The Effect of Pinna Position on Body Temperature Measurements Made with a Tympanic Membrane Thermometer in Pediatric Patients

Pages
132–138
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

During temperature measurement with pinna positioning and without pinna positioning,patients felt no discomfort.Conclusion: In measurements carried out using tympanic membrane thermometers in pediatric patients, positioning the auricle by pulling itdownward posteriorly yielded more reliable and correct outcomes.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Aim: The aim of this study was to investigate the effect of pinna position on body temperature measurements made with a tympanic membranethermometer in pediatric patients.Materials and Methods: This study was conducted with a quasi-experimental design employing a pre-test and post-test.
📊
Key Findings & Data:For analysis of thedata, frequencies, percentages, means and standard deviations were calculated, and the significance of the difference between paired valueswas tested in order to investigate the effects of the auricle position on measurement values.Results: The age of the patients included in the study ranged between 6 and 13 years, and the mean age was 10.25±1.83 years.
💡
Conclusion & Impact:During temperature measurement with pinna positioning and without pinna positioning,patients felt no discomfort.Conclusion: In measurements carried out using tympanic membrane thermometers in pediatric patients, positioning the auricle by pulling itdownward posteriorly yielded more reliable and correct outcomes.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Özet

Aim: The aim of this study was to investigate the effect of pinna position on body temperature measurements made with a tympanic membranethermometer in pediatric patients.Materials and Methods: This study was conducted with a quasi-experimental design employing a pre-test and post-test. For analysis of thedata, frequencies, percentages, means and standard deviations were calculated, and the significance of the difference between paired valueswas tested in order to investigate the effects of the auricle position on measurement values.Results: The age of the patients included in the study ranged between 6 and 13 years, and the mean age was 10.25±1.83 years. The meandifference between measurements in the two different positions was 0.35°C. The measurements made in each position were representedin a Bland Altman plot. It was seen that the differences between the two positions were not distributed around zero, but instead showeda systematic distribution around 0.35°C. There was a significant relationship between the differences and the mean values. The differencebetween the mean durations of the measurements was found to be 1.07 seconds longer with pinna positioning than without. The discomfortlevels of the patients during temperature measurement without pinna positioning and with pinna positioning ranged between 0-2 and 0-4according to the facial expressions scale respectively. During temperature measurement with pinna positioning and without pinna positioning,patients felt no discomfort.Conclusion: In measurements carried out using tympanic membrane thermometers in pediatric patients, positioning the auricle by pulling itdownward posteriorly yielded more reliable and correct outcomes.

🔖

Cite This Article

ORKUN, N., EŞER, İ. (2020). The Effect of Pinna Position on Body Temperature Measurements Made with a Tympanic Membrane Thermometer in Pediatric Patients. The Journal of Pediatric Research, Cilt: 7 - Sayı: 2, 132–138. https://ajindex.com/articles/362997

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (16)

  1. Chaglla JSE, Celik N, Balachandran W. Measurement of core body temperature using graphene-inked infrared thermopile. Sensor 2018;18:3315-23
  2. Annane D. Body temperature in sepsis: a hot topic. The Lancet 2018;6:162-3
  3. Bayhan C, Özsürekçi Y, Tekçam N, et al. Comparison of infrared tympanic thermometer with non-contact infrared thermometer. J Pediatr Inf 2014;8:52-5
  4. Yang WC, Kuo HT, Lin CH, et al. Tympanic temperature versus temporal temperature in patients with pyrexia and chills. Medicine 2016;95:1-6
  5. Çakırcalı E. Fundamentals of Nursing:Vital Signs., Istanbul: Akademi publishing, 2013:579-591
  6. Potter PA, Perry AG. Fundamentals of Nursing. St. Louis: Elsevier Inc, 7. th ed, 2009:98-116
  7. Arslan GG, Eşer İ, Khorshıd L. Analysis of the effect of lying on the ear on body temperature measurement using a tympanic thermometer. J Pak Med Assoc 2011;61:1065-8
  8. Karamanoğlu AY, Korkmaz FD. Evidences and practice recommendations for the measurement of non-invasive body temperature in the emergency department. International Refereed Journal of Nursing Researchers 2015;2:71-90
  9. Giuliano KK, Scott SS, Elliot S, et al. Temperature measurement in critically ill adults: a comparison of tympanic ve oral methods. Am J Crit Care 2000;9:254-61
  10. Poveda VB, Nascimento AS. Intraoperative body temperature control: esophageal thermometer versus infrared tympanic thermometer. Rev Esc Enferm USP 2016;50:946-52
  11. Muth M, Statler J, Gentıle DL, Hagle ME. Frequency of fever in pediatric patients presenting to the emergency depart¬ment with non-ıllness-related conditions. J Emerg Nurs 2013;39:389-92
  12. Çultu Ö, Yıldırım İ, Ceyhan M, et al. Comparing body temperature measurements by mothers and physicians using mercury-in-glass, digital mercury and infrared tympanic membrane thermometers in healthy newborn babies. Turk J Pediatr 2008;50:354-8
  13. Health Directorate of Istanbul. [Cited: 07 May 2015.] Istanbul, Turkey, Retrieved from: http://www.istanbulsaglik. gov. tr/ w/ sb/ecz/mevzuat/mevzuat.asp
  14. Pour HA, Yavuz M. Effect of body temperature height (fever) on the homodynamic parameters. Maltepe University Journal of Nursing Science and Art 2010;3:73-9
  15. Edelu BO, Ojinnaka NC, Ikefuna AN. Fever detection in under 5 children in a tertiary health facility using the infrared tympanic thermometer in the oral mode. Ital J Pediatr 2011;37:1-6
  16. Obermeyer Z, Samra JK, Mullainathan S. Individual differences in normal body temperature: longitudinal bi