AI Insight
⚡ TL;DR(Too Long; Didn't Read)

Basedon these results, we speculate that one of the most significant causes of adjacent segment disease is excessive bending rigidityand one of the most important causes of instrumentation failure is rotational instability.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:AIM: To assess biomechanical problems related to pedicle screw (PS) systems.MATERIAL and METHODS: Functional spinal units (L3-4) of deer were evaluated using a 6-axis material testing machine.
📊
Key Findings & Data:Eight directions were measured in the bending test: anterior, right-anterior, right,right-posterior, posterior, left-posterior, left, and left-anterior, and 2 directions were measured in the rotation test (right and left).RESULTS: ROMs of the PS model were smaller than those of the intact model in all directions.
💡
Conclusion & Impact:Basedon these results, we speculate that one of the most significant causes of adjacent segment disease is excessive bending rigidityand one of the most important causes of instrumentation failure is rotational instability.
🔬Design:Modeling & Simulation
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Özet

AIM: To assess biomechanical problems related to pedicle screw (PS) systems.MATERIAL and METHODS: Functional spinal units (L3-4) of deer were evaluated using a 6-axis material testing machine. For thespecimen models, we prepared an intact model, a damaged model, a PS model, and a crosslink model. We checked the range ofmotion (ROM) during bending and rotation tests. Eight directions were measured in the bending test: anterior, right-anterior, right,right-posterior, posterior, left-posterior, left, and left-anterior, and 2 directions were measured in the rotation test (right and left).RESULTS: ROMs of the PS model were smaller than those of the intact model in all directions. However, ROMs of the PS modelin the rotation test were smaller than those of the damaged model and larger than those of the intact model. The stability of thecrosslink model was better than that of the PS model during the bending test, but ROMs of the crosslink model were larger thanthose of the intact model during the rotation test.CONCLUSION: Excessive bending rigidity and rotational instability are the biomechanical problems related to PS systems. Basedon these results, we speculate that one of the most significant causes of adjacent segment disease is excessive bending rigidityand one of the most important causes of instrumentation failure is rotational instability.

🔖

Cite This Article

MIZUNO, T., SAKAKIBARA, T., YOSHIKAWA, T., INABA, T., KATO, T., KASAI, Y. (2019). Biomechanical Problems Related to the Pedicle Screw System. Turkish Neurosurgery, Cilt: 29 - Sayı: 1, 53–58. https://ajindex.com/articles/404413

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (1)

  1. Cheh G, Bridwell KH, Lenke LG, Buchowsli JM: Adjacent segment disease following lumbar/thoracolumbar fusion with pedicle screw instrumentation. Spine 32(20): 2253-2257, 2007 2. Fujiwara M, Masuda T, Inaba T, Katoh T, Kasai Y, Satoru I: Development of 6-axis material tester for measuring mechanical spine properties. Journal of Robotics and Mechatronics 18(2):160-165, 2006 3. Grob D, Benini A, Junge A, Mannion AF: Clinical experience with the Dynesys semirigid fixation system for the lumbar spine. Spine 30(3):324-331, 2005 4. Hadlow SV, Fagan AB, Hillier TM, Fraser RD: The Graf ligamentoplasty procedure. Spine 23(10):1172-1179, 1998 5. Jutte PC, Castelein RM: Complication of pedicle screws in lumbar and lumbosacral fusions in 105 consecutive primary operations. Eur Spine J 11:594-598, 2002 6. Kanayama M, Hashimoto T, Shigenobu K, Harada M, Oha F, Ohkoshi Y, Tada H, Yamamoto K, Yamane S: Adjacentsegment morbidity after Graf ligamentoplasty compared with posterolateral lumbar fusion. J Neurosurg 95:5-10, 2001 7. Kitahara K, Uni K, Hasegawa K, Takano K, Hara T: Biomechanical effect of fusion and stabilization on the adjacent segments in the lumbar spine. Journal of Japanese Society for Clinical Biomechanics 25:133-137, 2004 8. Kumar N, Kukreti S, Ishaque M, Mulholland R: Anatomy of deer spine and its comparison to the human spine. Anat Rec 260:189-203, 2000 9. Kumar N, Kukreti S, Ishaque M, Sengupta DK, Mulholland R: Functional anatomy of the deer spine: An appropriate biomechanical model for the human spine? Anat Rec 266:108- 117, 2002 10. Liu GM, Li YQ, Xu CJ, Zhu XM, Liu Y: Feasibility of vertebral internal fixation using deer and sheep as animal models. Chin Med J (Engl) 123(17):2379-2383, 2010 11. Lynn G, Mukherjee DP, Kruse RN, Sadasivan KK, Albright JA: Mechanical stability of thoracolumbar pedicle screw fixation. The effect of crosslink. Spine 22(14):1568-1573, 1997 12. Masaoka T, Inaba T, Kasai Y, Uchida A, Mogi M, Kato T, Nakamata T, Tokuda M: Biomechanical effect of spinal instrumentation on adjacent segments. Journal of Japanese Society for Clinical Biomechanics 29:305-310, 2008 13. Niosi CA, Zhu QA, Wilson DC, Keynan O, Wilson DR, Oxland TR: Biomechanical characterization of the three-dimensional kinematic behavior of the Dynesys dynamic stabilization system: An in vitro study. Eur Spine J 15:913-922, 2006 14. Park P, Garton HJ, Gala VC, Hoff JT, McGillicuddy JE: Adjacent segment disease after lumbar or lumbosacral fusion: Review of the literatu