Effects of local anesthetic use on non-contact tonometry measurement results and patient comfort

Pages
2625–2629
Publication Date
2022-10-12
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

The measurements with Shin Nippon NCT-10 with or without localanesthesia overestimated IOP compared to the GAT and cannot be used interchangeably with the GAT.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Aim: To evaluate the effects of local anesthetic use on non-contact tonometry (NCT) measurement results and patient comfortand to compare the intraocular pressure (IOP) readings obtained with Shin Nippon NCT-10 tonometer and Goldmann applanationtonometry (GAT).Material and Methods: 31 healthy participants were included in this study.
📊
Key Findings & Data:After the measurements, participants were asked to rate the pain felt during IOP measurements.
💡
Conclusion & Impact:The measurements with Shin Nippon NCT-10 with or without localanesthesia overestimated IOP compared to the GAT and cannot be used interchangeably with the GAT.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

Aim: To evaluate the effects of local anesthetic use on non-contact tonometry (NCT) measurement results and patient comfortand to compare the intraocular pressure (IOP) readings obtained with Shin Nippon NCT-10 tonometer and Goldmann applanationtonometry (GAT).Material and Methods: 31 healthy participants were included in this study. Central corneal thicknesses (CCT) were measured. All IOPmeasurements were performed in the same order as NCT without anesthesia, NCT with local anesthesia and GAT with anesthesia,respectively. After the measurements, participants were asked to rate the pain felt during IOP measurements. The range was from0 to 10; 0= very painful/reading impossible, 10=no sensation at all/reading very easily obtained. Less pain was evaluated as morecomfort.Results: NCT (p=0.007) and NCT-local (P=0.001) measurements overestimated IOP compared to GAT. IOP measured in NCT andNCT-local groups were similar (p=1.00). Mean patient comfort score was higher in GAT group than NCT group (p=0.006). Comparedwith NCT-local group, mean patient comfort score was lower in NCT group (P=0.033). Mean patient comfort scores were similarin GAT and in NCT-local groups (p=0.615). A positive but statistically insignificant correlation was found among NCT (r=0.201,p=0.277), NCT-local (r=0.259, p=0.160), GAT (r=0.272, p=0.139) measurements and CCT.Conclusion: Compared to NCT, IOP measurement with GAT is more comfortable. Performing NCT measurements with local anesthesiaincreases the patient comfort without changing the IOP values. The measurements with Shin Nippon NCT-10 with or without localanesthesia overestimated IOP compared to the GAT and cannot be used interchangeably with the GAT.

🔖

Cite This Article

BİLGİN, B., GÜLER, M., URFALIOĞLU, S. (2022). Effects of local anesthetic use on non-contact tonometry measurement results and patient comfort. Annals of Medical Research, Cilt: 26 - Sayı: 11, 2625–2629. https://ajindex.com/articles/446845

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (12)

  1. Lanza M, Borrelli M, De Bernardo M, et al. Corneal parameters and difference between Goldmann applanation tonometry and dynamic contour tonometry in normal eyes. J Glaucoma 2008;17:460- 4
  2. Pakrou N, Gray T, Mills R, et al. Clinical comparison of the Icare tonometer and Goldmann applanation tonometry. J Glaucoma 2008;17:43-7
  3. Mansoori T, Balakrishna N. Effect of central corneal thickness on intraocular pressure and comparison of Topcon CT-80 non-contact tonometry with Goldmann applanation tonometry. Clin Exp Optom 2018;101:206- 12
  4. Schiano Lomoriello D, Lombardo M, Tranchina L, et al. Repeatability of intra-ocular pressure and central corneal thickness measurements provided by a non-contact method of tonometry and pachymetry. Graefes Arch Clin Exp Ophthalmol 2011;249:429-34
  5. Yaoeda K, Shirakashi M, Fukushima A, et al. Measurement of intraocular pressure using the NT-4000:a new non-contact tonometer equipped with pulsesynchronous measurement function. J Glaucoma 2005;14:201-5
  6. Tonnu PA, Ho T, Newson T, et al. The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, noncontact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry. Br J Ophthalmol 2005;89:851- 4. 7. van der Jagt LH, Jansonius NM. Three portable tonometers, the TGDc-01, the ICARE and the Tonopen XL, compared with each other and with Goldmann applanation tonometry. Ophthalmic Physiol Opt 2005;25:429-35
  7. Ugalahi MO, Seidu MA, Olusanya BA, et al. Does patient comfort influence the choice of tonometer for the measurement of intraocular pressure? Int Ophthalmol 2016;36:385-9
  8. Almubrad TM, Ogbuehi KC. Clinical investigation of the effect of topical anesthesia on intraocular pressure. Clin Ophthalmol 2007;1:305-9
  9. Nam SM, Lee HK, Kim EK, et al. Comparison of corneal thickness after the ınstillation of topical anesthetics: proparacaine versus oxybuprocaine. Cornea 2006;25:51-4
  10. Lam AK, Chen D. Effect of proparacaine on central corneal thickness values: An evaluation using noncontact specular microscopy and pentacam. Cornea 2007;26:55-8
  11. Kato Y, Nakakura S, Matsuo N, et al. Agreement among Goldmann applanation tonometer, ICare, and Icare PRO rebound tonometers; non-contact tonometer; and Tonopen XL in healthy elderly subjects. Int Ophthalmol 2018;38:687-96
  12. Bang SP, Lee CE, Kim YC. Comparison of intraocular pressure as measured by three different non-contact tonometers and Goldmann applanation tonometer for n