Journals / ARI The Bulletin of the Istanbul Technical University / 2004 / Cilt: 54 - Sayı: 1

Standardization of bovine lung thromboplastin according to the WHO report: Calculation of the ISI value and the prothrombin time as INR

Pages
22–25
Publication Date
2022-10-11
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

In conclusion the results of our evaluation show that our thromboplastin extract can be used safely as a reagent for PT tests instead of commercial PT kit.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Thromboplastin is an important reagent for prothrombin time (PT) test.
📊
Key Findings & Data:For this reason we prepared thromboplastin extract from bovine lung and calibrated according to the thirty-third report of the WHO Expert Committee on Biological Standardization.
💡
Conclusion & Impact:In conclusion the results of our evaluation show that our thromboplastin extract can be used safely as a reagent for PT tests instead of commercial PT kit.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

Thromboplastin is an important reagent for prothrombin time (PT) test. Various types of thromboplastin are prepared commercially and it is essential that each reagent is correctly standardized, in order to be able to interpret the results of the prothrombin time test. This will ensure that the results of tests with different products and batches are reproducible and can be compared. For this reason we prepared thromboplastin extract from bovine lung and calibrated according to the thirty-third report of the WHO Expert Committee on Biological Standardization. With this procedure, standardization parameter called International Sensitivity Index (ISI) was calculated and it is used for converting the PT values from seconds to International Normalized Ratio (INR). By this calculation it is possible to compare the PT results, which were performed by different thromboplastins. After standardization we used this thromboplastin extract for PT test and compared this extract with the commercial kit for PT test. In conclusion the results of our evaluation show that our thromboplastin extract can be used safely as a reagent for PT tests instead of commercial PT kit.

🔖

Cite This Article

TUNALI, T., YARAT, A., EMEKLİ, N. (2022). Standardization of bovine lung thromboplastin according to the WHO report: Calculation of the ISI value and the prothrombin time as INR. ARI The Bulletin of the Istanbul Technical University, Cilt: 54 - Sayı: 1, 22–25. https://ajindex.com/articles/448151

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (14)

  1. G.I.G. Ingrain and M. Hills, Thrombosis and Haemostasis 36, 237 (1976)
  2. S.M. Lewis, Brt. J. Haematol 66, 1 (1987)
  3. WHO Expert Committee on Biological Stan­dardization, Thirty-third Report (Geneva, 1983)
  4. WHO Expert Committee on Biological Series (Geneva, 1999)
  5. D.L. McGlasson, Clin. Lab. Sci. 12, 353 (1999)
  6. E. Bachlie, Br. J. Haematol 110, 248 (2000)
  7. T.B.L. Kirkwoodd, Thromb Haemostas 49, 238 (1983)
  8. L. Poller, Br. J. Haematol 67, 166 (1987)
  9. H.T. Moriorty, P.R.L. Lom-po Tang, and N. Anasta, Pathology 22, 71 (1990). [10]E.F. Mammen, Purification of tissue throm­boplastin in: Thrombosis and Bleeding dis­orders, (George Thieme Verlag, Stutgart Aca­demic Press, Newyork, London, 1971). [11]A.M.H.P. van den Besselaar, Thromb Haemostas 79, 439 (1996). [12]A. Tripodi, Thromb Haemostas 79, 439 (1998)
  10. L.N.G. Valerie, R.F. Comin, E.L. Gottfried, S.A.Echaves, A.G. Stead, and R.F. Eberd, Am. J.Clin. Pathol. 109, 338 (1998)
  11. J.M.K. Dufty, S. Craig, K.J. Stevenson, D.A.Taberner, J. Clin. Pathol. 50, 40 (1997)
  12. K.J. Stevenson, S. Craig, J.M.K. Dufty, and D.A. Taberner, Br. J. Haematol 96, 435 (1997)
  13. L. Poller, Acta Haematol 32, 292 (1968)
  14. H.J. Weiss, J.W. Eichelberger, and W.H.Crosby, J. Clin. Invest 40, 205 (1961).