Could you say that was an atrial flutter or not?

Pages
608–610
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

As Hippocrates saidcenturies ago “First, do no harm.”

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Muscle-tremor artefact is a potential cause of misdiagnosis of atrialarrhythmias on electrocardiography (ECG) monitoring.
📊
Key Findings & Data:Such errors maylead to inappropriate and potentially dangerous therapies in some patients.We present a case of a patient with uncontrolled seizures whose bedsideelectrocardiogram monitor analysis appeared to demonstrate atrial flutterwith 4:1 conduction through the AV node.
💡
Conclusion & Impact:As Hippocrates saidcenturies ago “First, do no harm.”
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Özet

Muscle-tremor artefact is a potential cause of misdiagnosis of atrialarrhythmias on electrocardiography (ECG) monitoring. Such errors maylead to inappropriate and potentially dangerous therapies in some patients.We present a case of a patient with uncontrolled seizures whose bedsideelectrocardiogram monitor analysis appeared to demonstrate atrial flutterwith 4:1 conduction through the AV node. The ECG monitor and ECGrhythm strip additionally showed a regular ventricular rate of 94 bpm with anunderlying regular ‘saw-tooth’ baseline. We applied cardioversion to convertto sinus rhythm. Amiodarone was loaded and added to the patients therapywho had atrial flutter after cardioversion. Echocardiogram was performed bya pediatric cardiologist and they noted that the atrial rate and ventricular ratewere equal. After this, we began to suspect this situation might be a pseudoflutter due to his muscle contractions. We applied rocuronium to the patientto understand whether this was a pseudo-flutter or not. We saw that the ECGreturned to normal sinus rhythm.Physicians especially working in intensive care units should be aware ofartifact to avoid unnecessary therapeutic procedures. As Hippocrates saidcenturies ago “First, do no harm.”

🔖

Cite This Article

ÖZSOYLU, S., AKYILDIZ, B. N., DURSUN, A., PAMUKÇU, Ö. (2019). Could you say that was an atrial flutter or not?. Turkish Journal of Pediatrics, Cilt: 61 - Sayı: 4, 608–610. https://ajindex.com/articles/412029

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (7)

  1. Krasnow AZ, Bloomfield DK. Artifacts in portable electrocardiographic monitoring. Am Heart J 1976; 91: 349-357
  2. Casta A, McCann ME. Electrocardiographic inscription of diaphragmatic contractions during anaesthesia. Paediatr Anaesth 1999; 9: 175-176
  3. Knight BP, Pelosi F, Michaud GF, Strickberger SA, Morady F. Clinical consequences of electrocardiographic artifact mimicking ventricular tachycardia. N Engl J Med 1999; 341: 1270-1274
  4. Patel S. Electrocardiographic artifact mimicking ventricular tachycardia during high-frequency oscillatory ventilation: A case report. Am J Crit Care 2006; 15: 310-311
  5. Boos CJ, Khan MY, Thorne S. An unusual case of misdiagnosed ventricular tachycardia. Emerg Med J 2008; 25: 173-174
  6. Samaniego NC, Morris F, Brady WJ. Electrocardiographic artefact mimicking arrhythmic change on the ECG. Emerg Med J 2003; 20: 356-357
  7. Finsterer J, Stollberger C, Gatterer E. Oral anticoagulation for ECG tremor artefact simulating atrial fibrillation. Acta Cardiol 2003; 58: 425-429.