Journals / Gıda / 2017 / Cilt: 42 - Sayı: 3

CONSTRUCTION OF A pH ELECTRODE BASED ON ZOFENOPRIL CALCIUM FOR THE DETERMINATION OF TITRATABLE ACIDITY IN SOME BEVERAGES

BAZI İÇECEKLERDE TİTRASYON ASİTLİĞİNİN TAYİNİ İÇİN ZOFENOPRİL KALSİYUMA DAYANAN pH ELEKTROT YAPIMI

Journal
Gıda
Pages
211–218
Publication Date
2017-06-15
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

Also, it was found to display good selectivity for H+ ions in the presence of various ions.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:In this study, a new PVC membrane pH electrode based on zofenopril calcium (ZFNCa) available for the determination of titratable acidity in some beverages was developed.
📊
Key Findings & Data:The electrode exhibited a slope of 44.1±1.7 mV/pH in the pH range 1.7–9.5 at 20±1 °C.
💡
Conclusion & Impact:Also, it was found to display good selectivity for H+ ions in the presence of various ions.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:PVC membran · pH elektrot · zofenopril kalsiyum

Abstract

In this study, a new PVC membrane pH electrode based on zofenopril calcium (ZFNCa) available for the determination of titratable acidity in some beverages was developed. The electrode exhibited a slope of 44.1±1.7 mV/pH in the pH range 1.7–9.5 at 20±1 °C. Furthermore, it was observed that the electrode had good repeatability and reproducibility with a response time of 15 s and a lifetime of at least 12 months. Also, it was found to display good selectivity for H+ ions in the presence of various ions. The applicability of the proposed electrode for the determination of titratable acidity in some beverages (orange juice, apple juice, fizzy drink, beer, wine and vinegar) was illustrated. It was seen that there were no significant differences between the results obtained with the proposed electrode and the traditional glass pH electrode at the 95% CL (confidence level). As a consequence, it was seen that the electrode based on ZFNCa could be successfully used as an alternative for glass electrode to determine the titratable acidity in beverages.

Özet

Bu çalışmada, bazı içeceklerde titrasyon asitliğinin tayini için kullanılabilecek zofenopril kalsiyum’a (ZFNCa) dayanan yeni bir PVC membran pH elektrot geliştirilmiştir Elektrot, 20±1 °C’de, 1.7–9.5 pH aralığında, 44.1±1.7 mV/pH’lik bir eğim göstermektedir. Ayrıca, bu elektrodun, 15 s’lik cevap süresi ve en az 12 aylık ömürle, iyi bir tekrarlanabilirliğe ve yeniden üretilebilirliğe sahip olduğu gözlenmiştir. Aynı zamanda, çeşitli iyonlar varlığında H+ iyonuna karşı iyi bir seçicilik gösterdiği belirlenmiştir. Önerilen elektrot kullanılarak, bazı içeceklerdeki (portakal suyu, elma suyu, gazlı içecek, bira, şarap ve sirke) titrasyon asitliğinin tayininin yapılabileceği gösterilmiştir. Bulunan sonuçlar geleneksel cam pH elektrotla elde edilenlerle karşılaştırıldığında, %95 GS’de (güven seviyesi), aralarında anlamlı bir fark olmadığı belirlenmiştir. Sonuç olarak, ZFNCa’a dayanan bu elektrodun, içeceklerdeki titrasyon asitliği tayininde cam elektroda alternatif olarak başarıyla kullanılabildiği görülmüştür.

🔖

Cite This Article

ERTÜRÜN, H. E. K. (2017). CONSTRUCTION OF A pH ELECTRODE BASED ON ZOFENOPRIL CALCIUM FOR THE DETERMINATION OF TITRATABLE ACIDITY IN SOME BEVERAGES. Gıda, Cilt: 42 - Sayı: 3, 211–218. https://ajindex.com/articles/39303

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (14)

  1. Cemeroğlu B.S (ed). 2013. Gıda Analizleri. Bizim Grup Basımevi Ajans Tan. Org. Yay. Dağ.San. Tic. Ltd. Şti. Ankara, Türkiye, 475 s
  2. Crespo G.A., Afshar M.G., Bakker E. 2012. Direct Detection of Acidity, Alkalinity, and pH with Membrane. Anal. Chem., 84, 10165-10169
  3. Tayfur M (ed). 2014. Gıda Katkı Maddeleri. Detay Yayıncılık Ankara, Türkiye, 230 s
  4. Ansari R., Arvand M., Heydari L. 2014. The behaviour of polyaniline-coated PVC membrane based on 7,16-didecyl-1, 4, 10, 13-tetraoxa-7, 16-diazacyclooctadecane for pH measurements in highly acidic media. J. Chem. Sci., 126, 41-48
  5. Peng L.B., Heng L.Y., Hasbullah S.A., Ahmad M. 2007. A solid-state pH transducer fabricated from a self-plasticized methacrylic-acrylic membrane for potentiometric acetylcholine chloride biosensor. J. Anal. Chem., 62, 884-888
  6. Kim B., Shim J., Chung K.C. 2011.Study on hydrogen ion-selective solid contact electrodes based on decamethylcyclopentasiloxane, Anal. Lett., 44, 2138-2149
  7. Crespo G.A., Gugsa D., Macho S., Rius F. X. 2009. Solid-contact pH-selective electrode using multi-walled carbon nanotubes. Anal. Bioanal. Chem., 395, 2371-2376
  8. Michalska A., Hulanicki A., Lewenstam A. 1994. All solid-state hydrogen ion-selective electrode based on a conducting poly(pyrrole) solid contact. Analyst, 119, 2417-2420
  9. Faria R.C., Bulhões L.O.S. 1998. Hydrogen ion selective electrode based on poly(1-aminoanthracene) film. Anal. Chim. Acta, 377, 21-27
  10. Zine N., Bausells J., Ivorra A., Aguiló J., Zabala M., Teixidor F., Masalles C., Viñas C., Errachid A. 2003. Hydrogen-selective microelectrodes based on silicon needles. Sensor Actuat B, 91, 76-82
  11. Han W.S., Chung K.C., Kim M.H., Ko H.B., Lee Y.H., Hong T.K. 2004. A hydrogen ion-selective poly(aniline) solid contact electrode based on dibenzylpyrenemethylamine ionophore for highly acidic solutions, Anal. Sci., 20, 1419-1422
  12. Han W.S., Park M.Y., Chung K.C., Cho K.C., Hong T.K. 2001. All solid state hydrogen ion selective electrode based on a tribenzylamine neutral carrier in a poly(vinyl chloride) membrane with a poly(aniline) solid contact. Electroanal., 13, 955-959
  13. Han W.S., Park M.Y., Chung K.C., Cho K.C., Hong T.K., 2001. Potentiometric sensor for hydrogen ion based on N,N'-dialkylbenzylethylenediamine neutral carrier in a poly(vinyl chloride) membrane with polyaniline solid contact. Talanta, 54, 153-159
  14. Alexander P.W., Dimitrakopoulos T., Hibbert D.B. 1997. Photo-cured ammonium and hydrogen ion se