Should physicians instead of industry representatives be the main actor of cardiac implantable electronic device follow-up? (Super Follow-up)

Pages
23–30
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

Therefore, our study strongly demonstrates that CIED follow-up should not be allowed to be performed entirely by manufacturers’ representatives alone

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Objective: This retrospective study sought to research the adequacy of the follow-up and optimization of cardiac implantable electronic devices (CIEDs) performed by industry representatives.Methods: A total of 403 consecutive patients (35% females; median age, 67 years; age range 18–97 years) with either pacemakers (n=246), implantable cardioverter-defibrillators (ICDs), (n=117) or cardiac resynchronization therapy with defibrillator (CRT-D) (n=40) applied to our hospital’s outpatient pacemaker clinic for follow-up.
📊
Key Findings & Data:These patients had been followed up by industry representatives alone until September 2013 and then by a cardiologist who is dealing with cardiac electrophysiology and has a knowledge of CIED follow-up.Results: It was ascertained that 117 (47.6%) of 246 patients with pacemakers had a programming error.
💡
Conclusion & Impact:Therefore, our study strongly demonstrates that CIED follow-up should not be allowed to be performed entirely by manufacturers’ representatives alone
🔬Design:Modeling & Simulation
📍Scope:2013 Verileri
⚙️Variables:Domain Parameters

Abstract

Objective: This retrospective study sought to research the adequacy of the follow-up and optimization of cardiac implantable electronic devices (CIEDs) performed by industry representatives.Methods: A total of 403 consecutive patients (35% females; median age, 67 years; age range 18–97 years) with either pacemakers (n=246), implantable cardioverter-defibrillators (ICDs), (n=117) or cardiac resynchronization therapy with defibrillator (CRT-D) (n=40) applied to our hospital’s outpatient pacemaker clinic for follow-up. These patients had been followed up by industry representatives alone until September 2013 and then by a cardiologist who is dealing with cardiac electrophysiology and has a knowledge of CIED follow-up.Results: It was ascertained that 117 (47.6%) of 246 patients with pacemakers had a programming error. Forty-three (36.8%) of 117 patients were symptomatic, and after reprogramming, all symptoms diminished partially or completely during the follow-up. Moreover, 30 (25.6%) of 117 patients with ICDs had a programming error. Furthermore, 6 (15%) of 40 patients with CRT-Ds had a programming error. To conclude, when all patients with CIEDs were assessed together, it was ascertained that 153 (38%) of 403 patients had programming errors.Conclusion: The prevalence of inappropriate programming of CIEDs by industry representatives was quite higher than expected. Therefore, our study strongly demonstrates that CIED follow-up should not be allowed to be performed entirely by manufacturers’ representatives alone

🔖

Cite This Article

ÜREYEN, Ç. M., BAŞ, C. Y., YÜKSEL, İ. Ö., KUŞ, G., ÇAĞIRCI, G., ARSLAN, Ş. (2017). Should physicians instead of industry representatives be the main actor of cardiac implantable electronic device follow-up? (Super Follow-up). The Anatolian Journal of Cardiology, Cilt: 18 - Sayı: 1, 23–30. https://ajindex.com/articles/366721

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (10)

  1. Eucomed. Medical Technology - Key Facts and Figures.http://www. eucomed.org/ medical-technology/facts-figures Greenspon AJ, Patel JD, Lau E, Ochoa JA, Frisch DR, Ho RT, et al. Trends in permanent pacemaker implantation in the United States from 1993 to 2009: increasing complexity of patients and proce- dures. J Am Coll Cardiol 2012; 60: 1540-5.
  2. John Camm A, Nisam S. European utilization of the implantable de- fibrillator: has 10 years changed the 'enigma'? Europace 2010; 12: 1063-9.
  3. Wilkoff BL, Auricchio A, Brugada J, Cowie M, Ellenbogen KA, Gillis AM, et al. HRS/EHRA expert consensus on the monitoring of car- diovascular implantable electronic devices (CIEDs): description of techniques, indications, personnel, frequency and ethical conside- rations. Heart Rhythm 2008; 5: 907-25.
  4. Shapland JE, MacCarter D, Tockman B, Knudson M. Physiologic benefits of rate responsiveness. Pacing Clin Electrophysiol 1983; 6: 329-32.
  5. Andersen HR, Nielsen JC, Thomsen PE, Thuesen L, Mortensen PT, Vesterlund T, et al. Long-term follow-up of patients from a ran- domised trial of atrial versus ventricular pacing for sick sinus syn- drome. Lancet 1997; 350: 1210-6.
  6. Sweeney MO, Hellkamp AS, Ellenbogen KA, Greenspon AJ, Freed- man RA, Lee KL, et al. Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation 2003; 107: 2932-7.
  7. Gillis AM, Pürerfellner H, Israel CW, Sunthorn H, Kacet S, Anelli- Monti M, et al. Reducing unnecessary right ventricular pacing with the managed ventricular pacing mode in patients with sinus node disease and AV block. Pacing Clin Electrophysiol 2006; 29: 697-705.
  8. Nielsen JC, Kristensen L, Andersen HR, Mortensen PT, Pedersen OL, Pedersen AK. A randomized comparison of atrial and dual- chamber pacing in 177 consecutive patients with sick sinus syn- drome: echocardiographic and clinical outcome. J Am Coll Cardiol 2003; 42: 614-23.
  9. McMurray JJ, Adamopoulos S, Anker SD, Auricchio A, Böhm M, Dickstein K, et al. ESC Guidelines for the diagnosis and treatment of acute and Chronic Heart Failure 2012: The Task Force for the Diagnosis and Treatment of Acute and Chronic Heart Failure 2012 of the European Society of Cardiology. Developed in collaboration with the Heart Failure Association (HFA) of the ESC. Eur Heart J 2012; 33: 1787-847.
  10. Kachboura S, Ben Halima A, Fersi I, Marrakchi S, Zouaoui W, Kam- moun I. Assessment of hea