Journals / Turkish Journal of Gastroenterology / 2021 / Cilt: 32 - Sayı: 8

Efficacy of Biodegradable Polydioxanone and Polylactic Acid Braided Biodegradable Biliary Stents for the Management of Benign Biliary Strictures

Pages
651–660
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

Conclusion: Our results project that biodegradable PLA and PDO braided biliary stents could be a better choice to treat BBS, with different rates of degradation.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Background: Benign biliary strictures (BBS) are widely treated by endoscopic procedures involving temporary stent placement.
📊
Key Findings & Data:Methods: Monofilaments of 3 different diameters, of PDO (0.30, 0.35, and 0.40 mm) and PLA (0.10, 0.15, and 0.20 mm) were braided to biliary stents and their mechanical properties were studied.
💡
Conclusion & Impact:Conclusion: Our results project that biodegradable PLA and PDO braided biliary stents could be a better choice to treat BBS, with different rates of degradation.
🔬Design:Modeling & Simulation
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Özet

Background: Benign biliary strictures (BBS) are widely treated by endoscopic procedures involving temporary stent placement. Occasionally, stents are required to be removed, making the treatment process very painful as well as expensive. Until now, no effective biodegradable biliary stents (BDBS) have been available for clinical applications. This study aims to investigate the safety and efficacy of biodegradable polydioxanone (PDO) and polylactic acid (PLA) braided BDBS both in vitro and in vivo. Methods: Monofilaments of 3 different diameters, of PDO (0.30, 0.35, and 0.40 mm) and PLA (0.10, 0.15, and 0.20 mm) were braided to biliary stents and their mechanical properties were studied. The stents were placed in an ex vivo bile duct model perfused with porcine bile, taken out, and observed every week until they were completely degraded. After the bile duct stenosis model was established successfully in piglet, stents with appropriate mechanical properties were further examined under endoscopy; the changes in hematology, patency time, and pathology of the stents were observed for 8 months. Results: A total of 10 pigs were included (2 groups; 5 PDO, 5 PLA) in the study. The patency time of the stents in the PLA group was significantly longer than that in the PDO group (25.7 ± 5.6 weeks vs 11.3 ± 3.4 weeks, respectively). Conclusion: Our results project that biodegradable PLA and PDO braided biliary stents could be a better choice to treat BBS, with different rates of degradation.

🔖

Cite This Article

ZHANG, W., KANWAL, F., REHMAN, M. F. U., WAN, X. (2021). Efficacy of Biodegradable Polydioxanone and Polylactic Acid Braided Biodegradable Biliary Stents for the Management of Benign Biliary Strictures. Turkish Journal of Gastroenterology, Cilt: 32 - Sayı: 8, 651–660. https://ajindex.com/articles/399379

References (13)

  1. Kaffes A. Benign biliary strictures: how we are evolving to the perfect endoscopic strategy. Endoscopy. 2019
  2. 51(12):1115-1116. [CrossRef] 2. Singhi AD, Nikiforova MN, Chennat J, et al. Integrating next-generation sequencing to endoscopic retrograde cholangiopancreatography (ERCP)-obtained biliary specimens improves the detection and management of patients with malignant bile duct strictures. Gut. 2020
  3. 69(1):52-61. [CrossRef] 3. Kapoor BS, Mauri G, Lorenz JM. Management of biliary strictures: state-of-the-art review. Radiology. 2018
  4. 289(3):590-603. [CrossRef] 4. Mukai S, Itoi T, Sofuni A, et al. EUS-guided antegrade intervention for benign biliary diseases in patients with surgically altered anatomy (with videos). Gastrointest Endosc. 2019
  5. 89(2):399-407. [CrossRef] 5. Dumonceau JM, Tringali A, Blero D, et al. Biliary stenting: indications, choice of stents and results: European Society of Gastrointestinal Endoscopy (ESGE) Clinical Guideline. Endoscopy. 2012
  6. 44(3):277-298. [CrossRef] 6. Wong MYW, Saxena P, Kaffes AJ. Benign biliary strictures: a systematic review on endoscopic treatment options. Diagnostics (Basel). 2020
  7. 10(4):221. [CrossRef] 7. Dumonceau JM, Tringali A, Papanikolaou IS, et al. Endoscopic biliary stenting: indications, choice of stents, and results: European Society of Gastrointestinal Endoscopy (ESGE) clinical guideline– updated October 2017. Endoscopy. 2018
  8. 50(9):910-930. [CrossRef] 8. Mauri G. Endoscopic plastic stents: still the preferred option of treatment for benign biliary stenosis? Gastrointest Endosc. 2019
  9. 90(6):998-999. [CrossRef] 9. Tarantino I, Amata M, Cicchese N, et al. Sequential multistenting protocol in biliary stenosis after liver transplantation: a prospective analysis. Endoscopy. 2019
  10. 51(12):1130-1135. [CrossRef] 10. Häner JD, Räber L, Windecker S. Biodegradable vs. permanent polymer drug-eluting stents: the need for a new nomenclature to classify drug-eluting stent technology. Eur Heart J. 2019
  11. 40(31):2616- 2619. [CrossRef] 11. Sato T, Kogure H, Nakai Y, et al. Long‐term outcomes of endoscopic treatment for duct‐to‐duct anastomotic strictures after living donor liver transplantation. Liver Int. 2019
  12. 39(10):1954-1963. [CrossRef] 12. Iwasaki H, Mizushima T, Suzuki Y, et al. Factors that affect stent-related complications in patients with malignant obstruction of the esophagus or gastric cardia. Gut Liver. 2017
  13. 11(1):47-54. [CrossRef] 13. Buiten RA, Ploumen EH, Zocca P, et al. Thin, very thin, or ultrathin strut biodegr