Journals / Hand and Microsurgery / 2021 / Cilt: 10 - Sayı: 3

What must be the volar plate screw length to prevent extensor tendon injuries for distal radius fractures fixation?

Pages
269–273
Publication Date
2022-10-19
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

Conclusion: To prevent extensor tendon injury/rupture, downsizing of screws is recommended as is careful checking of screw length on intraoperative fluoroscopic views often performed in multiple planes

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Objectives: We aimed to determine the appropriate screw lengths to prevent dorsal screw penetration in fixation with volar plate for distal radius fractures and to demonstrate the danger of relying on intraoperative fluoroscopic views to determine screw penetration.
📊
Key Findings & Data:Methods: Magnetic resonance imaging (MRI) scans of fifty patients’ wrists conducted with a 1.5-T scanner (Siemens, Erlangen, GY) were used.
💡
Conclusion & Impact:Conclusion: To prevent extensor tendon injury/rupture, downsizing of screws is recommended as is careful checking of screw length on intraoperative fluoroscopic views often performed in multiple planes
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

Objectives: We aimed to determine the appropriate screw lengths to prevent dorsal screw penetration in fixation with volar plate for distal radius fractures and to demonstrate the danger of relying on intraoperative fluoroscopic views to determine screw penetration. Methods: Magnetic resonance imaging (MRI) scans of fifty patients’ wrists conducted with a 1.5-T scanner (Siemens, Erlangen, GY) were used. MRI scans evaluated independently and at different times by two musculoskeletal specialist radiologists and the means of each measurements from each patient’s images were used for the statistical analysis. All measurements were performed with standard MRI evaluation techniques. Volar to dorsal measurements were performed for extensor carpi radialis longus (ECRL), extensor carpi radialis brevis (ECRB), Lister’s tubercle (LT), extensor pollicis longus (EPL) and extensor digitorum communis (EDC) tendon compartments to find the real screw length and compared with the lateral view of the distal radius especially Lister’s tubercle. Results: Patients ages ranged from 20 to 64 years and the mean age was 39.9±13.2 years. For female patients, the mean height lengths for LT and ECRL, ECRB, EPL and EDC compartments were 20.7±2.5 mm, 13.0±1.1 mm, 16.4±1.3 mm, 18.1±1.1 mm and 17.0±1.0 mm, respectively. For male patients, the mean height lengths for LT and ECRL, ECRB, EPL and EDC compartments were 23.5±2.2 mm, 15.1±1.8 mm, 18.9±2.4 mm, 20.6±2.1 mm and 19.5±2.4 mm, respectively. Conclusion: To prevent extensor tendon injury/rupture, downsizing of screws is recommended as is careful checking of screw length on intraoperative fluoroscopic views often performed in multiple planes

🔖

Cite This Article

Erden, T., Yücetürk, A., Toker, B. (2022). What must be the volar plate screw length to prevent extensor tendon injuries for distal radius fractures fixation?. Hand and Microsurgery, Cilt: 10 - Sayı: 3, 269–273. https://ajindex.com/articles/531503

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (14)

  1. Haug LC, Glodny B, Deml C, Lutz M, Attal R. A new radiological method to detectdorsally pen- etrating screws when using volar locking plates in distal radialfractures. Bone Joint J 2013;95- B:1101–5
  2. Kamano M, Koshimune M, Toyama M, Kazuki K. Palmar plating system for Colles’ fractures–a pre- liminary report. J Hand Surg Am 2005;30:750–5
  3. Yuan ZZ, Yang Z, Liu Q, Liu YM. Complications following open reduction and internal fixation versus external fixation in treating unstable distal radius frac-tures: grading the evidence through a meta-analysis. Orthop Traumatol Surg Res2018;104:95–103
  4. Arora R, Lutz M, Hennerbichler A, Krappinger D, Espen D, Gabl M. Complications following internal fixation of unstable distal radius fracture with a palmar locking-plate. J Orthop Trauma 2007;21:316-22
  5. Benson EC, DeCarvalho A, Mikola EA, Veitch JM, Moneim MS. Two potential causes of EPL rupture after distal radius volar plate fixation. Clin Orthop Relat Res 2006;451:218-22
  6. Rozental TD, Blazar PE. Functional outcome and complications after volar plating for dorsally dis- placed, unstable fractures of the distal radius. J Hand Surg Am 2006;31:359-65
  7. Hill BW, Shakir I, Cannada LK. Dorsal Screw Pen- etration With the Use of Volar Plating of Distal Ra- dius Fractures: How Can You Best Detect? J Or - thop Trauma 2015;29:e408-13
  8. Pichler W, Windisch G, Schaffler G, Rienmuller R, Grechenig W. Computer tomography aided 3D analysis of the distal dorsal radius surface and the effects on volar plate osteosynthesis. J Hand Surg Eur Vol 2009;34:598-602
  9. Joseph SJ, Harvey JN. The dorsal horizon view: detecting screw protrusion at the distal radius. J Hand Surg Am 2011;36:1691-3
  10. Ozer K, Wolf JM, Watkins B, Hak DJ. Compari- son of 4 fluoroscopic views for dorsal cortex screw penetration after volar plating of the distal radius. J Hand Surg Am 2012;37:963-7
  11. Park DH, Goldie BS. Volar plating for distal radius fractures–do not trust the image intensifier when judging distal subchondral screw length. Tech Hand Up Extrem Surg 2012;16:169-72
  12. Smith DW, Henry MH. The 45 degrees pronated oblique view for volar fixed-angle plating of distal radius fractures. J Hand Surg Am 2004;29:703-6
  13. Clement H, Pichler W, Nelson D, Hausleitner L, Tesch NP, Grechenig W. Morphometric analysis of lister's tubercle and its consequences on volar plate fixation of distal radius fractures. J Hand Surg Am 2008;33:1716-9
  14. Gurbuz Y, Kucuk L, Gunay H, Ozaksar K, Sugun TS,