Overall, 5 cardiac deaths and 2 definite stent thrombosis were observed in the study.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Objective: Even with the immense progress achieved in the field of percutaneous coronary interventions (PCIs), treatment of diffuse long atherosclerotic coronary artery disease continues to remain a challenge for durable outcomes.
📊
Key Findings & Data:The downstream reduction in diameter along the lesion length of a coronary artery may compel the cardiologist to use either 2 overlapping stents of different diameters or a single long stent leading to stent-vessel mismatch at the edges.
💡
Conclusion & Impact:Overall, 5 cardiac deaths and 2 definite stent thrombosis were observed in the study.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Özet

Objective: Even with the immense progress achieved in the field of percutaneous coronary interventions (PCIs), treatment of diffuse long atherosclerotic coronary artery disease continues to remain a challenge for durable outcomes. The downstream reduction in diameter along the lesion length of a coronary artery may compel the cardiologist to use either 2 overlapping stents of different diameters or a single long stent leading to stent-vessel mismatch at the edges. Recently, Meril Life Sciences Pvt. Ltd., India, has introduced a long-tapered sirolimus-eluting stent (SES) system, BioMime Morph, which conforms to the normal tapered geometry of coronary arteries along with adequate lesion coverage. In this study, we aimed to provide real world experience regarding the safety and efficacy of the BioMime Morph SES over a follow-up of one year. Methods: This was a single center, retrospective study involving 172 participants who underwent PCI with the BioMime Morph SES. Mean length of the target lesion was 34.4±10.4 mm, and mean stent length was 53.2±8.7 mm. The most frequent revascularized vessel was the left anterior descending artery (LAD) in 97 lesions (54.4%). Results: Major adverse cardiac events (MACE) (defined as a composite of target vessel myocardial infarction, target lesion revascularization, and death due to a cardiac cause) at 1, 6, and 12 months were seen in 4 (2.3%), 7 (4.0%), and 8 (4.7%) patients, respectively. Overall, 5 cardiac deaths and 2 definite stent thrombosis were observed in the study. Conclusion: The study suggests that the novel BioMime Morph SES is an effective and a safe option for PCI in the treatment of long diffuse atherosclerotic lesions.

🔖

Cite This Article

SHARMA, Y. P., UPPAL, L., PANDA, P., MOHANTY, S., KASİNADHUNİ, G., KRİSHNAPPA, D., MEHROTA, S., GUPTA, A., PRASAD, K., SANTOSH, K., BOOTLA, D., GHOSH, S. (2021). One-year outcomes of novel BioMime Morph tapered stent in long and multiple coronary artery lesions. The Anatolian Journal of Cardiology, Cilt: 25 - Sayı: 12, 896–901. https://ajindex.com/articles/365124

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (12)

  1. Russo RJ, Silva PD, Yeager M. Coronary artery overexpansion increases neointimal hyperplasia after stent placement in a porcine model. Heart 2007
  2. [Crossref] 2. Phan TQ, Nguyen LH, Nguyen LV, Lee WS, Won H, Cho I, et al. Imaging Characteristics of Mismatch Lesions: An Angiographic and Intravascular Ultrasound Analysis of 1369 Coronary Lesions. Angiology 2019
  3. [Crossref] 3. Zubaid M, Buller C, Mancini GB. Normal angiographic tapering of the coronary arteries. Can J Cardiol 2002
  4. Zhang LR, Xu DS, Liu XC, Wu XS, Ying YN, Dong Z, et al. Coronary artery lumen diameter and bifurcation angle derived from CT coronary angiographic image in healthy people. Zhonghua Xin Xue Guan Bing Za Zhi 2011
  5. [Article in Chinese] 5. Rikhtegar F, Edelman ER, Olgac U, Poulikakos D, Kurtcuoglu V. Drug deposition in coronary arteries with overlapping drug-eluting stents. J Control Release 2016
  6. [Crossref] 6. Patted SV, Jain RK, Jiwani PA, Suryavanshi S, Raghu TR, Raveesh H, et al. Clinical Outcomes of Novel Long-Tapered Sirolimus-Eluting Coronary Stent System in Real-World Patients With Long Diffused De Novo Coronary Lesions. Cardiol Res 2018
  7. [Crossref] 7. Lupi A, Ugo F, De Martino L, Infantino V, Iannaccone M, Iorio S, et al. Real-World Experience With a Tapered Biodegradable Polymer-Coated Sirolimus-Eluting Stent in Patients With Long Coronary Artery Stenoses. Cardiol Res 2020
  8. [Crossref] 8. Mintz GS, Popma JJ, Pichard AD, Kent KM, Satler LF, Chuang YC, et al. Patterns of calcification in coronary artery disease. A statistical analysis of intravascular ultrasound and coronary angiography in 1155 lesions. Circulation 1995
  9. [Crossref] 9. Sianos G, Papafaklis MI, Serruys PW. Angiographic thrombus burden classification in patients with ST-segment elevation myocardial infarction treated with percutaneous coronary intervention. J Invasive Cardiol 2010
  10. 22 (10 Suppl B): 6B-14B. 10. Eleid MF, Guddeti RR, Tweet MS, Lerman A, Singh M, Best PJ, et al. Coronary artery tortuosity in spontaneous coronary artery dissection: angiographic characteristics and clinical implications. Circ Cardiovasc Interv 2014
  11. [Crossref] 11. Mauri L, Hsieh WH, Massaro JM, Ho KK, D'Agostino R, Cutlip DE. Stent thrombosis in randomized clinical trials of drug-eluting stents. N Engl J Med 2007
  12. [Crossref] 12. Applegate RJ, Sacrinty MT, Kutcher MA, Santos RM, Gandhi SK, Little WC. Effect of lengt