bouiadjra B., Benbarek S., Drai A. Analysis of the effect of load direction on the stress distribution in dental implant. Materials & Design 2010; 31: 2097-2101. doi: 10.1016/j.matdes.2009.10.042 Ekici B. Numerical Analysis of a dental system in three-dimension. Advances in Engineering Software 2002; 33: 109-113. doi: 10.1016/S0965-9978(01)00056-4
  2. Figueiredo R.B., Barbosa E. R. C., Zhao X., Yang X., Liu X., Cetlin P.R., Langdon T.G. Improving the fatigue behavior of dental implants through processing commercial pu