Journals / Gazi Medical Journal / 2006 / Cilt: 17 - Sayı: 3

Elektrocerrahi: Tehlikeler ve öneriler

Electrosurgery: Pitfalls and recommendationd

Pages
145–151
Publication Date
2022-10-19
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

It is recognized that simple precautions would sufficiently eradicate injuries caused by electro­surgery.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Purpose: In spite of the fact that the electrosurgical unit is the most frequ­ently used energy source in the operating room, it is probably the one least understood by the operating room personnel, including the physicians.
📊
Key Findings & Data:Materials and Methods: We documented all electrosurgery injuries occur­ring in our operating rooms over the previous 20 months.
💡
Conclusion & Impact:It is recognized that simple precautions would sufficiently eradicate injuries caused by electro­surgery.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

Purpose: In spite of the fact that the electrosurgical unit is the most frequ­ently used energy source in the operating room, it is probably the one least understood by the operating room personnel, including the physicians. This study was performed to develop a conceptual understanding of some basic principles concerning the adept use of electrosurgery instruments. Materials and Methods: We documented all electrosurgery injuries occur­ring in our operating rooms over the previous 20 months. We analyzed the injuries, and revealed causative factors and preventive measures to minimi­ze these injuries. Results: We encountered 12 electrosurgery injuries, six of which were glove bites suffered by the surgeons. Six of the injuries were different ele­ctrosurgery complications experienced by the patients, two of which were eventually fatal. Conclusion: Electrosurgery injuries in the operating room are infrequent, but we think that the issue deserves attention, because basic electrosurgical principles are important for appropriate, effective and safe use of the ins­truments and the prevention of stray current injuries. It is recognized that simple precautions would sufficiently eradicate injuries caused by electro­surgery.

Özet

Amaç: Elektrocerrahi üniteler, ameliyathanelerde en sık kullanılan enerji kaynaklan olmalarına karşın muhtemelen hem cerrahlar hem de diğer ame­liyathane personeli tarafından en az anlaşılan cihazlardır. Bu çalışma, elek­trocerrahi cihazlarının uygun kullanılabilmesi için bazı temel prensipleri ortaya koymak amacıyla yapılmıştır. Gereç ve Yöntemler: Son 20 ayda amliyathanelerimizde ortaya çıkan elek­trocerrahi yaralanmalarının dökümünü çıkardık. Yaralanmaları inceledik, sebep olabilecek faktörleri gözden geçirdik ve bu yaralanmaları en aza indi­rebilmek için alınabilecek önlemleri belirledik. Bulgular: On iki elektrocerrahi yaralanması tespit ettik, bunlardan altısı cerrahlarda meydana gelen eldiven yanıklarıydı. Altı yaralanma hastaların başına gelen değişik elektrocerrahi komplikasyonlarıydı ve bunlardan ikisi ölümcüldü. Sonuç: Ameliyathanelerde meydana gelen elektrocerrahi yaralanmaları seyrek olmakla birlikte konunun önemli olduğuna inanmaktayız. Çünkü ci­hazların uygun, etkili ve güvenli kullanımı ve kaçak akımlara bağlı yaralan­maların önlenebilmesi için temel elektrocerrahi prensipler önemlidir.

🔖

Cite This Article

YAVUZER, R., AYHAN, S., LATİFOĞLU, O., DEMİRTAŞ, Y., ÇELEBİ, C. (2022). Elektrocerrahi: Tehlikeler ve öneriler. Gazi Medical Journal, Cilt: 17 - Sayı: 3, 145–151. https://ajindex.com/articles/520432

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (11)

  1. Cushing H. Electrosurgery as an aid to the removal of intracranial tumors. Surg Gynecol Obstet. 1928
  2. Goldwyn RM. Bovie: the man and the machine. Ann Plast Surg. 1979
  3. O'Connor JL, Bloom DA. William T. Bovie and electrosurgery. Sur­ gery. 1996
  4. Licht S. History of therapeutic heat and cold. In: Lehmann JF (Ed.) Therapeutic Heat and Cold. Baltimore, Williams & Wilkins, 1984, Pp: 1-31. 5.Ulmer BC. The Valleylab Institute of Clinical Education electrosur­ gery continuing education module (Valleylab web site). July 2001. Available at: http://www.valleylabeducation.org. Accessed August 15,2004. 6.Absten GT. Practical electrosurgery for clinicians (Electrosurgery Professional Medical Education Association, Inc. Associate of the Caribbean Medical Association web site). September 2002. Availab­ le at: http://www.lasertraining.org/electros.htm. Accessed September 15,2004. 7.Demirtas Y, Mentes BB, Tatlicioglu E. Elektrocerrahi: temel prensip­ ler. Ankara Cerrahi Derneği, 2003 (suppl). 8.Sudhindra TV, Joseph A, Haray PN, Hacking BC. Are surgeons aware of the dangers of diathermy? Ann R Coll Surg Eng. 2000
  5. 82 (suppl): 189-190. 9.Mitchell J, Lumb G. The hazards of diathermy in urology. Proc R SocMed. 1960, 53:348-354. 10.Mitchell JP. The isolated circuit diathermy. Ann R Coll Surg Eng. 1979
  6. Miller E, Paull DE, Morrissey K, Cortese A, Nowak E. Scalpel ver­ sus electrocautery in modified radical mastectomy. Am Surg. 1988
  7. Shafighi M, Huemer GM, Meirer R, Piza-Katzer H. Comparison of epigastric skin flap survival in sharp versus electrocautery dissection in a rat-model. Plast Reconstr Surg. 2003
  8. Rappaport WD, Hunter GC, Allen R, Lick S, Halldorsson A, Chva- pil T, Holcomb M, Chvapil M. Effect of electrocautery on wound healing in midline laparotomy incisions. Am J Surg. 1990
  9. Schemmel M, Warren JS, Haefner HK, Termin CS, Selvaggi SM, Hurd WW. Comparison of the ultrasonic scalpel to CO2 laser and electrosurgery in terms of tissue injury and adhesion formation in a rabbit model. Fertil Steril. 1997
  10. Smith TL, Smith JM. Electrosurgery in otolaryngology-head and neck surgery: principles, advances, and complications. Laryngosco­ pe. 2001
  11. Pollinger HS, Mostafa G, Harold KL, Austin C, Kercher KW, Mat­ thews BD. Comparison of wound-healing characteristic with feed­ back circuit electrosurgical generators in a porcine model. Am Surg.