Journals / Selcuk Dental Journal / 2021 / Cilt: 8 - Sayı: 3

Evaluation of Interdevice Reliability and Repeatability of TwoIdentical Spectrophotometers

Özdeş İki Spektrofotometrenin Cihazlar Arası GüvenilirlikVe Tekrarlanabilirliğinin İncelenmesi

Pages
605–610
Publication Date
2022-10-18
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

The ∆E00 values of the shade tabsrange from 0.21 to 1.78, which were lower than the acceptabilitythreshold (∆E00

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Background: The purpose of this in-vitro study was to evaluatethe repeatability and interdevice reliability of two identicalspectrophotometers with different software.
📊
Key Findings & Data:Methods: Twenty-six shade tabs of a shade guide (Vita 3D-Master,Vita Zahnfabrik) were measured with two identicalspectrophotometers (Vita EasyShade V) with different software(v509i: Device A; v505p: Device B), and CIELab values wererecorded (n=260).
💡
Conclusion & Impact:The ∆E00 values of the shade tabsrange from 0.21 to 1.78, which were lower than the acceptabilitythreshold (∆E00
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

Background: The purpose of this in-vitro study was to evaluatethe repeatability and interdevice reliability of two identicalspectrophotometers with different software. Methods: Twenty-six shade tabs of a shade guide (Vita 3D-Master,Vita Zahnfabrik) were measured with two identicalspectrophotometers (Vita EasyShade V) with different software(v509i: Device A; v505p: Device B), and CIELab values wererecorded (n=260). The color difference (∆E00) values betweenmeasurements of spectrophotometers were calculated. Forrepeatability, the 3M2 shade tab of the same shade guide wasmeasured 50 times with each spectrophotometer. The matchingaccuracy rate of the devices with shade guide was calculated.Interdevice reliability and differences between devices werecalculated using intraclass correlation coefficient (ICC),Cronbach's alpha, and Paired T-test (α=0.05). Results: Interdevice reliability was found to be high for L*, a*, andb* parameters (ICC>0.99). The ∆E00 values of the shade tabsrange from 0.21 to 1.78, which were lower than the acceptabilitythreshold (∆E00

Özet

Amaç: Bu in-vitro çalışmanın amacı, farklı yazılımlara sahipözdeş iki spektrofotometrenin tekrarlanabilirliğini ve cihazlararası güvenilirliğini değerlendirmektir. Gereç ve Yöntemler: Bir renk skalasına (Vita 3D-Master, VitaZahnfabrik) ait 26 adet renk tabı, farklı yazılımlara sahip (v509i:A cihazı, v505p: B cihazı) iki özdeş spektrofotometre (VitaEasyShade V) ile ölçüldü ve CIELab değerleri kaydedildi(n=260). Spektrofotometrelerin ölçümleri arasındaki renkfarkları (∆E00) hesaplandı. Tekrarlanabilirlik değerlendirilmesiiçin aynı skalanın 3M2 renk tabı, iki spektrofotometre ile ölçüldü(n=50). Cihazların skala ile renk eşleştirme doğruluk oranıhesaplandı. Cihazlar arası güvenilirlik ve cihazlar arasındakiölçüm farkları, sınıf içi korelasyon katsayısı (ICC), Cronbach’salfa ve Paired T-testi kullanılarak hesaplandı (α=0.05). Bulgular: Cihazlar arası güvenilirlik L*, a* ve b* parametreleriiçin yüksek düzeyde bulunmuştur (ICC>0.99). Renk skalasının∆E00 değerleri 0.21 ile 1.78 arasında değişmektedir veskaladaki bütün renklerin ∆E00 değerleri kabul edilebilirlik eşikdeğerinin altındadır (∆E00

🔖

Cite This Article

SAVAŞ, A., SAVAŞ, T. Y., Tunçdemir, A. R. (2022). Evaluation of Interdevice Reliability and Repeatability of TwoIdentical Spectrophotometers. Selcuk Dental Journal, Cilt: 8 - Sayı: 3, 605–610. https://ajindex.com/articles/591405

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (8)

  1. Kim HK. Evaluation of the repeatability and matching accuracy between two identical intraoral spectrophotometers: an in vivo and in vitro study. J Adv Prosthodont 2018;10:252-8
  2. Van der Burgt T, Ten Bosch J, Borsboom P, Kortsmit W. A comparison of new and conventional methods for quantification of tooth color. J Prosthet Dent 1990;63:155-62
  3. Chu SJ, Trushkowsky RD, Paravina RD. Dental color matching instruments and systems. Review of clinical and research aspects. J Dent 2010;38 2-16
  4. Brewer JD, Wee A, Seghi R. Advances in color matching. Dent Clin North Am 2004;48:341-58
  5. Dagg H, O'connell B, Claffey N, Byrne D, Gorman C. The influence of some different factors on the accuracy of shade selection. J Oral Rehabil 2004;31:900-4
  6. Paul SJ, Peter A, Rodoni L, Pietrobon N. Conventional visual vs spectrophotometric shade taking for porcelainfused-to-metal crowns: a clinical comparison. J Prosthet Dent 2004;92:577
  7. International Organization for Standardization. ISO 5725- 1. Accuracy (trueness and precision) of measurement methods and results - Part 1: General principles and definitions. Geneva: ISO; 1994. Available at: https://www.iso.org/obp/ ui/#iso:std:iso:5725:-1:ed1:v1:en 8. Kim-Pusateri S, Brewer JD, Davis EL, Wee AG. Reliability and accuracy of four dental shade-matching devices. J Prosthet Dent 2009;101:193-9
  8. Hammad IA. Intrarater repeatability of shade selections with two shade guides. J Prosthet Dent 2003;89:50-3. 10.Karaagaclioglu L, Terzioglu H, Yilmaz B, Yurdukoru B. In vivo and in vitro assessment of an intraoral dental colorimeter. J Prosthodont 2010;19:279-85. 11.Tsiliagkou A, Diamantopoulou S, Papazoglou E, Kakaboura A. Evaluation of reliability and validity of three dental colour-matching devices. Int J Esthet Dent 2016;11(1):2-16. 12.Klotz AL, Habibi Y, Corcodel N, Rammelsberg P, Hassel AJ, Zenthofer A. Laboratory and clinical reliability of two spectrophotometers. J Esthet Restor Dent. 2018:1-5. 13.Luo MR, Cui G, Rigg B. The development of the CIE 2000 colour‐difference formula: CIEDE2000. Col Res Appl 2001;26:340-50. 14.Paravina RD, Ghinea R, Herrera LJ, Bona AD, Igiel C, Linninger M et al. Color difference thresholds in dentistry. J Esthet Restor Dent 2015;27 Suppl 1:S1-9. 15.Yuan JC-C, Brewer JD, Monaco Jr EA, Davis EL. Defining a natural tooth color space based on a 3-dimensional shade system. J Prosthet Dent 2007;98:110-9. 16.Öngül D, Şermet B, Balkaya MC. Visual and instrumental evaluation of color match ability