Comparison of intraocular pressure measurements performed with noncontact tonometer and Goldmann applanation tonometer

Nonkontakt tonometre ve Goldmann aplanasyon tonometresi ile yapılan göziçi basıncı ölçümlerinin karşılaştırılması

Pages
343–345
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

CONCLUSION: IOP measurements obtained with XPERT were in correlation with GAT results particularly at normal IOP levels, however XPERT measurements should be confirmed with GAT in glaucomapatients.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:PURPOSE: To evaluate the reliability of intraocular pressure (IOP) measurements obtained with a new generation noncontact tonometer, Reichert XPERT plus (XPERT).
📊
Key Findings & Data:METHODS: IOP measurements obtained with XPERT was compared with those from Goldmann applanation tonometer (GAT) in 100 eyes of 50 volunteers.
💡
Conclusion & Impact:CONCLUSION: IOP measurements obtained with XPERT were in correlation with GAT results particularly at normal IOP levels, however XPERT measurements should be confirmed with GAT in glaucomapatients.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Glokom · Göz içi basınç · Tonometri

Abstract

PURPOSE: To evaluate the reliability of intraocular pressure (IOP) measurements obtained with a new generation noncontact tonometer, Reichert XPERT plus (XPERT). METHODS: IOP measurements obtained with XPERT was compared with those from Goldmann applanation tonometer (GAT) in 100 eyes of 50 volunteers. RESULTS: The difference between IOP measurements obtained with XPERT and GAT was less than 2mmHg in 61.5% and less than 3mmHg in 77.1% of the eyes. There was a high correlation between two measurement methods (r=0.81, p<0.001). The equation of regression was XPERT=3.4 + (0.77 x GAT), (p<0.0001). The Standard deviation of differences was 2.66mmHg. CONCLUSION: IOP measurements obtained with XPERT were in correlation with GAT results particularly at normal IOP levels, however XPERT measurements should be confirmed with GAT in glaucomapatients.

Özet

AMAÇ: Yeni nesil nonkontakt tonometrelerden birisi olan Reichert XPERT plus NCT'nin (XPERT) güvenilirliğinin değerlendirilmesi. YÖNTEM: Elli gönüllünün 100 gözünde XPERT ve Goldmann aplanasyon tonometresi (GAT) ile yapılan göziçi basıncı (GİB} ölçümleri çeşitli istatistiksel yöntemlerle karşılaştırıldı. BULGULAR: XPERT ve GAT ile yapılan GİB ölçümleri arasındaki fark, gözlerin %61.5'inde 2 mmHg'dan az ve %77.1'inde 3.0 mmHg'dan düşüktü. İki yöntem arasında yüksek oranda korelasyon saptandı (r=0.81, p<0.001). Regresyon eğrisi XPERT=3.4 + (0.77 X GAT) olarak hesaplandı (p<0.0001). Eş ölçümler arası farkın standart sapması 2.66mmHg olarak bulundu. SONUÇ: XPERT ile yapılan ölçümlerinin özellikle normal GİB değerlerinde GAT ile uyumlu olduğu, ancak glokom hastaları gibi ölçüm hatalarının tedaviyi etkileyebileceği hastalarda XPERT Ölçümlerinin GAT ile kontrol edilmesi gerektiği kanısına varıldı.

🔖

Cite This Article

AKMAN, A., YAYLALI, V., ÜNAL, M., SÖNMEZ, M., ÖRGE, Y. (1999). Comparison of intraocular pressure measurements performed with noncontact tonometer and Goldmann applanation tonometer. MN Oftalmoloji, Cilt: 6 - Sayı: 4, 343–345. https://ajindex.com/articles/296505